GAMMA 1320-3010Recalls & reported problems
Stryker GmbH · One Shot Target Device · Class 2
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540396860), received between Sep 23, 2015 and Jun 8, 2022 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Align” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Align2
- Material Deformation2
- Fracture1
- Malposition of Device1
- Positioning Failure1
Patient problems
- No Known Impact Or Consequence To Patient4
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2022Positioning Failure; Positioning Failure
AS REPORTED: "DURING THE GAMMA 3 SURGERY, WHEN THE ONE-SHOT GUIDE WAS USED, THE IMAGE WAS CLEARLY MISALIGNED AND COULD NOT BE ACCURATELY TARGETED. WHEN OTHER INSTRUMENT WAS USED AS A SUBSTITUTE, IT FUNCTIONED AS NORMAL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14630778 ↗InjuryJul 20, 2020Fracture; Fracture
THE CUSTOMER REPORTED THAT : "DURING THE INSTALLATION OF A NAIL GAMMA3 BY [SURGEON], A RUPTURED INSTRUMENT SET OCCURRED DURING THE LOCKING OF THE CEPHALIC SCREW; RESULTING IN THE LOSS OF A PART OF THE DEVICE, WHICH CAME IN THE SOFT PARTS OF THE PATIENT'S HIP." THE CUSTOMER REPLIED: "THE EVENT OCCURRED DURING THE INTERVENTION, WHEN THE SCREW WAS LOCKED. THE PROCEDURE MAY HAVE BEEN COMPLETED, BUT ADDITIONAL MANIPULATION WAS REQUIRED WHEN THE UNSUCCESSFUL ATTEMPT WAS MADE TO RETRIEVE THE PIECE OF THE DEVICE THAT WAS LOST FROM THE PATIENT. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT AT THIS TIME."
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 10299946 ↗MalfunctionOct 21, 2019Failure to Align
IT WAS REPORTED THAT THE INSERTION POSITION INDICATED BY THE ONE SHOT GUIDE AND THE POSITION OF THE ACTUALLY INSERTED GUIDE PIN WERE MISALIGNED DURING GAMMA3 OPERATION. THEREFORE, THE ONE SHOT GUIDE WAS NOT USED AND THE OPERATION WAS CONTINUED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES OR SURGICAL DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9216232 ↗MalfunctionAug 22, 2019Failure to Align; Material Deformation; Material Deformation
IT WAS REPORTED THAT WHEN THE ONE SHOT DEVICE WAS USED IT WAS NOTICED THAT THE LINES ONE X-RAY INDICATING LAG SCREW POSITION DID NOT ALIGN WITH THE HOLE IN THE NAIL. LOOKING ALONG THE LENGTH OF THE ONE SHOT DEVICE IT WAS NOTED THAT THERE WAS A TWIST ALONG ITS LENGTH. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY OR ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8922764 ↗MalfunctionNov 5, 2018Material Deformation
THE ONE SHOT TARGET DEVICE IS BENT, CAUSING AN INACCURATE DRILLING FOR FEMORAL NECK SCREWS. THERE IS A PLASTIC DEFORMATION OF THE DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8039731 ↗MalfunctionSep 23, 2015Malposition of Device
DURING NAIL SURGERY, WHEN THE SURGEON USED THE ONE SHOT GUIDE, THE INDICATOR (WIRE) WASN'T SEEN AT THE NORMAL LOCATION. THEREFORE THE SURGEON DID NOT USE THE ONE SHOT GUIDE AND THE SURGERY WAS COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5096896 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page