Skip to content

Target M0036154080Recalls & reported problems

Stryker Corporation · 3D Detachable Coil · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540711670), received between Jun 7, 2022 and Sep 13, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Migration1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 13, 2024Migration; Migration

    IT WAS REPORTED THAT DURING A COIL EMBOLIZATION PROCEDURE FOR A 4 MM ANEURYSM WAS ABOUT TO BE TREATED WITHOUT A STENT. A MICROCATHETER WAS GUIDED INTO THE ANEURYSM, AND THE FIRST COIL (SUBJECT DEVICE) AND SECOND COIL WERE IMPLANTED. HOWEVER, DURING THE INSERTION OF THE THIRD, IT DID NOT ENGAGE WITH THE MASS, AND WHILE REPOSITIONING THE THIRD COIL, IT BECAME ENTANGLED WITH NEAR THE TIP OF THE FIRST SUBJECT COIL IMPLANTED AND MADE A KNOT, MAKING IT IMPOSSIBLE TO RETRIEVE THE THIRD COIL. WHILE PUSHING AND PULLING THE THIRD COIL TO UNTIE THE KNOT, THE SUBJECT COIL DEVIATED FROM THE MASS, AND A STENT THAT WAS NOT PLANNED TO USE HAD TO BE USED. BY PUSHING AND PULLING THE COILS, THE KNOT WAS UNTIED, AND THE THIRD COIL WAS MANAGED TO BE RETRIEVED. THREE ADDITIONAL COILS WERE INSERTED INTO THE ANEURYSM AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT. HOWEVER, AS A STENT WAS USED, THE PATIENT WILL REQUIRE LONG-TERM MEDICATION. THE EXTENDED OPERATIVE TIME ALSO RESULTED IN INCREASED RADIATION EXPOSURE. THE PATIENT¿S CURRENT CONDITION IS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212871 ↗
  • MalfunctionJan 5, 2024Premature Separation; Premature Separation

    THE SUBJECT COIL WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE MAIN COIL PREMATURELY DETACHED/SEPARATED DURING USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18457366 ↗
  • MalfunctionJun 7, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING THE COIL EMBOLIZATION PROCEDURE, THE PHYSICIAN PLACED THE COIL (SUBJECT DEVICE) INTO THE ANEURYSM. THE PROXIMAL PART OF DELIVERY WIRE OF THE SUBJECT COIL BROKE AND THE PHYSICIAN WAS UNABLE TO DETACH THE SUBJECT COIL. THE PHYSICIAN REMOVED THE SUBJECT COIL FROM THE PATIENT ANATOMY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14617796 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates