OASYS 48551071Recalls & reported problems
Stryker Corporation · TRANSVERSE CONNECTOR · Class 2 · Implant
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540596086), received between Nov 15, 2016 and Mar 13, 2017 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break3
- Fracture3
- Detachment Of Device Component1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2017Break; Detachment Of Device Component; Fracture; Fracture
IT WAS REPORTED THAT; DURING TIGHTENING OF TRANSVERSE CROSS CONNECTOR BOTTOM TABS BROKE OFF. THIS HAPPENED 8-10 TIMES WITH ALL SIZES OF CROSS CONNECTORS. PROPER TECHNIQUE WAS FOLLOWED DURING TIGHTENED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6399899 ↗MalfunctionMar 13, 2017Break; Fracture; Fracture
IT WAS REPORTED THAT; DURING TIGHTENING OF TRANSVERSE CROSS CONNECTOR BOTTOM TABS BROKE OFF. THIS HAPPENED 8-10 TIMES WITH ALL SIZES OF CROSS CONNECTORS. PROPER TECHNIQUE WAS FOLLOW DURING TIGHTENED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6399637 ↗MalfunctionNov 15, 2016Fracture; Break; Fracture
IT WAS REPORTED THAT; WHEN TORQUING THE TRANSVERSE CONNECTOR SET SCREW, THE TRANSVERSE CONNECTOR BROKE. IT WAS THE PIECE THAT SNAPS ONTO THE ROD THAT BROKE. THE COUNTER TORQUE WAS USED. ALL OTHER BLOCKERS WERE FINAL TIGHTENED. THE SET SCREW WAS REMOVED FROM THE CONNECTOR, AND THE ENTIRE CONNECTOR WAS REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6104746 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.