Skip to content

XIA 3 48232480Recalls & reported problems

Stryker Corporation · CP TI ROD · Class 2 · Implant

Quick summary

This Stryker XIA 3 rod is a titanium spinal implant used with pedicle screw systems to hold the spine steady. Most of the 31 reports describe the rod breaking or fracturing, sometimes years after it was implanted, often after a fall or with no obvious trigger, and several times more than one rod in the same patient. Patients are commonly treated for pain and sometimes need follow-up surgery. No deaths were reported and there were 20 injuries.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 04546540560575), received between Feb 9, 2016 and Jul 29, 2026 — classified as 22 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Fracture” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture32
  • Break4
  • Adverse Event Without Identified Device or Use Problem1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Pain10
  • Back pain3
  • Injury3
  • Insufficient Information3
  • Inflammation2
  • Muscle Weakness/Atrophy2
  • No Code Available2

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 29, 2026Fracture

    It was reported that a patient sustained a fall at home approximately 15 months post-operatively and experienced thoracic back pain. Imaging revealed that two Xia 3 Rods had fractured. The patient is scheduled to receive specialized medical care, however hospitalization nor surgical treatment have been required. This report captures the second of two Rods.

    Patient problems: Back pain

    MAUDE report 50074691 ↗
  • InjuryJul 29, 2026Fracture

    It was reported that a patient sustained a fall at home approximately 15 months post-operatively and experienced thoracic back pain. Imaging revealed that two Xia 3 Rods had fractured. The patient is scheduled to receive specialized medical care, however hospitalization nor surgical treatment have been required. This report captures the first of two Rods.

    Patient problems: Back pain

    MAUDE report 50074679 ↗
  • InjuryJul 21, 2026Fracture

    A HEALTH PROFESSIONAL REPORTED THAT A XIA 3 ROD WAS OBSERVED TO HAVE FRACTURED APPROXIMATELY 3 YEARS AND 8 MONTHS POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE EVENT RESULTED IN HOSPITALIZATION OR ITS EXTENSION. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED AND THE PATIENT REPORTED NOT EXPERIENCING ANY PAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25898679 ↗
  • InjuryApr 14, 2026Fracture

    ADDITIONAL IMAGING PROVIDED FROM THE FIELD INDICATED THAT THREE RODS FRACTURED. THIS REPORT CAPTURES THE FIRST OF THREE RODS.

    Patient problems: Back pain

    MAUDE report 24877573 ↗
  • InjuryDec 23, 2025Fracture; Fracture

    A HEALTH PROFESSIONAL REPORTED THAT AN UNKNOWN ROD FRACTURED INTRA-OPERATIVELY AND THAT THE PATIENT EXPERIENCED UNSPECIFIED ADVERSE CONSEQUENCES.

    Patient problems: Insufficient Information

    MAUDE report 23882515 ↗
  • MalfunctionMay 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT TWO XIA 3 TITANIUM RODS FRACTURED POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED AN UNSPECIFIED "ISSUE RELATED TO BONE DESTRUCTION". NO REVISION SURGERY HAS BEEN REPORTED. THIS REPORT CAPTURES THE FIRST OF TWO RODS.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 22069983 ↗
  • MalfunctionMay 21, 2025Fracture; Fracture

    NO NEW INFORMATION.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 22067266 ↗
  • InjuryDec 27, 2024Fracture; Fracture

    (B)(6) REPORTED THAT A PATIENT WITH PAIN IS TO BE REVISED APPROXIMATELY 3 YEARS POST-OPERATIVELY TO ADDRESS THREE FRACTURED RODS. THIS REPORT CAPTURES THE XIA 3 ROD.

    Patient problems: Pain

    MAUDE report 21025598 ↗
  • MalfunctionOct 25, 2024Fracture; Fracture

    IT WAS REPORTED THAT TWO XIA 3 RODS FRACTURED APPROXIMATELY ONE YEAR POST-OPERATIVELY. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. THE PATIENT WAS REPORTED TO HAVE NO CHANGES IN THEIR CONDITION AND NO ADVERSE CONSEQUENCES WERE REPORTED. THIS REPORT CAPTURES THE FIRST OF TWO XIA 3 RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20533351 ↗
  • MalfunctionOct 25, 2024Fracture; Fracture

    IT WAS REPORTED THAT TWO XIA 3 RODS FRACTURED APPROXIMATELY ONE YEAR POST-OPERATIVELY. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. THE PATIENT WAS REPORTED TO HAVE NO CHANGES IN THEIR CONDITION AND NO ADVERSE CONSEQUENCES WERE REPORTED. THIS REPORT CAPTURES THE SECOND OF TWO XIA 3 RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20533180 ↗
  • InjuryOct 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT TWO XIA RODS FRACTURED APPROXIMATELY FOUR YEARS POST-OPERATIVELY. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. THE PATIENT WAS REPORTED TO HAVE EXPERIENCED AN UNKNOWN SERIOUS ADVERSE REACTION. THIS REPORT CAPTURES THE FIRST OF TWO XIA RODS.

    Patient problems: Insufficient Information

    MAUDE report 20428397 ↗
  • InjuryOct 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT TWO XIA RODS FRACTURED APPROXIMATELY FOUR YEARS POST-OPERATIVELY. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. THE PATIENT WAS REPORTED TO HAVE EXPERIENCED AN UNKNOWN SERIOUS ADVERSE REACTION. THIS REPORT CAPTURES THE SECOND OF TWO XIA RODS.

    Patient problems: Insufficient Information

    MAUDE report 20428344 ↗
  • InjuryJul 31, 2024Fracture; Fracture

    IT WAS REPORTED VIA THE RUSSIAN HEALTH AUTHORITY THAT A PATIENT IS TO BE REVISED TO ADDRESS A XIA ROD FRACTURE WHICH WAS OBSERVED VIA IMAGING AT APPROXIMATELY 4.5 YEARS AFTER IMPLANTATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 19878177 ↗
  • MalfunctionMay 8, 2024Fracture; Fracture

    IT WAS REPORTED VIA THE RUSSIAN HEALTH AUTHORITY THAT A PATIENT EXPERIENCED PAIN APPROXIMATELY 26 MONTHS AFTER IMPLANTATION FOLLOWING A PHYSICAL ALTERCATION AND "HEAVY PHYSICAL LABOR." IMAGING REVEALED THE FRACTURE OF TWO XIA RODS. REVISION SURGERY HAS NOT BEEN PERFORMED. THIS REPORT CAPTURES THE FIRST OF TWO RODS.

    Patient problems: Pain

    MAUDE report 19274357 ↗
  • MalfunctionMay 8, 2024Fracture; Fracture

    IT WAS REPORTED VIA THE RUSSIAN HEALTH AUTHORITY THAT A PATIENT EXPERIENCED PAIN APPROXIMATELY 26 MONTHS AFTER IMPLANTATION FOLLOWING A PHYSICAL ALTERCATION AND "HEAVY PHYSICAL LABOR." IMAGING REVEALED THE FRACTURE OF TWO XIA RODS. REVISION SURGERY HAS NOT BEEN PERFORMED. THIS REPORT CAPTURES THE SECOND OF TWO RODS.

    Patient problems: Pain

    MAUDE report 19274356 ↗
  • InjuryApr 2, 2024Fracture; Fracture

    IT WAS REPORTED VIA THE RUSSIAN HEALTH AUTHORITY THAT A PATIENT WAS HOSPITALIZED DUE TO CONSTRUCT INSTABILITY APPROXIMATELY 25 MONTHS AFTER IMPLANTATION AND IMAGING REVEALED THE FRACTURE OF TWO XIA RODS AT THE BASE OF THE CONSTRUCT. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE FIRST OF TWO RODS.

    Patient problems: Inflammation

    MAUDE report 19029702 ↗
  • InjuryApr 2, 2024Fracture; Fracture

    IT WAS REPORTED VIA THE RUSSIAN HEALTH AUTHORITY THAT A PATIENT WAS HOSPITALIZED DUE TO CONSTRUCT INSTABILITY APPROXIMATELY 25 MONTHS AFTER IMPLANTATION AND IMAGING REVEALED THE FRACTURE OF TWO XIA RODS AT THE BASE OF THE CONSTRUCT. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE SECOND OF TWO RODS.

    Patient problems: Inflammation

    MAUDE report 19029701 ↗
  • MalfunctionFeb 23, 2024Fracture; Fracture

    THE RUSSIAN HEALTH AUTHORITY REPORTED THAT A PATIENT REPORTED PAIN APPROXIMATELY THREE TO FOUR MONTHS POST-OPERATIVELY AND IMAGING REVEALED MIGRATION OF THE LEFT L4 SCREW AND FRACTURE OF TWO XIA RODS. IT IS UNKNOWN IF A REVISION SURGERY IS PLANNED. THIS REPORT CAPTURES THE SECOND OF THE TWO FRACTURED RODS.

    Patient problems: Pain

    MAUDE report 18773084 ↗
  • InjuryJan 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WILL BE REVISED TO ADDRESS ONE FRACTURED ES2 STRAIGHT ROD AND TWO FRACTURED XIA TITANIUM RODS. THE PATIENT HAS REPORTED NUMBNESS AND WEAKNESS IN THEIR LOWER EXTREMITIES. THIS REPORT CAPTURES THE FIRST OF TWO XIA TITANIUM RODS.

    Patient problems: Muscle Weakness/Atrophy; Numbness

    MAUDE report 18540156 ↗
  • InjuryJan 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WILL BE REVISED TO ADDRESS ONE FRACTURED ES2 STRAIGHT ROD AND TWO FRACTURED XIA TITANIUM RODS. THE PATIENT HAS REPORTED NUMBNESS AND WEAKNESS IN THEIR LOWER EXTREMITIES. THIS REPORT CAPTURES THE SECOND OF TWO XIA TITANIUM RODS.

    Patient problems: Muscle Weakness/Atrophy; Numbness

    MAUDE report 18540155 ↗
  • InjuryJun 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR FRACTURED XIA 3 TITANIUM RODS AT THE T11 TO L1 LEVELS. THIS REPORT CAPTURES THE SECOND OF FOUR RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17078472 ↗
  • InjuryJun 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR FRACTURED XIA 3 TITANIUM RODS AT THE T11 TO L1 LEVELS. THIS REPORT CAPTURES THE THIRD OF FOUR RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17078369 ↗
  • InjuryJun 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR FRACTURED XIA 3 TITANIUM RODS AT THE T11 TO L1 LEVELS. THIS REPORT CAPTURES THE FOURTH OF FOUR RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17078252 ↗
  • InjuryJun 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS FOUR FRACTURED XIA 3 TITANIUM RODS AT THE T11 TO L1 LEVELS. THIS REPORT CAPTURES THE FIRST OF FOUR RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17078249 ↗
  • InjuryDec 16, 2021Fracture; Fracture

    IT WAS REPORTED THAT APPROXIMATELY 2.5 YEARS POST-OPERATIVELY, AFTER HEARING A "POP" AND EXPERIENCING BACK PAIN, THE PATIENT PRESENTED WITH TWO FRACTURED XIA 3 RODS. THE PATIENT WAS REVISED AND THEIR CONSTRUCT WAS EXTENDED. THIS RECORD CAPTURES THE SECOND OF THE TWO RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 13022344 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.