Skip to content

TARGET M0035469300Recalls & reported problems

Stryker Corporation · 360 Standard Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540676351), received between Oct 30, 2020 and Jan 16, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation2

Patient problems

  • Cerebral Edema1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Premature Separation

    IT WAS REPORTED THAT THE COIL (SUBJECT DEVICE) WAS USED IN THE MEDICAL PROCEDURE. HOWEVER, THE SUBJECT COIL GOT DETACHED WHEN ATTEMPTED TO DEPLOY INTO THE ANEURYSM. THE PROCEDURE WAS DELAYED BY 15 MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24090849 ↗
  • InjuryOct 30, 2020Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE TO TREAT SUBARACHNOID HEMORRHAGE, THE COIL (SUBJECT DEVICE) GOT STRETCHED AND DETACHED PREMATURELY AT THE AORTA ARTERY. THE PHYSICIAN ATTEMPTED TO REMOVE THE COIL WITH A SNARE DEVICE, BUT WAS UNSUCCESSFUL. IT WAS REPORTED THAT THE PATIENT PASSED AWAY AFTER THE PROCEDURE DUE TO BRAIN SWELLING CAUSED BY THROMBOGENIC PROGRESSION AND SPASM DEVELOPED DUE TO THE PREEXISTING SUBARACHNOID HEMORRHAGE.

    Patient problems: Cerebral Edema; Foreign Body In Patient; Thrombosis/Thrombus

    MAUDE report 10767041 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page