ASAHI Gaia PV18 PHW18R302PRecalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04547327106917), received between Nov 22, 2022 and May 30, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Material Integrity Problem1
- Material Separation1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Intraoperative Cardiac Valve Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 30, 2025Material Integrity Problem; Material Integrity Problem
LESION DETAILS: PATIENT'S SFA WAS CHRONICALLY OCCLUDED & RECONSTITUTED SOMEWHERE BETWEEN THE POPLITERAL AND TIBIAL PERONEAL TRUNK. EVENT DESCRIPTION: THE PHYSICIAN WAS SUBINTIMAL THROUGH THE SFA INTO THE POPLITEAL AND AT. HE WAS ATTEMPTING TO RE-ENTER WITH THE GAIA PV AT THE TAKE OFF OF THE AT. THE GAIA PV SEEMED BUCKLED AND COLLAPSED ONTO ITSELF. UPON TRYING TO REMOVE THE WIRE, THE TIP OF THE WIRE BEGAN TO UNRAVEL. WE WERE ABLE TO REMOVE THE WIRE VIA THE SUPPORT CATHETER WITHOUT LOSING ANY PARTS OF THE WIRE. THE PHYSICIAN WAS TORQUING THE WIRE AT A SLOW STEADY PACE AND IS VERY FAMILIAR WITH THE WIRE IN HIS ALGORITHM. REMEDIAL ACTION: THE PHYSICIAN WAS ABLE TO REMOVE THE WIRE WITHOUT ANY BREAKING APART OF ANY WIRE PIECES. POST PROCEDURE OUTCOME: THE PHYSICIAN WAS UNABLE TO REENTER THE TRUE LUMEN DISTAL TO THE CTO AND REVASCULARIZE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22119276 ↗InjuryMay 16, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ENDOVASCULAR TREATMENT (EVT) WAS PERFORMED FOR A HEAVILY CALCIFIED AND SEVERELY FLEXUOUS CHRONIC TOTAL OCCLUSION (CTO) IN THE POPLITEAL ARTERY. AN ATTEMPT WAS MADE WITH AN ASAHI GAIA PV GUIDE WIRE RETROGRADELY FROM THE POSTERIOR TIBIAL ARTERY TO THE POPLITEAL ARTERY TO APPROACH THE CTO CAP IN THE POPLITEAL ARTERY. THE GAIA PV GUIDE WIRE WANDERED IN BETWEEN THE CTO AND THE SUBINTIMAL SPACE, AND STOPPED TORQUEING DUE TO THE ENTRAPMENT. THE PHYSICIAN TRIED SPINNING THE GUIDE WIRE TO ADVANCE THROUGH THE LESION, BUT THE GUIDE WIRE WOULD NOT MOVE. WHEN THE PHYSICIAN REMOVED THE GAIA PV GUIDE WIRE, THE WIRE BEGAN TO UNRAVEL AND GOT DETACHED, LEAVING A WIRE FRAGMENT IN THE POSTERIOR TIBIAL ARTERY. IT WAS INFORMED THAT SURGICAL TREATMENT WAS PERFORMED TO REMOVE THE FRAGMENT OF THE GAIA PV GUIDE WIRE, AND THE PATIENT WAS DOING FINE AFTERWARDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22033349 ↗MalfunctionMar 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT AN ENDOVASCULAR TREATMENT (EVT) WAS PERFORMED FOR A MILDLY CALCIFIED AND MILDLY TORTUOUS CHRONIC TOTAL OCCLUSION (CTO) IN THE ANTERIOR TIBIAL ARTERY (ATA). AFTER FEMORAL PUNCTURE, AN ASAHI GAIA PV18 GUIDE WIRE WAS ADVANCED SUBINTIMAL TO THE ATA AS THE SUPERFICIAL FEMORAL ARTERY (SFA) HAD BEEN CHRONICALLY OCCLUDED AND RECONSTRUCTED SOMEWHERE BETWEEN THE POPLITEAL ARTERY AND THE TIBIOPERONEAL TRUNK. WHEN AN ATTEMPT WAS MADE TO ENTER THE GAIA PV18 GUIDE WIRE FROM THE ATA TAKEOFF, THE GUIDE WIRE LOOKED BUCKLED AND COLLAPSED ONTO ITSELF. DURING REMOVAL ATTEMPTS, THE TIP OF THE GUIDE WIRE BEGAN TO UNRAVEL. WITH USE OF A TRAILBLAZER SUPPORT CATHETER (MEDTRONIC), THE GAIA PV18 GUIDE WIRE WAS REMOVED EX SITU WITHOUT LOSING ANY PART OF IT. THE TRUE LUMEN DISTAL TO THE CTO COULD NOT BE ACCESSED AND REVASCULARIZATION COULD NOT BE ACHIEVED. IT WAS INFORMED THAT THE PHYSICIAN TORQUED THE GUIDE WIRE AT A SLOW STEADY PACE AND WAS VERY FAMILIAR WITH THE WIRE AND ALGORITHM. IT WAS ALSO INFORMED THAT THERE WAS NO HEALTH HAZARDS CAUSED TO THE PATIENT DUE TO THE REPORTED EVENT.
Patient problems: Intraoperative Cardiac Valve Injury
MAUDE report 21674526 ↗MalfunctionNov 23, 2022Fracture; Fracture
DURING THE PROCEDURE THE PHYSICIAN SAID THE WIRE FELT "TRAPPED" HE MOVES THE WIRE AND GOT IT TO RELEASE. WHEN HE DID THAT THE WIRE BEGAN TO UNRAVEL. HE WAS ABLE TO REMOVE THE WIRE WITHOUT ANY PIECES BEING LOST. THERE WAS NO ADVERSE REACTION TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15855556 ↗MalfunctionNov 22, 2022Fracture; Fracture
IT WAS REPORTED THAT AN ASAHI GAIA PV GUIDE WIRE WAS USED FOR A PERCUTANEOUS PERIPHERAL INTERVENTION TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED, 75-90% STENOSIS IN THE UNSPECIFIED LIMB ARTERY. DURING THE PROCEDURE, THE WIRE SEEMED "TRAPPED." WHEN ATTEMPTING TO REMOVE AND RELEASE THE TRAPPED WIRE, THE WIRE BEGAN TO UNRAVEL. THE WIRE WAS ABLE TO BE REMOVED WITHOUT ANY PIECES BEING LOST. THERE WAS NO ADVERSE REACTION TO THE PATIENT. THE PATIENT WAS FINE AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15841180 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page