MINAMO AH14R026SRecalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04547327130745), received between Mar 11, 2020 and Sep 30, 2026 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation3
- Activation, Positioning or Separation Problem1
- Break1
- Detachment of Device or Device Component1
- Physical Resistance/Sticking1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient2
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Material Separation
It was reported that an ASAHI MINAMO guide wire was used for a percutaneous coronary intervention (PCI) to treat an unspecified lesion in the coronary artery. The MINAMO guide wire was unraveled in the patient body. The physician tried to advance an IVUS catheter over the MINAMO guide wire, but the IVUS catheter got stuck. When the MINAMO guide wire was pulled out of the patient body, it was broken. It was informed that there was no adverse patient effect associated with this event and the patient was discharged.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942444 ↗InjurySep 29, 2026Material Separation
It was reported that an ASAHI MINAMO guide wire was used for a percutaneous coronary intervention (PCI) to treat an unspecified lesion in the coronary artery. The MINAMO guide wire was unraveled in the patient body. The physician tried to advance an IVUS catheter over the MINAMO guide wire, but the IVUS catheter got stuck. When the MINAMO guide wire was pulled out of the patient body, it was broken. It was informed that there was no adverse patient effect associated with this event and the patient was discharged.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934447 ↗InjuryFeb 4, 2026Material Separation
IT WAS REPORTED THAT A PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS PERFORMED FOR A LESION IN THE RIGHT CORONARY ARTERY (RCA). AN ASAHI MINAMO GUIDE WIRE WAS ADVANCED INTO THE CORONARY ARTERY FOR STENTING. AFTER THE PROCEDURE WAS COMPLETED WITH STENT DEPLOYMENT, CORONARY COMPUTED TOMOGRAPHY REVELED A TIP FRAGMENT OF THE MINAMO GUIDE WIRE IN THE PATIENT ANATOMY. AS THE TIP FRAGMENT WAS FOUND AFTER COMPLETION OF THE PROCEDURE, NO ATTEMPTS WERE MADE TO RETRIEVE IT. THE PHYSICIAN COMMENTED THAT RESISTANCE WAS MET DURING REMOVAL OF THE MINAMO GUIDE WIRE AFTER DELIVERY OF THE STENT. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 24264632 ↗MalfunctionFeb 4, 2026Activation, Positioning or Separation Problem; Physical Resistance/Sticking
THE WIRE WAS IN PLACE IN THE CORONARY ARTERY FOR PERCUTANEOUS CORONARY STENT PLACEMENT. AFTER STENT DEPLOYED, WIRE ATTEMPTED TO BE WITHDRAWN AND OPERATOR WAS MET WITH RESISTANCE WHICH WAS OVERCOME WHEN HYDROPHILIC TIP DISCONNECTED FROM WIRE AND WAS RETAINED IN STRUTS OF STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259111 ↗InjuryMay 19, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN ASAHI MINAMO GUIDE WIRE WAS BROKEN AND RETAINED IN THE CORONARY ARTERY DURING PERCUTANEOUS CORONARY INTERVENTION TO TREAT A HEAVILY CALCIFIED AND TORTUOUS SUB-TOTAL OCCLUSION IN THE RIGHT CORONARY ARTERY. IT OCCURRED DURING INITIAL WIRING OF THE LESION, BEFORE PRE-DILATATION BALLOON INSERTION. NO MICROCATHETER WAS USED. THE PROCEDURE WAS TERMINATED AND UNSUCCESSFUL. NO ATTEMPTS WERE MADE TO RETRIEVE THE WIRE TIP. THE PATIENT REMAINED STABLE THROUGHOUT AND WAS DISCHARGED HOME. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 14441141 ↗MalfunctionMay 12, 2022Break
DURING PCI, THE "TIP OF THE MINAMO WIRE" WAS BROKEN AND RETAINED IN THE CORONARY ARTERY. NO ADDITIONAL DETAILS WERE KNOWN. I WILL FOLLOW UP WITH ACCOUNT AND WILL BE ON-SITE ON MAY 17TH.
Patient problems: Foreign Body In Patient
MAUDE report 14367090 ↗MalfunctionMar 11, 2020Scratched Material
DR. & STAFF NOTICED RIBBON DISTAL SEGMENT OF WIRE HAVING AN UNCOVERED PORTION 15MM BACK FROM DISTAL TIP OF WIRE, PRECEDED WITH THAT WIRE FOR CASE. PRE-DIL STENTED -POST DIL. COMPLETED CASE AND EVERYTHING REMOVED. NOTHING LEFT IN PATIENT. THEY WERE DISCHARGED -NO INJURY RECOVERED IN NORMAL FASHION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9821829 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page