XIA 3 482317555Recalls & reported problems
Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant
All models of XIA 3 (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540562371), received between Oct 25, 2016 and Aug 11, 2025 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Dislodged or Dislocated3
- Detachment of Device or Device Component2
- Fracture2
- Material Integrity Problem2
- Mechanical Problem1
Patient problems
- Injury2
- Pain2
- Device Embedded In Tissue or Plaque1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2025Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
InjuryAug 11, 2025Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
MalfunctionFeb 28, 2017Break; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem
IT WAS REPORTED THAT; IMPLANTATION OF SCREW (B)(4) IN VERTEBRAL BODY S1, DURING REVISION SURGERY (B)(6) 2016 WITH ENLARGEMENT SPONDYLODESES CRANIAL, INSERTING NEW SIDE RAIL TO VERTEBRAL BODY S1, SCREWS REMAINED IN VERTEBRAL BODY. ACTUAL BROKEN SCREWS WERE DETECTED BOTH-SIDED IN S1. CLINICAL ROOT CAUSE MOST LIKELY TO PSEUDOARTHROSES L5/S1 DUE TO NOT HEALED CAGE IN SPINAL DISK SHELF L5/S1. CONSPICUOUS IS THE BREAK OF SCREWS BOTH-SIDED. BOTH SCREWS ARE FROM THE SAME LOT, THE HEADS OF SCREWS WERE REMOVED FOR EXHIBIT, THE BROKEN SCREW RESTS REMAIN IN VERTEBRAL BODY; NEW SCREWS WERE PLACED. HANDLING SCREWS EVERY TIME CORRESPONDING TO MANUFACTURERS´ INSTRUCTIONS.
Patient problems: Device Embedded In Tissue or Plaque; Injury
MAUDE report 6367302 ↗InjuryFeb 28, 2017Break; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem
IT WAS REPORTED THAT; IMPLANTATION OF SCREW (B)(6) 2016 IN VERTEBRAL BODY S1, DURING REVISION SURGERY (B)(6) 2016 WITH ENLARGEMENT SPONDYLOSDESE CRANIAL, INSERTING NEW SIDE RAIL TO VERTEBRAL BODY S1, SCREWS REMAINED IN VERTEBRAL BODY. ACTUAL BROKEN SCREWS WERE DETECTED BOTH-SIDED IN S1. CLINICAL ROOT CAUSE MOST LIKELY TO PSEUDOARTHROSE L5/S1 DUE TO NOT HEALED CAGE IN SPINAL DISK SHELF L5/S1. CONSPICUOUS IS THE BREAK OF SCREWS BOTH-SIDED. BOTH SCREWS ARE FROM THE SAME LOT, THE HEADS OF SCREWS WERE REMOVED FOR EXHIBIT, THE BROKEN SCREW RESTS REMAIN IN VERTEBRAL BODY; NEW SCREWS WERE PLACED. HANDLING SCREWS EVERY TIME CORRESPONDING TO MANUFACTURERS´ INSTRUCTIONS.
Patient problems: Injury; No Code Available
MAUDE report 6367301 ↗MalfunctionOct 25, 2016Mechanical Problem; Break; Mechanical Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT A SCREW POTENTIALLY DAMAGED INTEROPERATIVELY, PREVENTING USE. THE HEAD OF THE SCREW WAS DISCOVERED BY SURGEON DISLODGED FROM THE SHAFT. SCREW WAS REMOVED AND REPLACED WITH SIMILAR SCREW FROM SAME SYSTEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6055206 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.