Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540676306), received between Jan 17, 2022 and Feb 13, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cognitive Changes1
  • Visual Impairment1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING RIGHT INTERNAL CAROTID ARTERY (ICA) C5 ANEURYSM PROCEDURE, THE OPERATOR USED THE GUIDEWIRE TO BRING THE MICROCATHETER TO THE ANEURYSM. NEXT, THE OPERATOR FRAMED THE ANEURYSM WITH THE SUBJECT COIL AND INSERTED THE SUBJECT COIL TO GO INTO THE ANEURYSM. HOWEVER, THE OPERATOR DID NOT CONTROL THE TENSION AND THE MICROCATHETER WENT INTO THE PARENT VESSEL. THE OPERATOR TRIED TO RETRACT THE SUBJECT COIL TO DELIVER AGAIN BUT DURING THE RETRACTION INTO THE INTRODUCER SHEATH THE SUBJECT COIL FRACTURED AT THE DETACHMENT POINT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18698067 ↗
  • MalfunctionJul 21, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT COIL DETACHED PREMATURELY INSIDE THE MICROCATHETER. THE PHYSICIAN PUSHED THE SUBJECT COIL INTO THE ANEURYSM AND COMPLETED THE PROCEDURE SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15071825 ↗
  • InjuryJan 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IN A CLINICAL TRIAL PROCEDURE TO TREAT A RIGHT POSTERIOR COMMUNICATING ARTERY ANEURYSM, THE PATIENT HAD HEMIANOPIA AND DELIRIUM. ON IMAGING, OCCIPITAL MICRO EMBOLIC INFARCTS WERE SEEN WHICH RESULTED IN PROLONGED HOSPITALIZATION FOR THE PATIENT. THE EVENT WAS TREATED WITH PHYSIOTHERAPY AND ORAL OLANZAPIN 5 MG FOR DELIRIUM FOR 5 DAYS AND RECOVERED. IN THE PHYSICIAN'S OPINION, THE EVENT WAS POSSIBLY RELATED TO THE SUBJECT DEVICE AND OTHER STRYKER DEVICES USED IN THE PROCEDURE.

    Patient problems: Cognitive Changes; Visual Impairment

    MAUDE report 13272011 ↗