Stryker GmbH · U-Blade Set, Ti · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04546540595997), received between Nov 15, 2018 and Oct 16, 2025 — classified as 9 injuries.

The most frequently reported specific device problem is “Migration” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Migration5
  • Appropriate Device Problem Term/Code Not Available1
  • Material Deformation1

Patient problems

  • Perforation3
  • Failure of Implant2
  • Fall2
  • Insufficient Information2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1
  • Hip Fracture1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "OVER-TELESCOPING OF THE LAG SCREW DUE TO NON-UNION. THE SCREW DOES NOT HAVE A CUTOUT. THE LAG SCREW PROTRUDED TO LATERAL DUE TO COLLAPSE OF THE FEMORAL NECK. THE PATIENT REQUIRED A REVISION SURGERY."

    Patient problems: Perforation

    MAUDE report 23318451 ↗
  • InjuryJun 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: PRIMARY G3 SURGERY WAS PERFORMED ON (B)(6) 2015. THE PATIENT FELL AND RE-FRACTURED THE BONE AT UNDER THE U-LAG SCREW OF G3. REVISION SURGERY FOR REMOVING G3 AND REPLACING LONG NAIL WAS PERFORMED ON (B)(6) 2025. DURING SURGERY, THE TIP OF THE T-HANDLE BECAME BENT DUE TO THE HARDNESS OF THE BONE DURING REMOVAL. THIS RECORD COVERS THE REVISION SURGERY.".

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 22287630 ↗
  • InjuryApr 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "PRIMARY SURGERY WITH GAMMA3 NAIL AND U-LAG SCREW WAS PERFORMED ON (B)(6) 2022. ON (B)(6) 2025, WHEN THE PATIENT VISITED THE HOSPITAL WITH AN IRREGULAR HEARTBEAT, LAG SCREW CUT-OUT WAS OBSERVED. ON (B)(6) 2025, THE REVISION SURGERY WAS PERFORMED FOR REMOVING U-LAG SCREW."

    Patient problems: Perforation

    MAUDE report 21762669 ↗
  • InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED FROM A TROCHANTERIC NAIL GAMMA4 NAIL TO A TOTAL HIP DUE TO THE ROTATION OF LAG SCREW NO T BEING MAINTAINED.

    Patient problems: Insufficient Information

    MAUDE report 20092081 ↗
  • InjuryMay 7, 2022Migration; Migration

    AS REPORTED: "GAMMA3 PRIMARY SURGERY WAS PERFORMED ON FEBRUARY 18, 2022 WITHOUT ANY PROBLEM. APPROXIMATELY TWO MONTHS POSTOPERATIVELY, THE U-LUG SCREW PENETRATED THE FEMORAL HEAD. REVISION SURGERY WITH BIPOLAR WAS PERFORMED ON (B)(6) 2022."

    Patient problems: Insufficient Information

    MAUDE report 14318034 ↗
  • InjuryOct 5, 2021Material Deformation; Migration; Material Deformation; Migration

    ON (B)(6) 2021, GAMMA 3 SURGERY WAS PERFORMED ON THE RIGHT PROXIMAL FEMUR. ON (B)(6) 2021, ALTHOUGH THE DETAILS OF THE BREAKAGE ARE UNKNOWN, THE U-BLADE LAG SCREW THAT WAS FIXED TO THE GAMMA 3 NAIL WAS FOUND TO HAVE FAILED AND MOVED UPWARD TO THE HEAD OF THE BONE. THE U-LAG PORTION OF THIS LAG SCREW WAS DEFORMED SO THAT IT OPENED FORWARD. THE SURGEON PLANS TO COLLECT THE DEFECTIVE DEVICE WHEN PERFORMING THE GIRDLE STONE LATER.

    Patient problems: Failure of Implant

    MAUDE report 12579446 ↗
  • InjuryMar 4, 2021Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    AS REPORTED: "AFTER THA SURGERY (RIGHT SIDE), CUTOUT OF A LAG SCREW OCCURRED AT THE FEMORAL HEAD. AFTER THAT, THE LAG SCREW INVADED THE PELVIS. THE INITIAL SURGERY PERFORMED ON (B)(6) 2021, THE EVENT WAS CONFIRMED ON (B)(6) 2021. THE RE-OPERATION WAS PERFORMED ON (B)(6) 2021. ADDITIONAL INFORMATION RECEIVED ON 17-FEB-202: "AT THE TIME OF THE INITIAL OPERATION, THE OPERATION WAS COMPLETED WITHOUT INSERTING THE U BLADE OF THE U LAG SCREW AND THE END CAP FOR THE LAG SCREW. IN ADDITION, IT HAS BEEN CONFIRMED THAT THE LAG SCREW DID NOT TURN WHEN THE SET SCREW WAS NOT LOOSENED IN THE REMOVAL PROCESS. SO, THE SURGEON RECOGNIZED THAT THE SET SCREW WAS FIXED TO SOME EXTENT."

    Patient problems: Hip Fracture; Perforation

    MAUDE report 11417551 ↗
  • InjuryApr 4, 2019Appropriate Device Problem Term/Code Not Available; Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT GAMMA3 U-LAG SCREW SURGERY ON (B)(6) 2019. AFTER THE SURGERY, THE CUTOUT OF U-LAG SCREW AND THE BENDING OF U-BLADE WERE CONFIRMED. THEN GAMMA3 EXTRACTION AND REVISION TO THA WAS PERFORMED ON (B)(6) 2019.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8482949 ↗
  • InjuryNov 15, 2018Migration

    CUT OUT OF U LAG SCREW OCCURS. PATIENT FELT PAIN ON HIS HIP FOR SEVERAL WEEKS AND DIFFICULT TO WALK. THEN, WHEN FALLING ON 18TH, VISITED IN THE HOSPITAL AND AN X - RAY FILM WAS TAKEN, IT TURNED OUT THAT THE LAG SCREW HAD CUT OUT AND REACHED THE ACETABULAR. THERE IS NO DISLOCATION OF THE PIECES OF THE BONY SKULL, AND ONLY THE LAG SCREW HAS CUT OUT. AS A MEASURE AFTER OCCURRENCE, ON 12TH OCTOBER REPLACED 120 ° GAMMA NAIL.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Fall; Pain

    MAUDE report 8074990 ↗