NGENIUS 6000-014-000Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · NGENIUS UNIVERSAL TRACKER · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 04546540651563), received between Feb 7, 2017 and Mar 1, 2024 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Image Display Error/Artifact” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Image Display Error/Artifact8
  • Imprecision5
  • Loose or Intermittent Connection5
  • Incorrect Measurement4
  • Detachment of Device or Device Component1
  • Disassembly1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Patient Involvement5
  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 1, 2024Incorrect Measurement; Detachment of Device or Device Component; Incorrect Measurement; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A CASE, THE NGENIUS UNIVERSAL TRACKER WAS NOT FULLY SEATED WHICH CAUSED THE ACCURACY TO BE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18821744 ↗
  • MalfunctionNov 10, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, THE DEVICE WAS INACCURATE AND NOT TRACKING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18115143 ↗
  • MalfunctionJul 25, 2022Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER THE DEVICE WAS FOUND TO BE LOOSE WHICH COULD LEAD TO A POTENTIAL INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15093547 ↗
  • MalfunctionJun 27, 2022Imprecision; Incorrect Measurement; Imprecision; Incorrect Measurement

    PER THE CUSTOMER, THERE WAS A LOSS OF ACCURACY DURING THE IMAGE GUIDED PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14827714 ↗
  • MalfunctionDec 21, 2021Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE TRACKER WAS INACCURATE DURING A CASE.  THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13056236 ↗
  • MalfunctionSep 13, 2021Imprecision; Imprecision

    IT WAS REPORTED THAT DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY IT WAS OBSERVED THAT THE TRACKER INTERFACE IS LOOSE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12461515 ↗
  • MalfunctionDec 18, 2020Imprecision; Imprecision

    IT WAS REPORTED THAT DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY THE DEVICE WAS INACCURATE DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11039449 ↗
  • MalfunctionSep 20, 2019Image Display Error/Artifact; Loose or Intermittent Connection; Image Display Error/Artifact; Loose or Intermittent Connection

    IT WAS REPORTED THAT IT WAS OBSERVED DURING A PREVENTATIVE MAINTENANCE VISIT THAT THE BASE OF THE DEVICE IS LOOSE WHICH COULD CAUSE INACCURACIES. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9098108 ↗
  • MalfunctionSep 20, 2019Image Display Error/Artifact; Loose or Intermittent Connection; Image Display Error/Artifact; Loose or Intermittent Connection

    IT WAS REPORTED THAT IT WAS OBSERVED DURING A PREVENTATIVE MAINTENANCE VISIT THAT THE BASE OF THE DEVICE IS LOOSE WHICH COULD CAUSE INACCURACIES. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9098107 ↗
  • MalfunctionSep 20, 2019Image Display Error/Artifact; Loose or Intermittent Connection; Image Display Error/Artifact; Loose or Intermittent Connection

    IT WAS REPORTED THAT IT WAS OBSERVED DURING A PREVENTATIVE MAINTENANCE VISIT THAT THE BASE OF THE DEVICE IS LOOSE WHICH COULD CAUSE INACCURACIES. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9098104 ↗
  • MalfunctionSep 20, 2019Image Display Error/Artifact; Loose or Intermittent Connection; Image Display Error/Artifact; Loose or Intermittent Connection

    IT WAS REPORTED THAT IT WAS OBSERVED DURING A PREVENTATIVE MAINTENANCE VISIT THAT THE BASE OF THE DEVICE IS LOOSE WHICH COULD CAUSE INACCURACIES. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9098039 ↗
  • MalfunctionSep 20, 2019Image Display Error/Artifact; Loose or Intermittent Connection; Image Display Error/Artifact; Loose or Intermittent Connection

    IT WAS REPORTED THAT IT WAS OBSERVED DURING A PREVENTATIVE MAINTENANCE VISIT THAT THE BASE OF THE DEVICE IS LOOSE WHICH COULD CAUSE INACCURACIES. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9098033 ↗
  • MalfunctionDec 18, 2018Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT THE ACCURACY OF THE DEVICE WAS FOUND TO BE OFF DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8175235 ↗
  • MalfunctionNov 26, 2018Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT PRIOR TO A SURGICAL PROCEDURE AT THE USER FACILITY THE METAL PIECE THAT THE TRACKER CONNECTS TO THE INSTRUMENTATION IS LOOSE, WHICH COULD CAUSE AN INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8102697 ↗
  • MalfunctionJul 2, 2018Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE WAS INACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7654374 ↗
  • MalfunctionFeb 7, 2017Disassembly; Imprecision; Disassembly; Imprecision

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE CONDUCTED AT THE USER FACILITY THE LED INTERFACE HAD LOOSENED CAUSING THE NAVIGATION TO BE INACCURATE. NO ADVERSE CONSEQUENCES TO THE PATIENT WERE REPORTED OR PROCEDURAL DELAYS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6306679 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →