HITACHI Aloka Arietta 850Recalls & reported problems
Hitachi, Ltd. · Diagnostic Ultrasound System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04562122958974), received between Apr 20, 2022 and May 5, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 14, 2021Class IIOpen, Classified
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Z-0190-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Display/Image3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2022No Display/Image; No Display/Image
ON (B)(6) 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT REGARDING THE ARIETTA 850 ULTRASOUND SYSTEM. THE SITE REPORTED THAT THE SCREEN BLACKED OUT DURING A PROCEDURE. THE INCIDENT RESULTED IN A MINOR DELAY, HOWEVER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING THE SAME DEVICE WITHOUT HARM TO THE PATIENT. THERE IS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14287387 ↗MalfunctionApr 27, 2022No Display/Image; No Display/Image
ON MARCH 31, 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT REGARDING THE ARIETTA 850 ULTRASOUND SYSTEM. THE SITE REPORTED THAT THE SCREEN BLACKED OUT DURING THE PROCEDURE. THERE IS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14220628 ↗MalfunctionApr 20, 2022No Display/Image; No Display/Image
ON (B)(6) 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT REGARDING THE ARIETTA 850 ULTRASOUND SYSTEM. THE SITE REPORTED THAT DURING THE EXAM, THE ARIETTA 850 SCREEN WENT BLANK WITH NOTHING ON THE TOUCH SCREEN AND THE OPERATION PANEL LIGHTS STARTED TO FLASH LIKE IT WAS REBOOTING. THE PATIENT WAS UNDER ANESTHESIA AT THIS TIME. THE SITE CHANGED TO THE OTHER ARIETTA 850 AND WERE ABLE TO COMPLETE THE PROCEDURE. THERE WAS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14163658 ↗
Other System, Imaging, Pulsed Echo, Ultrasonic devices
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed echo, ultrasonic1 recall8 deaths5,133 reports
- Site~Rite 8 Ultrasound System 9770600Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic629 reports
- Site-Rite 9770501Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic546 reports
- Transducer X8-2t 989605455171Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic8 recalls19 reports
- Transducer C10-3v 989605409562Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic7 recalls0 reports
- Site~Rite 8 Ultrasound System 9770600YBard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic330 reports
- Transducer C9-2 989605391092Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
- Transducer S5-1 989605412082Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page