FUJIFILM EB-530TRecalls & reported problems
Fujifilm Corporation · The device is intended for the observation, diagnosis, and endoscopic treatment of trachea and bronchial tree. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04547410330151), received between Jan 11, 2018 and Jan 2, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem2
- Microbial Contamination of Device2
- Break1
- Fracture1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Airway Obstruction1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 2, 2024Break; Fracture; Break; Fracture
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EB-530T. IT WAS REPORTED THAT THE LENS IS BROKEN. IN ADDITION, AIRWAY ENDO TRACHEAL TUBE BECAME CLOGGED WITH BLOOD. ENDOTRACHEAL TUBE CHANGED WITH A 7.5 SIZE. PROVIDER FORCED SCOPE IN A TUBE THAT WAS TOO SMALL.
Patient problems: Airway Obstruction; Hemorrhage/Blood Loss/Bleeding
MAUDE report 18431995 ↗MalfunctionJan 11, 2018Microbial Contamination of Device; Contamination /Decontamination Problem; Microbial Contamination of Device; Contamination /Decontamination Problem
FUJIFILM MEDICAL SYSTEMS U.S.A., INC. (FMSU) WAS INFORMED ON OCTOBER 25, 2017 BY A SERVICE PARTNER THAT A HOSPITAL HAD EXPERIENCED SEVERAL PSEUDOMONAS INFECTIONS IN THEIR RESPIRATORY DEPARTMENT AND THE DOCTORS WERE CONCERNED THEY COULD BE FROM ENDOSCOPES, ALTHOUGH THE SOURCE OF THE INFECTION HAD NOT BEEN DETERMINED. FUJIFILM ATTEMPTED TO CONTACT THE HOSPITAL AT THAT TIME, BUT THE HOSPITAL DID NOT RESPOND TO THE INQUIRY. ON (B)(6) 2017, FMSU WAS INFORMED BY SERVICE PARTNER THAT THERE WAS A SECOND INCIDENT OF A CERTAIN FUJIFILM BRONCHOSCOPE BEING USED RESULTING IN POSITIVE PATIENT BAL SAMPLE. ON JANUARY 5, 2018 FMSU WAS INFORMED BY THE HOSPITAL THAT ENDOSCOPE EB530T 1B084G148 WAS NOT INVOLVED IN THE SECOND INCIDENT REPORTED ON (B)(6) 2017 BUT WAS THE SUBJECT SCOPE REPORTED ON (B)(6) 2017. ENDOSCOPE EB530T 1B084G148 WAS USED ON (B)(6) 2017 TO OBTAIN PATIENT BAL SAMPLES WHICH RESULTED IN POSITIVE PSEUDOMONAS CULTURES. IN ADDITION, THE SCOPE EB530T 1B084G148 WAS NOTED TO HAVE EXCESSIVE AMOUNT OF FLUID COMING FROM SUCTION CHANNEL WHEN PROCESSING THE ENDOSCOPE, THEREFORE IT WAS REMOVED FROM SERVICE ON OR AROUND (B)(6) AND WAS THEN SENT TO A FUJIFILM AUTHORIZED SERVICE FACILITY FOR INSPECT
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7184539 ↗MalfunctionJan 11, 2018Microbial Contamination of Device; Contamination /Decontamination Problem; Microbial Contamination of Device; Contamination /Decontamination Problem
FUJIFILM MEDICAL SYSTEMS U.S.A., INC. (FMSU) WAS INFORMED ON OCTOBER 25, 2017 BY A SERVICE PARTNER THAT A HOSPITAL HAD EXPERIENCED SEVERAL PSEUDOMONAS INFECTIONS IN THEIR RESPIRATORY DEPARTMENT AND THE DOCTORS WERE CONCERNED THEY COULD BE FROM ENDOSCOPES, ALTHOUGH THE SOURCE OF THE INFECTION HAD NOT BEEN DETERMINED. FUJIFILM ATTEMPTED TO CONTACT THE HOSPITAL AT THAT TIME, BUT THE HOSPITAL DID NOT RESPOND TO THE INQUIRY. ON DECEMBER 12, 2017, FMSU WAS INFORMED BY A SERVICE PARTNER THAT THE HOSPITAL HAD USED THE ENDOSCOPE (EB530T SERIAL (B)(4)) TO OBTAIN PATIENT BAL SAMPLES WHICH RESULTED IN POSITIVE PSEUDOMONAS CULTURES. THE FIRST INSTANCE OF THE SUBJECT ENDOSCOPE SCOPE BAL SAMPLE RESULTING POSITIVE OCCURRED ON (B)(6) 2017. DURING THE (B)(6) EPISODE THE ENDOSCOPE WAS SUBSEQUENTLY TESTED FOR CONTAMINANTS AND RESULTED POSITIVE FOR PSEUDOMONAS. THE SUBJECT ENDOSCOPE WAS THEN SENT TO A FUJIFILM AUTHORIZED SERVICE FACILITY FOR INSPECTION AND REPAIR. THE SERVICE FACILITY COMPLETED THE REPAIR AND RETURNED THE SCOPE TO THE HOSPITAL ON NOVEMBER 29, 2017. THE ENDOSCOPE WAS REPROCESSED AND ON (B)(6) 2017, WAS USED AND RESULTED IN POSITIVE PATIENT BAL SAMPLES FOR PSEUDOMONAS. ALTHOUGH POSITIVE
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7182043 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page