ASAHI SION black APW14R010SRecalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04547327089685), received between Aug 16, 2021 and Nov 20, 2025 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Fracture2
- Material Separation2
- Difficult to Remove1
- Physical Resistance/Sticking1
- Unraveled Material1
Patient problems
- Foreign Body In Patient3
- Insufficient Information3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Fracture; Difficult to Remove
THE TIP OF THE GUIDEWIRE FRACTURED OFF AND WAS UNABLE TO BE RETRIEVED.
Patient problems: Insufficient Information
MAUDE report 23604344 ↗InjuryNov 6, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ASAHI SION BLACK GUIDE WIRE WAS USED DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR A HEAVILY TORTUOUS AND MILDLY CALCIFIED CHRONIC TOTAL OCCLUSION (CTO) IN THE LEFT CIRCUMFLEX ARTERY (LCX) AND THE OTHER CTO IN THE RIGHT CORONARY ARTERY (RCA) BY WAY OF THE RETROGRADE APPROACH. DURING WIRING THROUGH A SEPTAL BRANCH, THE DISTAL SEGMENT OF THE GUIDE WIRE WAS UNRAVELED, LEAVING A FRAGMENT IN THE SEPTAL BRANCH. THE LCX LESION WAS SUCCESSFULLY CROSSED AND BALLOONED, WHILE THE RCA LESION WAS ALSO SUCCESSFULLY CROSSED AND STENTED. BLOOD FLOW WAS REESTABLISHED AND THE PROCEDURE WAS COMPLETED. IT WAS INFORMED THAT THE PATIENT WAS DOING FINE AFTER THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 20616453 ↗MalfunctionOct 31, 2024Fracture; Unraveled Material; Fracture; Unraveled Material
WIRING THROUGH A SEPTAL BRANCH, THE SION BLACK TIP UNRAVELED AND LEFT A PIECE DISTAL IN THE SEPTAL BRANCH. PATIENT WAS OK AND RIGHT WAS CROSSED AND STENTED. PIECE LEFT IN DISTAL SEPTAL BRANCH.
Patient problems: Foreign Body In Patient
MAUDE report 20584404 ↗MalfunctionSep 1, 2021Detachment of Device or Device Component
WIRE TIP CAME OFF WHILE NAVIGATING STENT STRUTS. MANUFACTURER RESPONSE FOR WIRE, GUIDE, CATHETER, ASAHI SION BLACK (PER SITE REPORTER). GOOD FOLLOW-UP.
Patient problems: Insufficient Information
MAUDE report 12405438 ↗InjuryAug 27, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP OF AN ASAHI SION BLACK GUIDE WIRE BROKE OFF DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE LEFT CIRCUMFLEX ARTERY (LCX). A STENT WAS DEPLOYED INTO THE LEFT MAIN TRUNK (LMT) TO THE LCX. THEN, THE SION BLACK WAS USED TO WIRE THROUGH THE STENT TO DILATE THE STENT STRUTS AND FURTHER SUPPORT NORMAL BLOOD FLOW THROUGH STENT INTO LCX. THE SION BLACK NAVIGATED THE STENT STRUTS AND WAS GUIDED DOWN THE LCX FOR AN UNKNOWN DISTANCE. A RX CORONARY BALLOON WAS ADVANCED OVER THE SION BLACK WITH THE PLAN TO PASS THROUGH THE STRUTS AND DILATE THE OSTIUM. THE BALLOON WOULD NOT CROSS THE STRUTS AND WAS REMOVED. IT IS UNKNOWN IF THE BALLOON WAS INFLATED AT ANY TIME. AN ATTEMPT WAS THEN MADE TO REMOVE THE SION BLACK BUT IT WAS STUCK IN THE LCX. FORCE WAS USED AND THE DISTAL WIRE SEPARATED FROM THE BODY AT THE JUNCTION OF THE RADIOPAQUE TO THE BODY OF THE WIRE. THE PHYSICIAN COMMENTED THAT THE A CARDIAC INTERVENTIONAL FELLOW, WHO HAS JUST STARTED THEIR FOURTH YEAR/HANDS-ON, WAS DOING THE WIRING OF THE LCX AND MIGHT HAVE "STITCHED" THE WIRE IN AND OUT OF STENT OPENINGS BEFORE ENTERING THE LCX. WHEN THE PHYSICIAN WENT TO RETRI
Patient problems: Foreign Body In Patient
MAUDE report 12376236 ↗InjuryAug 16, 2021Material Separation; Physical Resistance/Sticking
THESE ARE THE EVENTS AS THEY WERE PROVIDED TO THE SALES REP. SALES REP WAS NOT PRESENT DURING THIS EVENT. THE SION BLACK WAS USED TO WIRE THROUGH FRESHLY PLACED LEFT MAIN STENT INTO THE LCX. THE GOAL WAS TO DILATE THE STENT STRUTS AND FURTHER SUPPORT NORMAL BLOOD FLOW THROUGH STENT INTO LCX. THE LCX WAS A CTO AT THE BEGINNING OF THE PROCEDURE. SALES REP DOES NOT KNOW IF THIS WAS A NEW WIRE OR IF IT HAD BEEN USED EARLIER IN THE PROCEDURE. SALES REP DOES NOT KNOW THE STATE OF THE LCX AFTER REVASCULARIZATION OR IF ANY STENTING THAT MAY HAVE BEEN DONE. THE SION BLACK NAVIGATED THE STENT STRUTS AND WAS GUIDED DOWN THE LCX FOR AN UNKNOWN DISTANCE. A RX CORONARY BALLOON WAS ADVANCED OVER THE SION BLACK WITH THE PLAN TO PASS THROUGH THE STRUTS AND DILATE THE OSTIUM. THE BALLOON WOULD NOT CROSS THE STRUTS AND WAS REMOVED, SALES REP DOES NOT HAVE INFORMATION IF IT WAS INFLATED AT ANY TIME. AN ATTEMPT WAS THEN MADE TO REMOVE THE SION BLACK BUT IT WAS STUCK IN THE LCX. FORCE WAS USED AND THE DISTAL WIRE SEPARATED FROM THE BODY AT THE JUNCTION OF THE RADIOPAQUE TO THE BODY OF THE WIRE. ATTEMPTS WERE MADE TO RETRIEVE THE DISTAL END OF THE WIRE USING A MICRO SNARE BUT THEY WERE UNSUCCESSFUL. FURT
Patient problems: Insufficient Information
MAUDE report 12329142 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page