The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540437143), received between May 4, 2021 and Dec 23, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Biocompatibility1
  • Device Damaged Prior to Receipt by User1
  • Difficult to Remove1
  • Structural Problem1

Patient problems

  • Insufficient Information3
  • Failure of Implant1
  • Hypersensitivity/Allergic reaction1
  • Limb Fracture1
  • Loss of Range of Motion1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 23, 2025Structural Problem; Structural Problem

    THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE T2 PROXIMAL HUMERAL NAILING (T2 PHN) IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF T2 PHN. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025. DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 3 HAD DEVICE REVISION OR REPLACEMENT AND STRUCTURAL PROBLEM.

    Patient problems: Insufficient Information

    MAUDE report 23878373 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE T2 PROXIMAL HUMERAL NAILING (T2 PHN) IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE", THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF T2 PHN. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JANUARY 2016 THROUGH 30 MARCH 2025. DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 45 HAD LOSS OF RANGE OF MOTION.

    Patient problems: Loss of Range of Motion

    MAUDE report 23877760 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE T2 PROXIMAL HUMERAL NAILING (T2 PHN) IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF T2 PHN. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025. DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 59 HAD DEVICE REVISION OR REPLACEMENT.

    Patient problems: Insufficient Information

    MAUDE report 23877325 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE T2 PROXIMAL HUMERAL NAILING (T2 PHN) IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF T2 PHN. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025. DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 52 HAD DEVICE REVISION OR REPLACEMENT.

    Patient problems: Insufficient Information

    MAUDE report 23877320 ↗
  • InjurySep 3, 2025Biocompatibility; Biocompatibility

    CUSTOMER SERVICE WAS CONTACTED BY A PATIENT INFORMING SHE HAS BEEN SUFFERING FROM SEVERE SKIN RASHES AND PAIN AFTER AN IMPLANT EMERGENCY SURGERY IN THE UPPER ARM AND REQUESTS SAMPLE OF THE COMPOSITION (METALS ETC) OF THE IMPLANTS AND SCREWS AS WELL AS A SAMPLE OF THE IMPLANTS SO THAT THE LAB/DERMATOLOGIST CAN DETERMINE WHETHER THE PARTS NEED TO BE REMOVED OR WHETHER THERE IS ANOTHER SOLUTION.

    Patient problems: Hypersensitivity/Allergic reaction; Pain

    MAUDE report 22950339 ↗
  • MalfunctionMar 12, 2024Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User

    AS REPORTED: "ORDERING ERROR BY SALES REP. A PROXIMAL HUMERAL LONG NAIL WAS INCORRECTLY ORDERED AND SHIPPED FOR A HUMERAL NAIL CASE. IT WAS USED WITHOUT BEING NOTICED AT THE SURGERY." NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18887396 ↗
  • InjuryMay 4, 2021Difficult to Remove; Difficult to Remove

    THE SURGEON REPORTED THAT NONE OF THE INSTRUMENTS ON THE METAL EXTRACTION FIT THE PROXIMAL HUMERAL NAIL. HE REPORTED THAT HE HAD ALREADY REMOVED THE TRANSVERSE SCREWS, BUT NEITHER THE EXTRACTION ROD NOR THE 6MM CONICAL ADAPTER WOULD GRAB THE PROXIMAL END OF THE NAIL. AS A RESULT, THE PROXIMAL HUMERUS WAS SPLIT TO LOOSEN THE NAIL. THERE WAS A 75 MINUTE SURGICAL DELAY AS A RESULT OF THE EVENT.

    Patient problems: Failure of Implant; Limb Fracture

    MAUDE report 11773329 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →