FUJIFILM EB-580SRecalls & reported problems
Fujifilm Corporation · The device is intended for the observation, diagnosis, and endoscopic treatment of trachea and bronchial tree · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 1
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04547410400380), received between Mar 25, 2026 and Sep 18, 2026 — classified as 1 death, 7 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Contamination3
- Image Display Error/Artifact1
- No Display/Image1
Patient problems
- Bacterial Infection3
- Fungal Infection2
- Bronchial Hemorrhage1
- Hemorrhage/Blood Loss/Bleeding1
- Unspecified Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
DeathSep 18, 2026No Display/Image
It was reported that visualization from EB-580S was lost during a procedure. The procedure treatment was delayed to bleeding in a lung. It took ten minutes to find a different room and to get different scope.Per customer, the patient died a few hours after the incident on 8/27 due to bleeding in the lung. The visualization was lost on reentry of the scope. Patient was intubated prior to start of procedure. The procedure was not completed after replacing the scope. The bleeding being treated was not the result of the device.
Patient problems: Bronchial Hemorrhage
MAUDE report 50815166 ↗InjurySep 16, 2026Image Display Error/Artifact
On 08/27/2026, FUJIFILM Healthcare Americas Corporation became aware of an event involving EB-580S. It was reported that visualization was lost during a procedure. The procedure treatment was delayed to bleeding in a lung. It took ten minutes to find a different room and to get different scope. Per customer, the patient died a few hours after the incident on (b)(6) 2026 due to bleeding in the lung. The visualization was lost on reentry of the scope. Patient was intubated prior to start of procedure. The procedure was not completed after replacing the scope. The bleeding being treated was not the result of the device. FUJIFILM Healthcare Americas Corporation is reporting this event, however the cause of death due to existing condition. Ref: FUJIFILM Healthcare Americas Corporation Complaint # (b)(4).
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50782620 ↗InjuryApr 3, 2026Break
IT WAS REPORTED THAT THE CUSTOMER WANTS THREE BRONCHOSCOPES EVALUATION TO BE PERFORMED DUE TO PATIENT(S) INFECTION EVENT. THE CUSTOMER CONFIRMED THERE WERE THREE PATIENTS INVOLVED AND POSITIVE CANDIDA WAS CONFIRMED.
Patient problems: Bacterial Infection
MAUDE report 24777001 ↗InjuryApr 3, 2026Break
IT WAS REPORTED THAT THE CUSTOMER WANTS THREE BRONCHOSCOPES EVALUATION TO BE PERFORMED DUE TO PATIENT(S) INFECTION EVENT. THE CUSTOMER CONFIRMED THERE WERE THREE PATIENTS INVOLVED AND POSITIVE CANDIDA WAS CONFIRMED.
Patient problems: Bacterial Infection
MAUDE report 24777000 ↗InjuryApr 3, 2026Break
IT WAS REPORTED THAT THE CUSTOMER WANTS THREE BRONCHOSCOPES EVALUATION TO BE PERFORMED DUE TO PATIENT(S) INFECTION EVENT. THE CUSTOMER CONFIRMED THERE WERE THREE PATIENTS INVOLVED AND POSITIVE CANDIDA WAS CONFIRMED.
Patient problems: Bacterial Infection
MAUDE report 24776999 ↗InjuryMar 25, 2026Contamination
ON 03/02/2026, FUJIFILM HEALTHCARE AMERICAS CORPORTATION BECAME AWARE OF AN EVENT INVOLVING EB-580S. THE CUSTOMER REPORTED A PATIENT INFECTION EVENT. THE PATIENT INFECTION ORGANISM WAS NOT IDENTIFIED. THE CUSTOMER IS UNABLE TO CONFIRM IF THE PATIENT REQUIRED ADDITIONAL MEDICAL TREATMENT TO TREAT THE INFECTION. DUE TO THE PATIENT INFECTION, FUJIFILM HEALTHCARE AMERICAS CORPORTATION IS REPORTING THIS EVENT IN AN ABDUNANCE OF CAUTION. REF: FUJIFILM HEALTHCARE AMERICAS CORPORTATION COMPLAINT #(B)(4).
Patient problems: Unspecified Infection
MAUDE report 24692429 ↗InjuryMar 25, 2026Contamination
ON 03/02/2026, FUJIFILM HEALTHCARE AMERICAS CORPORTATION BECAME AWARE OF AN EVENT INVOLVING EB-580S. THE CUSTOMER REPORTED A PATIENT INFECTION EVENT. THE PATIENT INFECTION WAS DUE TO YEAST- CANDIDA ALBICANS. THE CUSTOMER IS UNABLE TO CONFIRM IF THE PATIENT REQUIRED ADDITIONAL MEDICAL ATTENTION TO TREAT THE INFECTION. DUE TO THE PATIENT INFECTION, FUJIFILM HEALTHCARE AMERICAS CORPORTATION IS REPORTING THIS EVENT IN AN ABDUNANCE OF CAUTION. REF: FUJIFILM HEALTHCARE AMERICAS CORPORTATION COMPLAINT # (B)(4).
Patient problems: Fungal Infection
MAUDE report 24691946 ↗InjuryMar 25, 2026Contamination
ON 03/02/2026, FUJIFILM HEALTHCARE AMERICAS CORPORATION BECAME AWARE OF AN EVENT INVOLVING EB-580S. THE CUSTOMER REPORTED A PATIENT INFECTION EVENT. THE PATIENT INFECTION WAS DUE TO YEAST - CANDIDA ALBICANS. THE CUSTOMER IS UNABLE TO CONFIRM IF THE PATIENT REQUIRED ADDITIONAL MEDICAL ATTENTION TO TREAT THE IDENTIFIED INFECTION. DUE TO THE PATIENT INFECTION, FUJIFILM HEALTHCARE AMERICAS CORPORTATION IS REPORTING THIS EVENT IN AN ABDUNANCE OF CAUTION. REF: FUJIFILM HEALTHCARE AMERICAS CORPORATION COMPLAINT # (B)(4).
Patient problems: Fungal Infection
MAUDE report 24691676 ↗
Other Bronchoscope (Flexible Or Rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.