Skip to content

ASAHI Gaia Third AHW14R311PRecalls & reported problems

Asahi Intecc CO., Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547327094795), received between Oct 21, 2022 and Oct 30, 2022 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Material Separation1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 30, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TIP OF AN ASAHI GAIA THIRD GUIDE WIRE SEPARATED DURING A PERCUTANEOUS CORONARY INTERVENTION TO TREAT A MILDLY CALCIFIED, MODERATELY TORTUOUS, CHRONIC TOTAL OCCLUSION (CTO) IN THE OBTUSE MARGINAL BRANCH (OM). THE TIP DETACHED WHILE ATTEMPTING TO CROSS CTO OF OM. THE TIP BECAME STUCK BEHIND THE STRUT OF AN EXISTING STENT IN OM AND DETACHED. THE TIP REMAINED IN THE PATIENT. SNARING WAS DISCUSSED BUT THE PHYSICIAN FELT THAT IT WOULD NOT BE POSSIBLE TO RETRIEVE THE TIP SINCE IT WAS UNDER THE STENT STRUT. THE PROCEDURE WAS COMPLETED USING THE STAR TECHNIQUE. THE PATIENT WAS DISCHARGED AFTER TREATMENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 15697545 ↗
  • MalfunctionOct 21, 2022Material Separation; Material Separation

    TIP OF GAIA THIRD SEPARATED AND DETACHED WHILE ATTEMPTING TO CROSS CTO OF OM. THE TIP BECAME STUCK BEHIND A STENT STRUT AND DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15651805 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page