Skip to content

GAMMA 3525-0320SRecalls & reported problems

Stryker GmbH · Long Nail Kit R1.5, Ti, LEFT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540375728), received between May 11, 2018 and Oct 31, 2023 — classified as 5 injuries.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Break2

Patient problems

  • Failure of Implant4
  • Nonunion/Delayed-union Bone Fracture2
  • Injury1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 31, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "DUE TO POOR REDUCTION AT THE PRIMARY SURGERY ON (B)(6) 2023 A NON-UNION WAS OCCURRED AND THE NAIL FRACTURE WAS OCCURRED. REVISION SURGERY IS SCHEDULED FOR REPLACEMENT OF THE GAMMA NAIL."

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 18044280 ↗
  • InjuryJun 19, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A GAMMA NAIL WAS BROKEN DUE TO PSEUDARTHROSIS POSTOPERATIVELY.

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 17151413 ↗
  • InjuryJun 22, 2021Fracture; Fracture

    IT WAS REPORTED THE LONG GAMMA NAIL BROKE AS A RESULT OF CONTINUOUS STRESS CONCENTRATION ON THE NAIL DUE TO POOR REDUCTION. A REPLACEMENT OPERATION WAS CARRIED OUT ON (B)(6) 2021.

    Patient problems: Failure of Implant

    MAUDE report 12045887 ↗
  • InjuryJun 11, 2020Fracture; Fracture

    IT WAS REPORTED: A (B)(6) YEAR OLD, MALE PATIENT WAS DISCHARGED FROM THE HOSPITAL ONE WEEK AFTER ORIF PROCEDURE WITH GAMMA NAIL ON (B)(6) 2020. AT THAT TIME, HE WAS ABLE TO WALK NORMALLY USING CRUTCHES, AND FOLLOWED UP WITH ANOTHER HOSPITAL. HOWEVER; ON (B)(6) THE PATIENT WAS RETURNED FROM ANOTHER HOSPITAL TO THE HOSPITAL WHERE HE HAD SURGERY DUE TO A POSSIBLE NAIL BREAKAGE. HE WAS VERY HEAVY ((B)(6)) AND AT THE TIME OF THE INITIAL SURGERY THE SURGEON ATTEMPTED TO WRAP THE FEMUR WITH A CABLE, BUT IT COULD NOT WRAP WELL AND THERE WAS NO BONE CONTACT INSIDE. SO PROBABLY, ALL THE NAILS WERE LOADED AND BROKE. REVISION IS SCHEDULED ON (B)(6).

    Patient problems: Failure of Implant; Injury

    MAUDE report 10144817 ↗
  • InjuryMay 11, 2018Fracture

    ON (B)(6) 2016, LONG GAMMA NAIL WAS IMPLANTED AS A TREATMENT FOR TROCHANTER FRACTURE. PARTIAL LOAD WAS STARTED 1 WEEKS LATER AFTER SURGERY AND TOTAL LOAD STARTED AT 9 WEEKS. FELT SOMETHING STRANGE IN (B)(6) 2018, VISITED THE HOSPITAL ON (B)(6) 2018. FRACTURE OF THE PROXIMAL PART OF THE NAIL WAS DISCOVERED COMPARED WITH RADIOGRAPHING. IT WAS REPLACED ON (B)(6) WITH A REPLACEMENT OF LONG GAMMA.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7507091 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page