GAMMA 3425-3340SRecalls & reported problems
Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540406651), received between Jun 19, 2020 and Jun 9, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Fracture1
- Improper or Incorrect Procedure or Method1
Patient problems
- Bone Fracture(s)1
- Failure of Implant1
- Iatrogenic Source1
- Implant Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2023Break; Fracture; Break; Fracture
AS REPORTED: "INCREASING PAIN OF THE PATIENT IN THE HIP WITHOUT TRAUMA. X-RAY SHOWED NAIL FRACTURE. THE DEVICE WAS REVISED".
Patient problems: Failure of Implant; Implant Pain
MAUDE report 17097517 ↗InjuryJun 19, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
AS REPORTED: (B)(6) HOSPITAL WAS DOING A GAMMA NAIL AND INSERTED A RIGHT NAIL INTO A LEFT FEMUR. DUE TO THE BOW IN THE NAIL, THE NAIL FRACTURED THE DISTAL ANTERIOR CORTEX OF THE PT'S FEMUR BUSTING OUT THROUGH THE BONE. NOT KNOWING WHAT CAUSED THIS TO OCCUR, THE SURGEON OPENED A LONGER NAIL TO TRY AND BY-PASS THIS NEWLY CREATED FRACTURE BUT THIS TOO WAS A RIGHT (WRONG SIDED) NAIL AND THE SAME ISSUE OCCURRED WITH THE NAIL GOING THROUGH THE SAME DISTAL ANTERIOR FRACTURE. IT WAS AT THIS TIME THEY HAD REALISED THE WRONG SIDED NAILS WERE USED BOTH TIMES. IT WAS NOT LONG AFTER THIS I ARRIVED AT THE HOSPITAL. SURGICAL DELAY OF APPROX 60 MINUTES.
Patient problems: Bone Fracture(s); Iatrogenic Source
MAUDE report 10176359 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.