Skip to content

GAMMA 3525-0360SRecalls & reported problems

Stryker GmbH · Long Nail Kit R1.5, Ti, LEFT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540375742), received between Jul 12, 2019 and Oct 15, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Break3
  • Appropriate Device Problem Term/Code Not Available1
  • Device Contaminated During Manufacture or Shipping1

Patient problems

  • Failure of Implant3
  • Nonunion/Delayed-union Bone Fracture3
  • Discomfort1
  • Fall1
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025Fracture; Fracture

    AS REPORTED: "IMPLANT FRACTURE AFTER NON-CONSOLIDATED FRACTURE, IMPLANTATION (B)(6) 2023. EVIDENCE OF A NAIL FRACTURE, AT THE LEVEL OF THE FEMORAL NECK SCREW, IN PSEUDARTHROSIS AFTER PROXIMAL FEMUR FRACTURE AS A RESULT OF A TRAFFIC ACCIDENT WITH IMPLANTATION OF A LONG GAMMA NAIL ON (B)(6) 2023. ADIPOSITY PER MAGNA 160KG AT 1.6M. PAIN OVER THE AFFECTED HIP SINCE THE BEGINNING OF AUGUST. EXPLANTATION OF THE BROKEN IMPLANT, IMPLANTATION OF A NEW IMPLANT."

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 23300238 ↗
  • InjuryJul 8, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A CASE OF SUBCONDYLAR FRACTURE WAS TREATED WITH GAMMA 3 NAILS, BUT A PSEUDOARTHROSIS OCCURRED AND THE NAILS AND SCREWS WERE BROKEN OFF. THE PATIENT WAS REPLACED WITH A NAIL FROM ANOTHER COMPANY ON JUNE 13TH. NAIL HAS A BROKEN LAG SCREW HOLE SECTION, AND TWO LOCKING SCREWS WERE ALSO BROKEN.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 22444994 ↗
  • InjuryDec 12, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "POSTOPERATIVELY GAMMA3 LONG NAIL WAS BROKEN AND A REVISION SURGERY WAS PERFORMED. THE BROKEN PART IS THE LAG SCREW HOLE."

    Patient problems: Failure of Implant

    MAUDE report 15963396 ↗
  • InjuryMar 5, 2021Fracture; Fracture

    AS REPORTED: "PATIENT FELL, GAMMA NAIL BROKE AT LAG SCREW JUNCTION." PATIENT WAS REVISED TO A GMRS PROXIMAL FEMUR WITH BIPOLAR. ADDITIONAL INFO RECEIVED FROM THE REPRESENTATIVE (B)(6) 2021: DOCTOR HAS INDICATED THAT THE ¿PATHOLOGIC FRACTURE W RADIATION, BONE NEVER HEALED, PATHOLOGIC NON-UNION¿.

    Patient problems: Failure of Implant; Fall; Nonunion/Delayed-union Bone Fracture

    MAUDE report 11426288 ↗
  • MalfunctionJul 1, 2020Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED: "WHEN I OPENED THE LONG GAMMA 3 NAIL AND TRIED TO SET IT, I NOTICED THAT WHITE DEBRIS HAD ADHERED TO THE NAIL. I TOOK THE ONE THAT ADHERED TO THE SURFACE, BUT IT MIGHT STILL BE IN THE NAIL. A NEW NAIL WAS OPENED AND USED. THE WHITE DEBRIS WAS THE SCRAPED OFF SPACER THAT FIXED THE NAIL. IT WAS DISCARDED."

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10218330 ↗
  • InjuryJan 14, 2020Fracture; Fracture

    PRIMARY GAMMA3 SURGERY WAS PERFORMED ON (B)(6) 2019. 7 MONTHS AFTER THE SURGERY, THE PATIENT COMPLAINED DISCOMFORT AND THE NAIL BREAKAGE AT THE LAG SCREW HOLE WAS CONFIRMED.

    Patient problems: Discomfort; Failure of Implant; Injury

    MAUDE report 9586087 ↗
  • InjuryJul 12, 2019Break; Fracture; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT'S LEFT FEMUR WAS REVISED DUE TO FEMORAL PERIPROSTHETIC FRACTURE, AND BREAKAGE OF THE NAIL AT THE FRACTURE SITE. PATIENT WAS REVISED TO COMPETITOR DEVICES.

    Patient problems: No Information

    MAUDE report 8784562 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices