Skip to content

INZONE M00345100950Recalls & reported problems

Stryker Corporation · DETACHMENT SYSTEM · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540697950), received between Aug 3, 2020 and Feb 9, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Expulsion” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 6, 2026Class IIOpen, Classified

    Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.

    Z-2552-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Expulsion1

Patient problems

  • Insufficient Information1
  • Intracranial Hemorrhage1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026

    FROM STAFF: S [SITUATION]: MMA [MIDDLE MENINGEAL ARTERY] EMBOLIZATION BEING COMPLETED. A STRYKER COIL DETACHER WAS OPENED LOT # WMP137646 WAS OPENED AND THE PULL TAB WAS NOT THERE, AND THE DEVICE WOULD NOT TURN ON. B [BACKGROUND]: A STRYKER COIL NEEDED TO BE USED TO COMPLETE THE PROCEDURE TO PREVENT FURTHER GROWTH OF SDH [SUBDURAL HEMATOMA]. A [ASSESSMENT]: A DEVICE MALFUNCTION. A SECOND DEVICE WAS OPENED AND WAS SUCCESSFUL. R [RECOMMENDATION]: REP MAY WANT TO BE NOTIFIED TO LET THEM KNOW THERE WAS A MALFUNCTION IN PRODUCT. A SECOND COIL DETACHER WAS OPENED AND DID WORK.

    Patient problems: Insufficient Information

    MAUDE report 24290915 ↗
  • InjuryAug 3, 2020Expulsion; Adverse Event Without Identified Device or Use Problem; Expulsion; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING PROCEDURE FOR COILING OF ANEURYSM, SUBJECT COIL WAS NOTED TO BE UNABLE TO DETACH. THE DETACHMENT SYSTEM FAILED TO INDICATE THAT THE SUBJECT COIL DID NOT DETACH SUCCESSFULLY. ONE SUBJECT COIL LOOP REMAINED IN VESSEL. PATIENT EXPERIENCED ICB (INTRACRANIAL BLEEDING) AND SAH (SUBARACHNOID HEMORRHAGE). THE PROCEDURE WAS COMPLETED SUCCESSFULLY, NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Intracranial Hemorrhage

    MAUDE report 10360309 ↗

FDA clearance and classification

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the INZONE M00345100950

Has the INZONE M00345100950 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2552-2026 (Class II), began on Apr 6, 2026: Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the INZONE M00345100950?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 3, 2020 and Feb 9, 2026. In those reports, the most common specific device problem is “Expulsion” (1); the most common patient problem is “Intracranial Hemorrhage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the INZONE M00345100950?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the INZONE M00345100950, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand INZONE, model M00345100950). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page