TARGET M0035474060Recalls & reported problems
Stryker Corporation · 360 Soft Detachable Coil · Class 2 · Implant
All models of TARGET (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540676450), received between Apr 30, 2018 and Nov 8, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED DURING THE COIL EMBOLIZATION FOR AN UNRUPTURED CEREBRAL ANEURYSM IN A-COM (THE ANTERIOR COMMUNICATING ARTERY), THE SUBJECT COIL WAS GUIDED USING WITH A STRYKER CATHETER, AND WHILE REPOSITIONING SEVERAL TIMES TO ENGAGE WITH THE VESSEL, A RESISTANCE WAS ENCOUNTERED. FLUOROSCOPIC IMAGES SHOWED THAT PROXIMAL OF THE SUBJECT COIL SEEMED TO BE STRETCHED, SO IT WAS OPERATED MORE CAREFULLY AND PERFORMED THE DETACHMENT WITH THE MARKER IN THE APPROPRIATE POSITION, BUT THE SUBJECT COIL COULD NOT BE DETACHED. IT WAS REMOVED BY THE PHYSICIAN WITH A SNARE DEVICE. THE SUBJECT COIL WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18097634 ↗MalfunctionApr 30, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
DURING THE COIL EMBOLIZATION PROCEDURE, IT WAS REPORTED THAT THE PHYSICIAN THOUGHT THE FIRST COIL (SUBJECT DEVICE) WAS SMALLER THAN THE SIZE ON THE LABEL. HOWEVER, HE STILL DETACHED THE COIL. THEN THE PHYSICIAN ADVANCED THE SECOND COIL AND BIG RESISTANCE WAS FELT IN THE MICROCATHETER. SO HE WITHDREW THE SECOND COIL, BUT THE FIRST COIL CAME OUT WITH THE SECOND COIL TOGETHER AND THE FIRST COIL WAS FOUND STRETCHED. THE PROCEDURE WAS COMPLETED WITH OTHER COILS. NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473166 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.