GAMMA 1320-0330Recalls & reported problems
Stryker GmbH · U-Blade Lag Screw Connector · Class 2
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540377869), received between Aug 19, 2015 and Nov 16, 2020 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break4
- Fracture1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2020Fracture; Fracture
AS REPORTED: "DURING THE GAMMA U LAG OPERATION, WHEN THE SURGEON TRIED TO INSERT THE U BLADE WITH THE INSERTER ACCORDING TO THE NORMAL PROCEDURE, 13200330 (U LUG SCREW CONNECTOR) WAS DAMAGED AND BROKE. IT WAS SUCCESSFULLY REMOVED WITH THE PLIERS AT THE HOSPITAL. DAMAGE DUE TO METAL FATIGUE WAS SUSPECTED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10843650 ↗MalfunctionMay 17, 2019Break
IT WAS REPORTED THAT, DURING A REVISION SURGERY, SURGEON IMPLANTED A SHORT GAMMA NAIL, RC LAG SCREW, AND DISTAL LOCKING SCREW. DURING INSERTION OF THE RC LAG SCREW OP TECH PROTOCOL WAS FOLLOWED. THE U BLADE LAG SCREW CONNECTOR BROKE AFTER THE U BLADE INSERTER WAS PLACED OVER IT AND THE HANDLE WAS PRESSED 3-4 TIMES. PATIENT WAS NOT HARMED AS A RESULT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8621427 ↗MalfunctionNov 19, 2018Break
INSERT GAMMA NAIL AND INSERT U LAG SCREW. CONNECTING U LAG SCREW CONNECTOR TO THE END OF THE LAG SCREW, INSERTING THE U BLADE AND PUSHING IT IN 3 TIMES WITH THE U INSERTER, THE U LAG SCREW CONNECTOR BREAKS AT THE MIDDLE GROOVE. INSERTION OF THE U BLADE GAVE UP AND CONTINUED WITH THE LAG SCREW INSERTED, AND IT ENDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8084096 ↗MalfunctionOct 1, 2018Break
DURING GAMMA3 SURGERY, U-BLADE LAG SCREW CONNECTOR WAS BROKEN IN TWO AT BLACK MARKER PORTION WHEN INSERTING A U-BLADE USING U-BLADE INSERTER. THE BROKEN CONNECTOR WAS COMPLETELY REMOVED FROM THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7923905 ↗MalfunctionAug 19, 2015Break
IT WAS REPORTED THAT DURING G3 NAIL SURGERY, THE SURGEON TRIED TO INSERT THE U-BLADE. HOWEVER, THE U-BLADE INSERTION WAS DIFFICULT. WHEN THE SURGEON TRIED TO INSERT U-BLADE FORCIBLY USING AN INSERTER, CONNECTOR BROKE. THEREFORE THE SURGEON DID NOT INSERTED U-BLADE. AND THE SURGERY WAS FINISHED. ADDITIONAL INFORMATION: THERE IS A POSSIBILITY THAT AN END OF LAG SCREW WAS INSERTED MORE DEEPLY THAN A LATERAL CORTICAL BONE. THEREFORE THE SURGEON COULD NOT INSERT U-BLADE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5016144 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page