Skip to content

TARGET M0035478200Recalls & reported problems

Stryker Corporation · 360 Soft Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540676597), received between Dec 6, 2016 and Sep 16, 2022 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Fracture1

Patient problems

  • Embolism/Embolus1
  • Hernia1
  • No Consequences Or Impact To Patient1
  • Stroke/CVA1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathSep 16, 2022Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING RUPTURED ANEURYSM, WHEN PHYSICIAN TRIED TO REPOSITION THE SUBJECT COIL, IT WAS STUCK IN THE CATHETER AND DID NOT MOVE AT ALL. PHYSICIAN TRIED TO RETRACT THE SUBJECT COIL BUT WAS UNSUCCESSFUL, SO USED A SNAIL AND A STENT RETRIEVER AS A MEDICAL INTERVENTION. WHEN REMOVED THE SUBJECT COIL FOUND THAT THE LAST LOOP WAS HOOKED IN ANOTHER HOOP, AND IT WAS STRETCHED COMPLETELY. THERE WAS A SURGICAL DELAY OF 5 HOURS. AS A RESULT, THE PATIENT HAD A BIG STROKE ON THE LEFT SIDE. THE PATIENT HAS ANOTHER ANEURYSM ON THE OTHER SIDE THAT INITIALLY WANTED TO TREAT AT THE SAME TIME, BUT IT COULD NOT BE TREATED. PHYSICIAN DECIDED TO PROCEED IN THE COMING DAYS. PATIENT CURRENT STATUS IS UNKNOWN. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Embolism/Embolus; Hernia; Stroke/CVA; Thrombosis/Thrombus

    MAUDE report 15439906 ↗
  • MalfunctionDec 6, 2016Break; Break

    IT WAS REPORTED THAT DURING A COIL EMBOLIZATION PROCEDURE, THE SUBJECT COIL FRACTURED IN THE (NON-STRYKER) MICROCATHETER. THE COIL AND THE MICROCATHETER WERE REMOVED ALTOGETHER AS ONE UNIT. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6147450 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates