HITACHI NoblusRecalls & reported problems
Hitachi, Ltd. · Ultrasound Diagnostic Scanner · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04580292771704), received between Oct 18, 2021 and Mar 8, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Unintended Application Program Shut Down” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Apr 11, 2024Open, Classified
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Z-1884-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended Application Program Shut Down2
- Application Program Problem1
- Computer Software Problem1
Patient problems
- Insufficient Information2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 8, 2022Unintended Application Program Shut Down; Unintended Application Program Shut Down
ON (B)(6) 2021, FUJIFILM HEALTHCARE AMERICAS CORPORATION (FORMERLY HITACHI HEALTHCARE AMERICAS CORPORATION) RECEIVED A COMPLAINT REGARDING NOBLUS ULTRASOUND. THE SITE REPORTED THAT A PATIENT WAS UNDER ANESTHESIA AND UNDERGOING A PROSTATE CRYO PROCEDURE WITH A BI-PLACE C41L47RP PROBE. SITE TECHNICIAN TURNED ON THE NOBLUS SYSTEM AND ATTACHED THE PROBE. THE SYSTEM WOULD NOT BOOT UP READY AND REPORTED AN ERROR ASKING TO "PLEASE WAIT UNTIL THE ERROR LOG IS COMPLETE." A PROMPT WOULD POP UP WITH THE OPTION TO HIT "OK", AND WHEN HE DID, THE SYSTEM WOULD SHUT DOWN. TECH ATTEMPTED TO REBOOT SEVERAL TIMES, BOTH WITH AND WITHOUT A PROBE ATTACHED, AND EVEN TRIED TO REPLACE THE BATTERY. ERROR PERSISTED. THE PROCEDURE WAS RESCHEDULED. THERE WERE NO SIDE EFFECTS FROM THE ANESTHESIA AND THE PATIENT WAS CLEARED TO GO HOME. THERE WAS NO INJURY/HARM OR DEATH REPORTED WITH THIS EVENT. DURING A RETROSPECTIVE REVIEW, IT WAS DISCOVERED THAT THE MANUFACTURER HAD ASSESSED THIS EVENT AS AE LEADING TO SERIOUS INJURY; THEREFORE AN IMPORTER'S MDR IS BEING SUBMITTED.
Patient problems: Insufficient Information
MAUDE report 13686566 ↗InjuryOct 18, 2021Computer Software Problem; Application Program Problem; Unintended Application Program Shut Down; Computer Software Problem; Application Program Problem; Unintended Application Program Shut Down
ON SEPTEMBER 20 2021, FUJIFILM HEALTHCARE AMERICAS RECEIVED A REPORT OF AN ABORTED PROCEDURE WHILE PATIENT WAS UNDER ANESTHESIA. THE SITE WAS USING THE NOBLUS ULTRASOUND UN0581.
Patient problems: Insufficient Information
MAUDE report 12647907 ↗
Other System, Imaging, Pulsed Echo, Ultrasonic devices
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed echo, ultrasonic1 recall8 deaths5,133 reports
- Site~Rite 8 Ultrasound System 9770600Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic629 reports
- Site-Rite 9770501Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic546 reports
- Transducer X8-2t 989605455171Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic8 recalls19 reports
- Transducer C10-3v 989605409562Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic7 recalls0 reports
- Site~Rite 8 Ultrasound System 9770600YBard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic330 reports
- Transducer C9-2 989605391092Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
- Transducer S5-1 989605412082Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page