Skip to content

SUOH 03 AHW14R013SRecalls & reported problems

Asahi Intecc CO., Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547327116749), received between May 19, 2021 and Jun 3, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Fracture1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2021Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT INTERVENTION WAS PERFORMED TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE LEFT CIRCUMFLEX ARTERY (LCX). ANTEGRADE WIRE ESCALATION WAS PERFORMED THROUGH A NON-ASAHI MAMBA CATHETER: FIRST WITH A NON-ASAHI RUNTHROUGH WIRE, A NON-ASAHI FIGHTER WIRE. IT WAS AN AMBIGUOUS CAP. THE PHYSICIAN THEN WENT RETROGRADE DOWN THE APICAL LAD AND USED AN ASAHI SUOH 03 GUIDE WIRE TO ACROSS A HEAVILY TORTUOUS EPICARDIAL COLLATERAL. AS HE WAS ADVANCING AN ASAHI CORSAIR PRO XS MICROCATHETER, THE TIP OF THE GUIDE WIRE BROKE OFF INTO THE COLLATERAL VESSEL. HE WAS ABLE TO SUCCESSFULLY RETRIEVE THE TIP WITH A 4MM MICROSNARE. THE PHYSICIAN COMMENTED THAT THE GUIDE WIRE HAD BEEN PROLAPSED AND KINKED. UPON PULLING IT BACK INTO THE MICROCATHETER, THE MICROCATHETER SHEARED THE COIL OFF OF THE MANDREL OF THE WIRE. ANOTHER SUOH WAS USED AND THE CASE WAS SUCCESSFULLY COMPLETED. THE PATIENT DID GREAT AND EVERYTHING IS FINE. FURTHERMORE, THERE WAS NO ADVERSE REACTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11936369 ↗
  • MalfunctionMay 19, 2021Break

    AS I KNOW OF RIGHT NOW, DR. (B)(6) WAS USING A SUOH-03 ACROSS AN EPICARDIAL COLLATERAL. I DO NOT KNOW IF IT WAS RCA TO LCA OR VICE-VERSA. APPARENTLY, AS HE WAS ADVANCING A MC (DO NOT KNOW WHICH ONE), THE TIP OF THE SUOH BROKE OFF INTO THE COLLATERAL VESSEL. HE WAS ABLE TO SUCCESSFULLY RETRIEVE THE TIP AND THEY DO HAVE THE WIRE. I AM WAITING ON MORE INFORMATION TO COME BACK FROM KATE MULLER (CCL MANAGER).

    Patient problems: Insufficient Information

    MAUDE report 11853030 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page