FIELDER XT AGP140002Recalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04547327072854), received between Dec 11, 2015 and Dec 9, 2024 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment Of Device Component2
- Break1
- Detachment of Device or Device Component1
- Device Damaged by Another Device1
- Difficult to Remove1
- Failure to Advance1
- Improper or Incorrect Procedure or Method1
- Kinked1
Patient problems
- No Consequences Or Impact To Patient4
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryDec 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT ATTEMPTS WERE MADE TO ADVANCE A KNUCKLED ASAHI FIELDER XT GUIDE WIRE DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR A CHRONIC TOTAL OCCLUSION (CTO) IN THE UNSPECIFIED CORONARY ARTERY BUT FAILED. DURING WITHDRAWAL ATTEMPTS OF THE FIELDER XT GUIDE WIRE, THE DISTAL SEGMENT OF THE GUIDE WIRE BROKE OFF IN THE PATIENT ARTERY. THE WIRE FRAGMENT WAS JAILED WITH AN UNSPECIFIED STENT. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20876096 ↗MalfunctionDec 2, 2024Material Fragmentation; Material Fragmentation
WIRE WAS KNUCKLED AND PHYSICIAN WAS TRYING TO ADVANCE THE WIRE BUT IT WOULD NOT GO. WHEN ATTEMPTING TO REMOVE THE WIRE, THE TIP BROKE OFF IN THE PATIENT'S ARTERY. BROKEN TIP WAS JAILED WITH A STENT. NO ADVERSE AFFECT TO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20826594 ↗InjuryMar 19, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
USER FACILITY REPORT #: (B)(4) WAS RECEIVED ON MARCH 5, 2020. IT WAS REPORTED THAT THE TIP OF AN ASAHI FIELDER XT GUIDE WIRE BROKE OFF DURING A PCI TO TREAT AN UNSPECIFIED LESION IN THE LEFT HEART. THE GUIDE WIRE WAS USED TO SUPPORT UNSPECIFIED BALLOON AND CATHETER THAT WERE KEPT COMING OUT OF THE VESSEL. THE GUIDE WIRE WAS RETRIEVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9851810 ↗MalfunctionMar 6, 2020Break
WIRE TIP BROKE OFF WHEN PHYSICIAN WAS TRYING TO SUPPORT BALLOON AND CATHETER KEPT COMING OUT OF VESSEL. DEVICE RETRIEVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9803036 ↗InjuryJan 8, 2019Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A CHRONIC TOTALLY OCCLUDED CIRCUMFLEX CORONARY ARTERY. THE ASAHI GUIDE WIRE WAS ADVANCED TO THE TARGET LESION, BUT DURING TORQUEING OF THE GUIDE WIRE, THERE WAS TEFLON FLAKING ON THE PROXIMAL END OF THE GUIDE WIRE NEXT TO THE TUOHY. THE FLAKING OF TEFLON WAS OUTSIDE THE PATIENT ANATOMY. HOWEVER, IT WAS DECIDED TO REMOVE THE GUIDE WIRE. THE GUIDE WIRE WAS REMOVED WITHOUT ANY ISSUE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. IT WAS THEN DECIDED TO HAVE THE PATIENT SENT FOR CORONARY ARTERY BYPASS (CABG) DUE TO HAVING MULTI-VESSEL DISEASE, BUT NOT RELATED TO THE GUIDE WIRE USED OR THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8229701 ↗InjuryApr 20, 2018Detachment Of Device Component; Difficult to Remove; Improper or Incorrect Procedure or Method; Physical Resistance; Device Damaged by Another Device
USER FACILITY MEDWATCH REPORT RECEIVED THAT STATES: RESISTANCE WITH THE ANATOMY WAS FELT WHEN ADVANCING A FIELDER XT WIRE. THREE STENTS WERE DEPLOYED WITHOUT DIFFICULTY; HOWEVER, RESISTANCE REMOVING THE FIELDER XT WIRE WAS FELT AND FORCE WAS APPLIED. IT NOTICED THE TIP OF THE WIRE WAS DISLODGED/UNRAVELED IN THE LEFT ANTERIOR CORONARY ARTERY (LAD). AN ATTEMPT TO USE SNARES TO REMOVE THE WIRE TIP WAS MADE AND PIECES WERE REMOVED, BUT SOME TIP REMAINED. TWO WIRES WERE THEN ADVANCED BEYOND THE WIRE TIP AND THE WIRES WERE TWISTED TO CAPTURE THE REMAINING TIP. ON THE FIRST ATTEMPT, A PIECE OF THE WIRE WAS REMOVED. ON A SECOND ATTEMPT, THE REST OF THE WIRE TIP WAS SUCCESSFULLY REMOVED. INTRAVASCULAR ULTRASOUND WAS PERFORMED IN THE LAD TO VERIFY THE STENT WAS NO LONGER IN THE VESSEL. THE 3.0X22MM MEDTRONIC RESOLUTE ONYX STENT THAT HAD BEEN DEPLOYED EARLIER IN THE CASE HAD TO BE REMOVED AND REPLACED DUE TO MANIPULATION LOOKING FOR THE WIRE PIECES. NO DAMAGE TO THE NON-ABBOTT STENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7448584 ↗InjuryDec 11, 2015Detachment Of Device Component; Kinked; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A CHRONIC TOTAL OCCLUSION IN THE RIGHT CORONARY ARTERY. THE FIELDER XT190 GUIDE WIRE WAS CROSSED TO THE LESION; HOWEVER, KEPT HITTING THE OCCLUSION AND BOUNCING BACK. THE GUIDE WIRE WAS ABLE TO PARTIALLY ENTER THE LESION AND DURING DOTTERING OF THE GUIDE WIRE TO CROSS FULLY, THE GUIDE WIRE TIP KINKED AND SEPARATED. A SNARE DEVICE WAS USED TO SUCCESSFULLY CAPTURE THE GUIDE WIRE TIP WHICH WAS REMOVED FROM THE ANATOMY. THE PROCEDURE CONTINUED ON USING NEW DEVICES WITHOUT FURTHER INCIDENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 5290817 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page