Noblus AC Adapter used with the Noblus Ultrasound Imaging… recallZ-1884-2024 · FUJIFILM Healthcare Americas Corporation · initiated Apr 11, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FUJIFILM Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

FDA root cause category: Device Design

What the company is doing

FujiFilm issued Urgent Medical Device Recall Letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: Please complete and return the Field Action Verification Form attached on page 3 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Please accept the replacement of AC adapter that is implemented the countermeasure. Please contact our field action coordinator, Kotei Aoki, via email at [email protected], if you have any further questions regarding this field action.

Recall details

Recall number
Z-1884-2024
Status
Open, Classified
Initiated by the company
Apr 11, 2024
Posted by the FDA
May 22, 2024
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
418 units
Product code
IYN
FDA submissions
K160559
Distribution
US Nationwide distribution.

Product description

Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

Code information (models, lots, serial numbers)

UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.