Skip to content

T2 1897-6095SRecalls & reported problems

Stryker GmbH · Lag Screw, Recon · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540384799), received between Aug 14, 2018 and Mar 23, 2026 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Break1
  • Fracture1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Migration1

Patient problems

  • No Code Available2
  • Perforation2
  • Foreign Body In Patient1
  • Hip Fracture1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 23, 2026Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED: "THE LAG SCREW OF THE T2ALPHAGT WAS NOT SEATED WITHIN THE SCREW HOLE, RESULTING IN FEMORAL SHORTENING WHEN THE PATIENT APPLIED WEIGHT POSTOPERATIVELY, CAUSING THE PROXIMAL END OF THE NAIL TO PROTRUDE FROM THE FEMUR."

    Patient problems: Perforation

    MAUDE report 24670642 ↗
  • InjurySep 8, 2023Migration; Migration

    AS REPORTED: "T2ALPHAGT WAS USED IN THE RECOM MODE ON (B)(6) 2023 FOR A SUBTROCHANTERIC FRACTURE OF THE FEMUR . AFTER SURGERY, DR CONFIRMED CUTOUT BY BOTH LAG SCREWS AT FEMORAL HEAD WHEN POSTOPERATIVE FOLLOW-UP. REMOVAL SURGERY WAS DONE AT (B)(6) 2023. THE DR FELT THAT THE SET SCREWS SEEMED TO BE LOOSELY FASTENED TO THE LAG SCREWS DURING REMOVAL SURGERY."

    Patient problems: Hip Fracture; Perforation

    MAUDE report 17714474 ↗
  • MalfunctionDec 19, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "DURING SURGERY THE SURGEON TOOK AN X RAY AND IT SHOWED THE SCREW HAD BROKE. SURGEON USED ONE LAG SCREW IN T2 ALPHA RECON MODE, THEN LEFT THE HEAD OF THE BROKEN SCREW IN THE NAIL THREADS AS WAS UNABLE TO REMOVE AND FINISHED THE CASE BY USING ONE OF THE ANTEGRADE MODE SCREW HOLE OPTIONS.". THERE WAS A 15 MINUTE DELAY TO THE SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 16010174 ↗
  • InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available

    THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿T2 RECON IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE T2 RECON NAILING SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 2009 AND 2017. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 13 EVENTS WERE INITIATED RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO SEQUEL AFTER HIP FRACTURE.

    Patient problems: No Code Available

    MAUDE report 7780076 ↗
  • InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available

    THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿T2 RECON IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE T2 RECON NAILING SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 2009 AND 2017. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 13 EVENTS WERE INITIATED RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. A RE-OSTEOSYNTHESIS HAS BEEN PERFORMED.

    Patient problems: No Code Available

    MAUDE report 7779990 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page