Arietta 850 Ultrasound recallZ-0190-2022 · FujiFilm Healthcare Americas Corporation · initiated Oct 14, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from FujiFilm Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
FDA root cause category: Software design
What the company is doing
Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21. Letter states reason for recall, health risk and action to take: Customers/Users are advised to accept the corrected software and manual as soon as available. Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems. We will contact you to set up time to update the software and provide the revised manual. (7) Other Information: For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email [email protected]
Recall details
- Recall number
- Z-0190-2022
- Status
- Open, Classified
- Initiated by the company
- Oct 14, 2021
- Recalling firm
- FujiFilm Healthcare Americas Corporation
- Quantity
- 38 units
- Product code
- IYN
- FDA submissions
- K171708, K183456, K173739
- Distribution
- Nationwide
Product description
Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Serial Numbers: 205W3386 (01)04562122958974(11)171114(21)205W3386; 206F7900 (01)04562122958974(11)180926(21)206F7900; 206F7901 (01)04562122958974(11)180926(21)206F7901; 206F7936 (01)04562122958974(11)181113(21)206F7936; 206F7937 (01)04562122958974(11)181113(21)206F7937; 206F7940 (01)04562122958974(11)181113(21)206F7940; 206F7941 (01)04562122958974(11)181113(21)206F7941; 206F7942 (01)04562122958974(11)181113(21)206F7942; 206F7944 (01)04562122958974(11)181113(21)206F7944; 206F7945 (01)04562122958974(11)181113(21)206F7945; 206F7946 (01)04562122958974(11)181113(21)206F7946; 206F7947 (01)04562122958974(11)181114(21)206F7947; G3002552 (01)04562122958974(11)20190115(21)G3002552; G3024545 (01)04562122958974(11)190607(21)G3024545; G3037885 (01)04562122958974(11)20190719(21)G3037885; G3037887 (01)04562122958974(11)20190719(21)G3037887; G3042405 (01)04562122958974(11)190829(21)G3042405; G3042823 (01)04562122958974(11)20190807(21)G3042823; G3042828 (01)04562122958974(11)20190808(21)G3042828; G3045094 (01)04562122958974(11)20190822(21)G3045094; G3052105 (01)04562122958974(11)20191008(21)G3052105 G3052118 (01)04562122958974(11)20191025(21)G3052118 G3055549 (01)04562122958974(11)20191118(21)G3055549 G3055550 (01)04562122958974(11)20191118(21)G3055550 G3055552 (01)04562122958974(11)20191118(21)G3055552 G3055553 (01)04562122958974(11)20191118(21)G3055553 G3055554 (01)04562122958974(11)20191118(21)G3055554 G3055555 (01)04562122958974(11)20191118(21)G3055555 G3060234 (01)04562122958974(11)20191206(21)G3060234 G3060254 (01)04562122958974(11)20191206(21)G3060254 G3060275 (01)04562122958974(11)20191216(21)G3060275 G3060278 (01)04562122958974(11)20191216(21)G3060278 G3060690 (01)04562122958974(11)200127(21)G3060690 G3071074 (01)04562122958974(11)20200227(21)G3071074 G3080785 (01)04562122958974(11)201105(21)G3080785 G3090450 (01)04562122958974(11)201225(21)G3090450 G3110884 (01)04562122958974(11)210401(21)G3110884
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HITACHI Aloka Arietta 850Hitachi, Ltd. · System, Imaging, Pulsed Echo, Ultrasonic1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88840).
- Z-0189-2022 · Open, ClassifiedArietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3
- Z-0191-2022 · Open, ClassifiedLisendo 880 Ultrasound Software Version: V1.0 through V4.1.3
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page