Skip to content

FUJIFILM EB-580TRecalls & reported problems

Fujifilm Corporation · The device is intended for the observation, diagnosis, and endoscopic treatment of trachea and bronchial tree · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04547410400397), received between Sep 6, 2023 and Apr 15, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Stretched” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Stretched2
  • Break1
  • Contamination /Decontamination Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Bacterial Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 15, 2026Break

    THE TIP OF THE BRONCHOSCOPE CAME OFF AND FELL INTO THE PATIENT'S AIRWAY. THE PROCEDURE WAS COMPLETED AFTER SWITCHING OUT THE SCOPES. THERE WAS A 10-MINUTE DELAY TO REMOVE THE PIECE THAT FELL OFF AND TO SWITCH THE SCOPES. NO DELAYS THEREAFTER TO COMPLETE THE PROCEDURE.THE PATIENT DID NOT EXPERIENCE ANY HARM, INJURY, OR COMPLICATIONS AFTER THE PROCEDURE. NO ADDITIONAL TREATMENT OR MONITORING WAS REQUIRED. THE PATIENT RECOVERED AND WAS DISCHARGED HOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24882020 ↗
  • InjuryMar 20, 2026Detachment of Device or Device Component

    ON 02/24/2026, FUJIFILM HEALTHCARE AMERICAS CORPORATION BECAME AWARE OF AN EVENT INVOLVING EB-580T. IT WAS REPORTED THAT THE TIP OF THE BRONCHOSCOPE CAME OFF AND FELL INTO THE PATIENT'S AIRWAY. CUSTOMER WAS UNABLE TO USE THE SCOPE. CUSTOMER HAD TO RETRIEVE THE DISTAL TIP OUT OF THE AIRWAY OF THE PATIENT THEN SWITCH SCOPES. THE PATIENT HAS RECOVERED AND WAS DISCHARGED HOME. NO ADDITIONAL TREATMENT OR MONITORING WAS REQUIRED. THE PATIENT DID NOT EXPERIENCE ANY HARM, INJURY, OR COMPLICATIONS AFTER THE PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER SWITCHING THE SCOPE. THERE WAS A 10-MINUTE DELAY TO REMOVE THE PIECE OF THE SCOPE THAT FELL OFF AND SWITCH THE SCOPES. THE PATIENT REMAINED UNDER ANESTHESIA DURING THE DELAY. THE DELAY DID NOT IMPACT THE PATIENT OR PROCEDURE. DUE TO A PIECE OF THE DEVICE FALLING INTO THE PATIENT'S CAVITY, FUJIFILM HEALTHCARE AMERICAS CORPORATION IS REPORTING THIS EVENT IN AN ABDUNANCE OF CAUTION. REF: FUJIFILM HEALTHCARE AMERICAS CORPORATION COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651042 ↗
  • MalfunctionMar 7, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ON (B)(6) 2024, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EB-580T. IT WAS REPORTED THAT THERE IS A CONCERN FOR POTENTIAL DEFECT INSIDE THE SCOPE THAT COULD BE A SOURCE FOR HARBORING BACTERIA. THREE PATIENTS TEST POSITIVE WITH A RARE BACTERIUM AND ARE TRYING TO IDENTIFY A SOURCE.

    Patient problems: Bacterial Infection

    MAUDE report 18861833 ↗
  • MalfunctionSep 7, 2023Stretched; Stretched

    ON (B)(6) 2023, FUJIFILM CORPORATION WAS RECEIVED ADDITIONAL INFORMATION OF AN EVENT INVOLVING EB-580T. PRIOR TO THE ADDITIONAL INFORMATION, FUJIFILM CORPORATION DETERMINED THE INCIDENT NON-REPORTABLE. IT WAS REPORTED THAT DURING PROCEDURE USING A COMBINATION OF A BRONCHOSCOPE EB-580T AND AN ENDOTRACHEAL TUBE AT THE FACILITY, WHEN THE EB-580T WAS WITHDRAWN THROUGH THE ENDOTRACHEAL TUBE, THE EB-580T WAS DIFFICULT TO WITHDRAW DUE TO THE BENDING RUBBER MIGRATING TO THE DISTAL END. THE EB-580T WAS REMOVED ALONG WITH THE ENDOTRACHEAL TUBE AND REPLACED WITH ANOTHER SCOPE, AND THE PROCEDURE WAS CONTINUED AND COMPLETED SUCCESSFULLY. THE PATIENT HAS NOT COMPLAINED OF ANY HEALTH HAZARDS. THERE IS NO SERIOUS INJURY AND DEATH OCCURRED ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17699327 ↗
  • MalfunctionSep 6, 2023Stretched; Stretched

    ON AUGUST,17, 2023, FUJIFILM CORPORATION WAS RECEIVED ADDITIONAL INFORMATION OF AN EVENT INVOLVING EB-580T. PRIOR TO THE ADDITIONAL INFORMATION, FUJIFILM CORPORATION DETERMINED THE INCIDENT NON-REPORTABLE. IT WAS REPORTED THAT DURING PROCEDURE USING A COMBINATION OF A BRONCHOSCOPE EB-580T AND AN ENDOTRACHEAL TUBE AT THE FACILITY, WHEN THE EB-580T WAS WITHDRAWN THROUGH THE ENDOTRACHEAL TUBE, THE EB-580T WAS DIFFICULT TO WITHDRAW DUE TO THE BENDING RUBBER MIGRATING TO THE DISTAL END. THE EB-580T WAS REMOVED ALONG WITH THE ENDOTRACHEAL TUBE AND REPLACED WITH ANOTHER SCOPE, AND THE PROCEDURE WAS CONTINUED AND COMPLETED SUCCESSFULLY. THE PATIENT HAS NOT COMPLAINED OF ANY HEALTH HAZARDS. THERE IS NO SERIOUS INJURY AND DEATH OCCURRED ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17691248 ↗

Other Bronchoscope (Flexible Or Rigid) devices

All Bronchoscope (Flexible Or Rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page