Affirm Lateral Arm Upright Biopsy Accessory ASY-03723Recalls & reported problems
Hologic, Inc. · Affirm Lateral Arm Upright Biopsy Accessory · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15420045506817), received between Jun 16, 2017 and Jun 16, 2017 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (3 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 7, 2017Class IITerminated
The firm is implementing a product correction involving its Affirm Lateral Arm Upright Biopsy Accessory, ASY-09880. The correction has been initiated due to a component (the Left-hand Guide or the Blue Needle Guide) issue that could impact needle alignment for all the left-approach lateral biopsy procedures. This may have an impact on the position of the needle within the breast, and thus on the aspiration of the intended tissue. This impact is case-dependent and detectable when imagining is used to confirm needle position and/or tissue acquisition during and after a procedure. There are no known instances of illness or injury.
Z-0351-2018 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Operates Differently Than Expected3
- Failure to Align3
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2017Failure to Align; Device Operates Differently Than Expected; Failure to Align; Device Operates Differently Than Expected
DURING THE INITIAL BIOPSY APPLICATIONS TRAINING (PRIOR TO FIRST PATIENT USE), OUR APPLICATION SPECIALIST VALIDATED THE SPRING LOADED BIOPSY NEEDLE (BARD MAGNUM, NON VACUUM ASSISTED BIOPSY DEVICE). TO DO THIS THEY FOLLOWED A DEFINED PROCEDURE USING A TARGETING PHANTOM (PHANTOM WITH NEEDLE TIP INSIDE). AT THE END OF THE TEST, THE TIP OF THE BIOPSY NEEDLE NEEDS TO BE ALIGNED WITH THE NEEDLE TIP OF THE TARGETING PHANTOM. DURING THIS VALIDATION PROCESS THEY HAVE NOTICED THE BIOPSY NEEDLE WAS NOT ALIGNED WITH THE TARGETING PHANTOM. IT WAS OFF FOR APPROX. 2MM WHEN THE LATERAL ARM WAS INSTALLED ON THE LEFT HAND SIDE (BLUE SIDE).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6648560 ↗MalfunctionJun 16, 2017Failure to Align; Device Operates Differently Than Expected; Failure to Align; Device Operates Differently Than Expected
DURING THE INITIAL BIOPSY APPLICATIONS TRAINING (PRIOR TO FIRST PATIENT USE), OUR APPLICATION SPECIALIST VALIDATED THE SPRING LOADED BIOPSY NEEDLE (BARD MAGNUM, NON VACUUM ASSISTED BIOPSY DEVICE). TO DO THIS THEY FOLLOWED A DEFINED PROCEDURE USING A TARGETING PHANTOM (PHANTOM WITH NEEDLE TIP INSIDE). AT THE END OF THE TEST, THE TIP OF THE BIOPSY NEEDLE NEEDS TO BE ALIGNED WITH THE NEEDLE TIP OF THE TARGETING PHANTOM. DURING THIS VALIDATION PROCESS THEY HAVE NOTICED THE BIOPSY NEEDLE WAS NOT ALIGNED WITH THE TARGETING PHANTOM. IT WAS OFF FOR APPROX. 2MM WHEN THE LATERAL ARM WAS INSTALLED ON THE LEFT HAND SIDE (BLUE SIDE).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6648550 ↗MalfunctionJun 16, 2017Failure to Align; Device Operates Differently Than Expected; Failure to Align; Device Operates Differently Than Expected
DURING THE INITIAL BIOPSY APPLICATIONS TRAINING (PRIOR TO FIRST PATIENT USE), OUR APPLICATION SPECIALIST VALIDATED THE SPRING LOADED BIOPSY NEEDLE (BARD MAGNUM, NON VACUUM ASSISTED BIOPSY DEVICE). TO DO THIS THEY FOLLOWED A DEFINED PROCEDURE USING A TARGETING PHANTOM (PHANTOM WITH NEEDLE TIP INSIDE). AT THE END OF THE TEST, THE TIP OF THE BIOPSY NEEDLE NEEDS TO BE ALIGNED WITH THE NEEDLE TIP OF THE TARGETING PHANTOM. DURING THIS VALIDATION PROCESS THEY HAVE NOTICED THE BIOPSY NEEDLE WAS NOT ALIGNED WITH THE TARGETING PHANTOM. IT WAS OFF FOR APPROX. 2MM WHEN THE LATERAL ARM WAS INSTALLED ON THE LEFT HAND SIDE (BLUE SIDE).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6648534 ↗
Other System, x-ray, mammographic devices
- Affirm Breast Biopsy Guidance System STLC-00004Hologic, Inc. · System, x-ray, mammographic24 reports
- Affirm Prone Biopsy System PBX-SYS-AFFIRM-3DHologic, Inc. · System, x-ray, mammographic23 reports
- Multicare Platinum 8-004-0017Hologic, Inc. · System, x-ray, mammographic2 reports
- FDR-2000BPY POSITIONER KIT BPY-L HA USA EFujifilm Corporation · System, X-Ray, Mammographic2 reports
- Bella Blankets 311Beekley Corporation · System, X-Ray, Mammographic1 report
- Affirm Prone Biopsy System PBX-SYS-AFFIRM-2DHologic, Inc. · System, x-ray, mammographic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page