NA 4100700000Recalls & reported problems
Stryker Corporation · High Speed Bur Attachment · Class 1
Quick summary
This Stryker high speed bur attachment is a surgical instrument used in orthopedic procedures. The 74 reports mostly describe the bur coming loose or detaching from its handpiece, along with some overheating and breakage. No deaths and no injuries were reported, and most events were caught before use or outside a patient. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 74
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 04546540401717), received between Jul 24, 2018 and Apr 22, 2026 — classified as 74 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component55
- Overheating of Device8
- Fracture4
- Ejection Problem3
- Unintended Ejection2
- Break1
- Material Disintegration1
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- No Patient Involvement16
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: OVERHEATING OF DEVICE. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954005 ↗MalfunctionOct 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368682 ↗MalfunctionJul 29, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22648121 ↗MalfunctionJul 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22647007 ↗MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207525 ↗MalfunctionOct 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20494423 ↗MalfunctionJul 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19823274 ↗MalfunctionApr 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113365 ↗MalfunctionJan 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18557345 ↗MalfunctionOct 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17993226 ↗MalfunctionJul 27, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17402742 ↗MalfunctionJul 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401902 ↗MalfunctionMay 19, 2023Fracture; Fracture
THE DEVICE HAD A BROKEN BUR INSIDE DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16961290 ↗MalfunctionApr 25, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16808051 ↗MalfunctionApr 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807819 ↗MalfunctionApr 25, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16806619 ↗MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261998 ↗MalfunctionSep 2, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE DISASSEMBLED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15346619 ↗MalfunctionAug 25, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE HAD A COMPONENT MISSING DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15293306 ↗MalfunctionJul 22, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING A COMPONENT WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15085333 ↗MalfunctionJul 19, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE WAS DISASSEMBLED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15059316 ↗MalfunctionJul 8, 2022Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14961393 ↗MalfunctionJun 22, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE DISASSEMBLED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14776538 ↗MalfunctionMay 25, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS MISSING A SPRING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14497888 ↗MalfunctionMay 6, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE WAS DISASSEMBLED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14305715 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.