TARGET M0035478300Recalls & reported problems
Stryker Corporation · 360 Soft Detachable Coil · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04546540676603), received between Sep 8, 2018 and Oct 17, 2025 — classified as 4 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Premature Separation” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Separation8
- Fracture2
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2025Fracture; Fracture
THE COIL (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS, AND IT WAS REVEALED THAT THE MAIN COIL (SUBJECT DEVICE) AND COIL DELIVERY WIRE (SUBJECT DEVICE) WERE BROKEN/FRACTURED DURING USE. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23330729 ↗MalfunctionMar 12, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT COIL DETACHED WITHOUT USING A POWER SUPPLY AND THE SUBJECT COIL WAS STRETCHED. THE SUBJECT COIL WAS REMOVED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18887253 ↗MalfunctionDec 15, 2023Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT COIL DETACHED PREMATURELY WHEN ONLY TWO LOOPS WERE IN THE ANEURYSM. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18336971 ↗MalfunctionApr 26, 2023Premature Separation; Premature Separation
IT WAS REPORTED DURING COILING PROCEDURE, THE SUBJECT COIL WAS PREMATURELY DETACHED INSIDE THE MICROCATHETER . THE MICROCATHETER WAS COMPLETELY REMOVED FROM THE PATIENT. THERE WERE NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT DUE TO THIS EVENT. NO FURTHER INFORMATION AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16820893 ↗InjuryJul 11, 2022Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE TREATMENT OF POSTERIOR CEREBRAL ARTERY ANEURYSM THE SUBJECT COIL PREMATURELY DETACHED IN THE CATHETER AND WAS REMOVED USING A SNARE. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14981957 ↗MalfunctionJun 28, 2022Premature Separation; Premature Separation
THE DEVICE WAS RETURNED FOR ANALYSIS AND THE INVESTIGATION OF THE DEVICE REVEALED THAT THE COIL(SUBJECT DEVICE) PREMATURELY DETACHED DURING USE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14841984 ↗InjuryMar 20, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT COIL WAS STRETCHED AT THE DISTAL PART UPON DEPLOYMENT, AND IT COULD NOT BE ADVANCE OR RETRIEVED INSIDE THE MICROCATHETER. THE PHYSICIAN APPLIED FORCE TO REMOVE THE COIL (SUBJECT DEVICE) BUT THE MAIN COIL WAS FRACTURED AND WAS LEFT INSIDE THE ANATOMY. A SNARE WAS USED TO RETRIEVE THE FRAGMENT BUT REMOVAL WAS NOT SUCCESSFUL. THERE WAS A 40 MINS DELAY DURING THE PROCEDURE TO REMOVE THE COIL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 13831778 ↗InjuryNov 3, 2021Adverse Event Without Identified Device or Use Problem; Premature Separation; Adverse Event Without Identified Device or Use Problem; Premature Separation
IT WAS REPORTED THAT DURING POSTERIOR COMMUNICATING ARTERY (PCOMA) ANEURYSM PROCEDURE, THE PHYSICIAN PLACED A GUIDE CATHETER, GUIDEWIRE AND MICROCATHETER TO BUILD ACCESS TO THE ANEURYSM AND DEPLOYED A STENT. THEN PREPARED TO FRAME THE SUBJECT COIL TO THE TREATMENT SITE. PHYSICIAN FELT FRICTION WHILE ADVANCING THE SUBJECT COIL IN THE DISTAL OF THE MICROCATHETER AND THE SUBJECT COIL WAS DEPLOYED PREMATURELY INSIDE THE MICROCATHETER. A SNARE DEVICE WAS USED TO RETRIEVE THE SUBJECT COIL. THE PHYSICIAN REPLACED IT WITH A SAME NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12742494 ↗MalfunctionJun 10, 2019Premature Separation; Premature Separation
ANALYSIS OF THE DEVICE REVEALED THAT THE SUBJECT COIL PREMATURELY DETACHED INSIDE PATIENT. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8684354 ↗MalfunctionDec 29, 2018Break; Break
DURING THE STENT ASSISTING COIL EMBOLIZATION PROCEDURE FOR AN ANEURYSM, IT WAS REPORTED THAT BIG RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE 9TH COIL IN THE MICROCATHETER. SO THE PHYSICIAN WITHDRAW THE COIL AND FOUND IT WAS STRETCHED AND FRACTURED. THEN THE WHOLE COIL WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY AFTER CHANGING TO ANOTHER COIL. THE PATIENT¿S CONDITION WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8205262 ↗InjurySep 8, 2018Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING AVF (ARTERIOVENOUS FISTULA) SIGMOID SINUS PROCEDURE, THE SUBJECT DEVICE (COIL) WAS STUCK INSIDE THE COIL MASS. THE SUBJECT DEVICE WAS WITHDRAWN CAREFULLY BUT IT WAS PREMATURELY DETACHED INSIDE THE MICROCATHETER. SINCE THE SUBJECT DEVICE WAS TANGLED WITH THE COIL MASS, THE SUBJECT DEVICE WAS RETRIEVED AND SAFELY REMOVED FROM THE PATIENT BY USING SNARE DEVICE. THE PROCEDURE WAS FINISHED WITHOUT ADDITIONAL COIL. THE PATIENT¿S CURRENT CONDITION IS FINE.NO CLINICAL CONSEQUENCES TO THE PATIENT WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7858444 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page