T2 1826-0003SRecalls & reported problems
Stryker GmbH · End Cap, SCN · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540378613), received between Oct 1, 2015 and Jun 7, 2021 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Fracture1
- Loss of or Failure to Bond1
- Migration1
Patient problems
- Failure of Implant1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 7, 2021Migration; Migration
IT WAS REPORTED: "DOCTOR SENT AN X-RAY SHOWING THAT THE T2 SCN END CAP WAS OUT OF THE NAIL AND IN THE PATIENT'S JOINT. THE REPORTER IS UNAWARE IF A REVISION IS SCHEDULED TO HAPPEN".
Patient problems: Failure of Implant
MAUDE report 11949734 ↗MalfunctionJan 21, 2021Fracture; Fracture
AS REPORTED; "ABOUT 10 YEARS AGO, THE PATIENT INSERTED A T2 SUPRA CONDILLA NAIL AT ANOTHER HOSPITAL, AND OVER THE YEARS, THE END CAP HAD FALLEN OFF AND WAS IN THE JOINT. ADDITIONAL INFORMATION RECEIVED: DID THE LOOSE CAP HAD BEEN LOOSE IN THE JOINT DURING THE LAST 10 YEARS OR MORE RECENTLY? IT LOOSEN RECENT YEARS. ANY PLANS FOR MEDICAL INTERVENTION TO REMOVE THE END CAP? THE INFORMATION IS UNAVAILABLE, IF THE END CAP IS GOING TO REMOVE OR NOT.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 11207338 ↗InjuryNov 15, 2018Detachment of Device or Device Component
AT (B)(6) 2018, OPERATION WITH T2SCN WAS DONE. AFTER THE OPERATION, THE END CAP WAS CAME OUT AND THE MOST DISTAL CONDYLE SCREW WAS BACKED OUT. THEREFORE, AT (B)(6) 2018, THE END CAP AND CONDYLE SCREW WAS REMOVED AND NEW SAME DEVICES WERE IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8075190 ↗InjuryOct 1, 2015Loss of or Failure to Bond
IT WAS REPORTED THAT IMPLANTS WERE EXPLANTED FROM PATIENT'S LEFT FEMUR DUE TO INFECTION AND NON-UNION. THE SURGEON DID NOT FEEL THE INFECTION WAS CAUSED BY THE IMPLANT HOWEVER DUE TO THE INFECTION IT WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 5116381 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page