ASAHI SION AHW14R101SRecalls & reported problems
Asahi Intecc CO., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04547327132985), received between Sep 9, 2025 and Aug 10, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Material Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation4
Patient problems
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient2
- Vascular Dissection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2026Material Separation
It was reported that an Asahi SION guide wire was used during a percutaneous coronary intervention (PCI) to treat a modelately calcified lesion in the proximal left anterior descending artery (LAD). During the removal of the SION guide wire, the SION guide wire entered a subintimal space at the distal edge of the stent and was entrapped. The coil of the SION guide wire was unwound and fractured. An attempt was made to advance an unspecified microcatheter to the wire tip to facilitate retrieval; however, the attempt was unsuccessful. A TERUMO Runthrough guide wire was advanced into the vessel and an unspecified 2.0mm balloon catheter was inflated across the affected segment. The microcatheter was advanced over the SION guide wire. Inspection revealed that the fracture had occurred only at the extreme distal tip of SION guide wire, and the remainder of the SION guide wire was removed intact. Post-intervention angiography showed that the wire fragment was trapped by the newly deployed stent. It was informed that there was no adverse patient effect associated with this event and the patient was discharged in stable condition.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50189606 ↗InjuryAug 6, 2026Material Separation
It was reported that an Asahi SION guide wire was used during a percutaneous coronary intervention (PCI) to treat a modelately calcified lesion in the proximal left anterior descending artery (LAD). During the removal of the SION guide wire, the SION guide wire entered a subintimal space at the distal edge of the stent and was entrapped. The coil of the SION guide wire was unwound and fractured. An attempt was made to advance an unspecified microcatheter to the wire tip to facilitate retrieval; however, the attempt was unsuccessful. A TERUMO Runthrough guide wire was advanced into the vessel and an unspecified 2.0mm balloon catheter was inflated across the affected segment. The microcatheter was advanced over the SION guide wire. Inspection revealed that the fracture had occurred only at the extreme distal tip of SION guide wire, and the remainder of the SION guide wire was removed intact. Post-intervention angiography showed that the wire fragment was trapped by the newly deployed stent. It was informed that there was no adverse patient effect associated with this event and the patient was dischagrged in stable condition.Mfr Report #: 3003775027-2026-00161
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50164779 ↗InjuryOct 28, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ASAHI SION GUIDE WIRE WAS USED FOR STENT DEPLOYMENT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO TREAT AN UNSPECIFIED LESION IN THE CORONARY ARTERY. WHEN REMOVED AFTER STENTING, THE SION GUIDE WIRE BECAME ENTANGLED IN THE STENT. THE GUIDE WIRE THEN BROKE OFF AND THE WIRE FRAGMENT WAS TRAPPED WITHIN THE STENT. IT WAS INFORMED THAT THE PROCEDURE WAS PROLONGED AND BECAME MORE COMPLEX DUE TO AN IATROGENIC DISSECTION BUT COMPLETED. THE CONDITION OF THE PATIENT WAS REPORTEDLY STABLE AFTER THE PROCEDURE AND THE PATIENT WAS RECOVERING WELL.
Patient problems: Foreign Body In Patient; Vascular Dissection
MAUDE report 23399635 ↗InjurySep 9, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS PERFORMED FOR A MILDLY CALCIFIED AND MODERATELY TORTUOUS 90-99% STENOSIS IN THE LEFT CIRCUMFLEX ARTERY (LCX). DURING REMOVAL OF AN ASAHI SION GUIDE WIRE AFTER STENTING IN THE LCX, THE TIP OF THE SION GUIDE WIRE WAS UNRAVELLED AND EVENTUALLY SEPARATED. THE DISTAL SEGMENT OF THE SEPARATED WIRE TIP GOT STUCK IN THE STENT THAT WAS DEPLOYED IN THE LCX AND THE PROXIMAL SEGMENT OF THE FRAGMENT WAS EXTENDED INTO LEFT SUBCLAVIAN ARTERY. VOLCANO IVUS ON THE GUIDE WIRE THAT WAS PLACED IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD) CONFIRMED THE WIRE FRAGMENT IN THE LEFT MAIN TRUNK (LMT). MULTIPLE GUIDE WIRES WERE USED TO RE-CROSS INTO THE STENT IN THE LCX, WITHOUT SUCCESS. THE PROCEDURE WAS THEN ABORTED. THE WIRE FRAGMENT WAS LEFT IN THE PATIENT ANATOMY. THE PATIENT HAD NO CHEST PAIN AND THE ELECTROCARDIOGRAM (ECG) WAS BASELINE THROUGHOUT THE PROCEDURE. IT WAS INFORMED THAT THE PATIENT WAS STABLE AFTER THE PROCEDURE. ANOTHER PCI WILL BE PERFORMED TO REMOVE THE WIRE FRAGMENT ONCE THE PATIENT'S KIDNEY FUNCTION HAS RETURNED TO NORMAL.
Patient problems: Foreign Body In Patient
MAUDE report 23000994 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page