FDR-2000BPY POSITIONER KIT BPY-L HA USA ERecalls & reported problems
Fujifilm Corporation · The FDR-2000BPY provides the function that calculates the three-dimensional coordinates of the image obtained from an exposure and supports the positioning of the needle using the coordinates. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547410336047), received between Jul 19, 2019 and Sep 9, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K133972, K191495); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Malposition of Device2
Patient problems
- Perforation2
- Skin Tears2
- Tissue Damage1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 9, 2021Malposition of Device; Malposition of Device
FUJIFILM WAS INFORMED THAT DURING A BIOPSY EXAM A NEEDLE MISPUNCTURE OCCURRED IN GERMANY. A BIOPSY ON THE PATIENT WAS PERFORMED AND THE NEEDLE DROVE TOO DEEP INTO THE PATIENT'S BREAST, PASSING THROUGH THE CALCIFICATION (Z AXIS STROKE TOO LONG). THE (VACUUM) NEEDLE PENETRATED THE CAUDAL SKIN OF THE PATIENT. THE NEEDLE WAS PULLED BACK INTO THE POSITION OF THE CALCIFICATION AND THE OPERATOR (PHYSICIAN) WAS ABLE TO EXTRACT A BIOPSY SAMPLE. THE EXAM COULD BE FINALIZED AND THE PATIENT WAS SENT HOME AFTER TREATING THE CUT WITH AN ADHESIVE BANDAGE. THERE WAS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT; THIS INCIDENT IS BEING REPORTED IN AN ABUNDANCE OF CAUTION. THIS EVENT WAS REPORTED TO THE MANUFACTURER ON (B)(6) 2019. THE MANUFACTURER REPORTED THIS EVENT TO FDA UNDER REPORT # 3001722928-2019-00009.
Patient problems: Perforation; Skin Tears
MAUDE report 12445821 ↗MalfunctionJul 19, 2019Malposition of Device; Malposition of Device
FUJIFILM WAS INFORMED THAT DURING A BIOPSY EXAM A NEEDLE MISPUNCTURE OCCURRED IN (B)(6). PROCEDURE 1, PATIENT 1: PRE-SHOTS, USED FOR CALCULATING THE LOCATION OF THE MICROCALC AND BIOPSY POSITIONER / NEEDLE POSITION, WERE PERFORMED WITHOUT PROBLEMS. AFTER THE EXPOSURES THE BIOPSY POSITIONER (THE BIOPSY ADD-ON DEVICE HOLDING AND POSITIONING THE NEEDLE) GENERATED AN ERROR MESSAGE "PROCESS COMMUNICATION NOT POSSIBLE" FOLLOWED BY ERROR MESSAGE "ERROR WHILE SAVING TARGET INFO". PATIENT 1 WAS SENT HOME, AND NO BIOPSY WAS PERFORMED. THE OPERATOR REBOOTED THE UNIT AND NO ERROR WAS VISIBLE. THE OPERATOR PERFORMED DAILY ACCURACY CHECK AFTER THE REBOOT WITHOUT ERRORS OR OTHER PROBLEMS. PROCEDURE 2, PATIENT 2: A BIOPSY ON PATIENT 2 WAS PERFORMED AND THE NEEDLE DROVE TOO DEEP INTO THE PATIENT'S BREAST, PASSING THROUGH THE CALCIFICATION (Z AXIS STROKE TOO LONG). THE (VACUUM) NEEDLE PENETRATED THE CAUDAL SKIN OF THE PATIENT. THE NEEDLE WAS PULLED BACK INTO THE POSITION OF THE CALCIFICATION AND THE OPERATOR (PHYSICIAN) WAS ABLE TO EXTRACT A BIOPSY SAMPLE. THE EXAM COULD BE FINALIZED AND THE PATIENT WAS SENT HOME AFTER TREATING THE CUT WITH AN ADHESIVE BANDAGE. THERE WAS NO DEATH ASSOCIATED WITH THI
Patient problems: Perforation; Skin Tears; Tissue Damage
MAUDE report 8808829 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133972, K191495. One submission often covers many models, so check the product and lot details in each record.
- Z-0661-2018 · Aug 4, 2017 · TerminatedASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full
- Z-0386-2018 · Jul 14, 2017 · TerminatedASPIRE Cristalle (FDR MS-3500) Mammography machine The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digit
Other System, X-Ray, Mammographic devices
- Affirm Lateral Arm Upright Biopsy Accessory ASY-03723Hologic, Inc. · System, x-ray, mammographic1 recall3 reports
- Affirm Breast Biopsy Guidance System STLC-00004Hologic, Inc. · System, x-ray, mammographic24 reports
- Affirm Prone Biopsy System PBX-SYS-AFFIRM-3DHologic, Inc. · System, x-ray, mammographic23 reports
- Multicare Platinum 8-004-0017Hologic, Inc. · System, x-ray, mammographic2 reports
- Bella Blankets 311Beekley Corporation · System, X-Ray, Mammographic1 report
- Affirm Prone Biopsy System PBX-SYS-AFFIRM-2DHologic, Inc. · System, x-ray, mammographic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.