Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540494160), received between Mar 9, 2016 and Nov 2, 2016 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Remove1
  • Material Deformation1
  • Material Integrity Problem1

Patient problems

  • No Code Available2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 2, 2016Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT; THE PATIENT UNDERWENT REMOVING SURGERY ON (B)(6)2016. THE SURGEON REMOVED THE SCREW AND THE OFFSET CONNECTER. AFTER SURGERY, THE CONNECTER RING WAS REMAINING PATIENT BODY WHEN THE SURGEON CHECKED THE IMAGE. THEREFORE THE SURGEON IS MONITORING FOR THE PATIENT. (UPDATE (B)(6) 2016: THE REASON FOR THE REVISION SURGERY WAS REMOVAL OF THE IMPLANTS BECAUSE THE PATIENT FUSED.

    Patient problems: No Code Available

    MAUDE report 6072224 ↗
  • InjuryMar 9, 2016Material Deformation; Material Integrity Problem; Difficult to Remove; Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT; DURING SURGERY, THE SURGEON USED THE SCREW AND CONNECTOR FOR TH12 VERTEBRAL BODY FRACTURE. THE SURGEON REMOVED THE CONNECTOR ADAPTER FROM OFFSET CONNECTOR OF TH12 (LEFT) AND L1. TO REMOVE THE CONNECTOR ADAPTER FROM OFFSET CONNECTOR OF TH12 (RIGHT), THE SURGEON TRIED TO LOOSEN NUT, BUT IT WAS NOT POSSIBLE TO LOOSEN NUT. THEREFORE, THE SURGEON TRIED TO LOOSEN THE OFFSET CONNECTOR OF L1 (LEFT) AND REMOVE THE ROD. HOWEVER, IT WAS NOT POSSIBLE TO LOOSEN THE SET SCREW OF OFFSET CONNECTOR. THE SURGEON USED THE ROD CUTTER. AND HE USED BRAND-NEW SCREW AND THE OFFSET CONNECTOR. OPERATION TIME WAS EXCEEDED FOR ABOUT 3 HOURS.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 5490417 ↗