TARGET M0035424060Recalls & reported problems

Stryker Corporation · 360 Ultra Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540675910), received between Jul 20, 2017 and Nov 4, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Premature Separation2
  • Device Dislodged or Dislocated1
  • Expulsion1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1
  • Vasoconstriction1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2025Fracture; Fracture

    THE COIL (SUBJECT DEVICE) WAS USED IN THE MEDICAL PROCEDURE. HOWEVER, WHEN THE COIL WAS PLACED IN THE ANEURSYM DURING DETACHMENT A SMALL PORTION OF THE TAIL END OF THE COIL DELIVERY WIRE BROKE OFF IN THE DETACHMENT DEVICE AND THE COIL COULD NOT BE DETACHED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23468485 ↗
  • MalfunctionJul 24, 2025Fracture; Premature Separation; Fracture; Premature Separation

    IT WAS REPORTED DURING THE PROCEDURE THE SUBJECT COIL HAD TO BE REPOSITIONED FOR PLACEMENT. WHEN THE COIL WAS RETRIEVED BACK INTO THE CATHETER, IT DETACHED AND WAS FRACTURED INTO TWO PARTS. ONE FRAGMENT WAS REMOVED FROM THE CATHETER AND ONE FRAGMENT REMAINED IN THE CATHETER. THE DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT. THE SUBJECT COIL WAS RETURNED FOR ANALYSIS AND WAS EXAMINED, AND IT WAS DISCOVERED THAT THE REPORTED EVENT OF THE SUBJECT MAIN COIL WAS DETACHED AT THE DISTAL END OF MICROCATHETER SHAFT AND WAS BROKEN INTO TWO PARTS DURING USE WAS NOT CONFIRMED. THE SUBJECT COIL DELIVERY WIRE WAS FOUND TO BE KINKED, THE MAIN COIL WAS FOUND TO BE STRETCHED, AND THE SUTURE OF SUBJECT MAIN COIL WAS FOUND TO BE DAMAGED. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586491 ↗
  • MalfunctionMar 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT WHEN INSERTING THE SUBJECT COIL INTO THE RHV (ROTATING HEMOSTASIS VALVE) OF THE MICROCATHETER DURING AN ANEURYSM COIL EMBOLIZATION PROCEDURE, THE FLUORO-SAVER MARKERS WERE VISIBLE ON THE COIL DELIVERY WIRE. HOWEVER, THE SUBJECT COIL HAD ALREADY REACHED THE ANEURYSM WHEN THE FLUOROSCOPY WAS TURNED TO VISUALIZE THE COIL POSITION. THE SUBJECT COIL WAS IMPLANTED WITHOUT ANY ISSUES, THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21590255 ↗
  • InjurySep 28, 2022Expulsion; Premature Separation; Expulsion; Premature Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE THE COIL (SUBJECT DEVICE) WAS ACCIDENTALLY DEPLOYED PARTIALLY OUTSIDE OF AN ACOM ( ANTERIOR COMMUNICATING ARTERY) ANEURYSM AND WAS SUSPENDED IN THE PARENT ACA (ANTERIOR CEREBRAL ARTERY) VESSEL. THE PHYSICIAN SUCCESSFULLY RETRIEVED THE COIL USING A RETRIEVER AND CATHETER (UNDER ASPIRATION). THE PATIENT DEVELOPED VASOSPASM IN THE RIGHT MCA (MIDDLE CEREBRAL ARTERY) WHICH CLEARED AFTER 10 MINUTES OF THE PHYSICIAN ADMINISTERING VERAPAMIL. THE PHYSICIAN SUCCESSFULLY COMPLETED THE PROCEDURE BY PLACING FIVE COILS IN THE ANEURYSM AND THE PATIENT WAS STABLE POST PROCEDURE.

    Patient problems: Vasoconstriction

    MAUDE report 15500202 ↗
  • MalfunctionMar 17, 2021Fracture; Fracture

    ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE DELIVERY WIRE WAS BROKEN DURING USE. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11509810 ↗
  • InjuryJul 20, 2017Premature Activation; Premature Activation

    DURING A COIL EMBOLIZATION PROCEDURE, DETACHMENT OF THE COIL WAS PERFORMED; HOWEVER, THE COIL DID NOT DETACH. IT WAS REPORTED THAT DURING WITHDRAWAL OF THE COIL, IT DETACHED, AND A SNARE DEVICE WAS USED TO REMOVE THE COIL. THE PROCEDURE WAS CONTINUED WITH ANOTHER COIL WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6730367 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →