T2 1819-1015SRecalls & reported problems
Stryker GmbH · Ankle Arthrodesis Nail, right · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540673374), received between Mar 7, 2019 and Jul 15, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Failure to Advance1
- Fracture1
- Positioning Failure1
Patient problems
- Failure of Implant1
- Insufficient Information1
- No Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2025Positioning Failure; Positioning Failure
AS REPORTED: "THE POSTERIOR ANTERIOR SCREW ARRIVED NEXT TO THE NAIL. THE SURGEON WAS ABLE TO REMOVE THE SCREW AND CHOOSE TO NOT PUT IT BACK IN.".
Patient problems: Insufficient Information
MAUDE report 22509871 ↗InjuryDec 6, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT A T2 ANKLE NAIL THAT WAS IMPLANTED A FEW WEEKS PRIOR IS NOW REQUIRING A REVISION SURGERY WITHIN THE NEXT WEEK DUE TO THE NAIL FRACTURING. THE 10X150MM ANKLE NAIL USED FOR FRACTURE, WAS LOCKED WITH 1 PROXIMAL TIBIAL SCREW AND ONE DISTAL CALCANEAL SCREW. STRYKER HAD RECOMMENDED THE USE OF ADDITIONAL LOCKING SCREWS FOR EXTRA STABILITY AND THE ADDITION OF AN END CAP TO LOCK THE CONSTRUCT, HOWEVER STRYKER ADVICE WAS NOT IMPLEMENTED. THE PATIENT WILL BE REVISED WITH REMOVAL OF THE ANKLE NAIL AND FUSION OF THE ANKLE WITH PLATING.
Patient problems: Failure of Implant
MAUDE report 15923081 ↗MalfunctionMar 7, 2019Failure to Advance
CUSTOMER REPORTED THAT THE COMPRESSION SCREW WOULD NOT ADVANCE THROUGH THE NAIL.
Patient problems: No Information
MAUDE report 8401842 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page