TARGET M0035475100Recalls & reported problems
Stryker Corporation · 360 Soft Detachable Coil · Class 2 · Implant
All models of TARGET (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540676498), received between Mar 29, 2018 and Dec 16, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Inadequate User Interface1
- Migration or Expulsion of Device1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Aneurysm1
- Intracranial Hemorrhage1
- Rupture1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 16, 2025Fracture; Fracture
IT WAS REPORTED THAT IN A CASE OF CEREBRAL ANEURYSM EMBOLIZATION FOR SUBARACHNOID HEMORRHAGE (SAH) AT POSTERIOR CEREBRAL ARTERIES (PCA) THE SUBJECT COIL WAS PLACED AS THE THIRD COIL AND ATTEMPTED TO BE DETACHED; HOWEVER, IT COULD NOT BE DETACHED. WHEN THE COIL DELIVERY WIRE WAS REMOVED FROM THE FUNNEL OF THE POWER SUPPLY, RESISTANCE WAS ENCOUNTERED. IT WAS NOTICED THAT THE ELECTRODE PART WAS PREMATURELY DETACHED AND REMAINED INSIDE OF THE POWER SUPPLY. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT. THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE SUBJECT COIL DELIVERY WIRE WAS BROKEN/FRACTURED DURING USE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23817657 ↗MalfunctionJul 2, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE PROCEDURE THE OPERATOR RETRACTED THE SUBJECT COIL FOR REPOSITIONING. WHEN THE SUBJECT COIL WAS REINSERTED, IT GOT CAUGHT BETWEEN THE INTRODUCER SHEATH AND THE MICROCATHETER'S HUB, AND PREMATURELY DETACHED IN THE MICROCATHETER'S SHAFT. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22385367 ↗MalfunctionMay 2, 2025Fracture; Fracture
IT WAS REPORTED THAT THE COIL (SUBJECT DEVICE) WAS SELECTED FOR ANTERIOR COMMUNICATING ARTERY ANEURYSM PROCEDURE. HOWEVER, WHILE ADVANCING THE COIL WITHIN THE MICROCATHETER RESISTANCE WAS ENCOUNTERED. THE PHYSICIAN ATTEMPTED TO RETRACT THE COIL AND FOUND IT TO BE STRETCHED. THUS, THE COIL ALONG WITH THE MICROCATHETER WAS WITHDRAWN AND THE COIL WAS FOUND TO BE STRETCHED AND FRACTURED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21947697 ↗MalfunctionDec 9, 2022Fracture; Fracture
THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE COIL (SUBJECT DEVICE) DELIVERY WIRE WAS FOUND TO BE BROKEN/FRACTURE DURING USE. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15947778 ↗MalfunctionMar 1, 2021Inadequate User Interface; Inadequate User Interface
IT WAS REPORTED THAT DURING THE PROCEDURE THE MIDDLE PART OF THE SUBJECT COIL BECAME INVISIBLE UNDER FLUOROSCOPY AND PHYSICIAN STATED THAT THE PHYSICIAN MIGHT HAVE OVERTIGHTENED THE HUB WHEN THE SUBJECT COIL WAS INSERTED INTO THE MICRO CATHETER AND THE COIL GOT STRETCHED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11396334 ↗InjuryMar 29, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT DURING A COIL EMBOLIZATION PROCEDURE, THE COIL (SUBJECT DEVICE) PERFORATED AND PROTRUDED OUT OF THE ANEURYSM CAUSING THE ANEURYSM TO BURST AND BLEEDING TO OCCUR. THE PATIENT WAS TREATED WITH PROTAMINE AND THE ANEURYSM WAS COILED SUCCESSFULLY AND THE BLEEDING STOPPED. THE PATIENT IS RECOVERING. THERE IS NO FURTHER INFORMATION AVAILABLE.
Patient problems: Aneurysm; Intracranial Hemorrhage; Rupture
MAUDE report 7383134 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page