MIRACLEBROS 6 AG14M060Recalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547327056083), received between Jun 20, 2023 and Jun 23, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Fracture1
Patient problems
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2023Fracture; Fracture
THIS IS AN EXERT FROM THE PROCEDURAL NOTE DESCRIBING THE FRACTURE OF A ''MIRACLE BROTHER WIRE'. APPROXIMATELY 13.2 MM OF WIRE REMAINED BEHIND. WHILE "TRYING TO PENETRATE THE DISTAL CAP OF A CORONARY VESSEL WITH A MIRACLE BROTHERS WIRE, IT APPEARED AS THOUGH THE WIRE HAD GONE SUBINTIMAL ON RETROGRADE INJECTION. WHEN THE MIRACLE BROTHER WIRE WAS ATTEMPTED TO BE RETRACTED THERE WAS RESISTANCE SO INITIAL ATTEMPTS WERE MADE TO REMOVE THE WIRE AS A UNIT WITH THE TURNPIKE MICROCATHETER AND THEN THE XB 3.5 GUIDE CATHETER UNTIL THE NO MORE RESISTANCE WAS NOTED BUT ON ANGIOGRAPHY IT APPEARED THAT THE FLOPPY TIP OF THE MIRACLE BROTHER WIRE HAD FRACTURED INTO THE SUBINTIMAL SPACE WHICH WAS CONFIRMED ON ORTHOGONAL AND DUAL ANGIOGRAPHIES. AS THE WIRE WAS REMOVED FROM THE BODY IT WAS CONFIRMED ON GROSS EXAMINATION THAT IT WAS THE FLOPPY TIP OF THE WIRE WHICH HAD FRACTURED AND WAS LEFT IN THE SUBINTIMAL SPACE. AS THERE WAS NO CHANGE IN THE CLINICAL STATUS OF THE PATIENT WITH INTACT HEMODYNAMICS, STABLE EKG FINDINGS, AND LACK OF WORSENING CHEST PAIN WE DECIDED AGAINST SNARING THE FRACTURED WIRE TIP DUE TO THE RISKS VS BENEFITS AS IT WAS IN THE SUBINTIMAL SPACE AND UNLIKELY TO BECOME THROMBOGENIC IN
Patient problems: Insufficient Information
MAUDE report 17195704 ↗InjuryJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
USER FACILITY REPORT (REPORT #:MW5117733) WAS RECEIVED FROM THE US DISTRIBUTOR. IT WAS REPORTED THAT THE TIP OF AN ASAHI MIRACLEBROS 6 GUIDE WIRE HAD FRACTURED AND APPROXIMATELY 13.2MM OF THE WIRE TIP REMAINED IN THE PATIENT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE UNSPECIFIED CORONARY ARTERY. THE GUIDE WIRE WAS USED TO PENETRATE THE DISTAL CAP OF THE CTO. IT APPEARED AS THOUGH THE GUIDE WIRE HAD GONE SUBINTIMAL ON RETROGRADE INJECTION ANGIOGRAM. WHEN THE GUIDE WIRE WAS ATTEMPTED TO BE RETRACTED, THERE WAS RESISTANCE. INITIAL ATTEMPTS WERE MADE TO REMOVE THE GUIDE WIRE AS A UNIT WITH A NON-ASAHI TURNPIKE MICROCATHETER AND THEN A NON-ASAHI XB 3.5 GUIDE CATHETER UNTIL NO MORE RESISTANCE WAS NOTED. BUT ON ANGIOGRAPHY IT APPEARED THAT THE FLOPPY TIP OF THE GUIDE WIRE HAD FRACTURED INTO THE SUBINTIMAL SPACE WHICH WAS CONFIRMED ON ORTHOGONAL AND DUAL ANGIOGRAPHIES. AS THE GUIDE WIRE WAS REMOVED FROM THE BODY IT WAS CONFIRMED ON GROSS EXAMINATION THAT IT WAS THE FLOPPY TIP OF THE WIRE WHICH HAD FRACTURED AND WAS LEFT IN THE SUBINTIMAL SPACE. AS THERE WAS NO CHANGE IN THE CLINICAL STATUS OF THE PATIENT WITH INTACT HEMODYNAMICS, STABL
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 17167917 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page