Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04711605587570), received between May 17, 2024 and Jul 18, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Disintegration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Disintegration3
- No Apparent Adverse Event3
- Scratched Material3
- Structural Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2024Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event; Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event
THE CALCAR PLANER INSTRUMENT (PART NUMBER 9110-3002) AND BROACH #4 (PART NUMBER 9110-6004-RA) WERE USED DURING A HIP SURGERY ON (B)(6) 2024 AT (B)(6) BY THE SURGEON. THE SURGEON COMPLAINED THAT THE CALCAR PLANNER'S AND BROACH#4 WERE NOT CONNECTING AS INTENDED. SURGEON THEN USED A SMALLER SIZE CALCAR PLANNER TO CONNECT TO THE BROACH AND COMPLETED THE CASE. METAL SHAVING DEBRIS DID NOT COME OUT OF THE INSTRUMENT, HOWEVER, WHEN VISUALLY LOOKING INSIDE THE CALCAR PLANNER, METAL PEALING WAS OBSERVED. THE SURGEON CLOSELY INSPECTED THE WOUND TO CONFIRM THERE WAS NO METAL RESIDUE IN THE PATIENT AND THE BROKEN INSTRUMENT. THE INSTRUMENTS HAD ALREADY PERFORMED THEIR FUNCTION AND THERE WAS NO FURTHER INCIDENT. THE INSTRUMENTS WERE RETRIEVED ACCORDING TO COMPLAINT HANDLING PROCEDURES FOR EVALUATION AFTER THE CASE COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19768582 ↗MalfunctionJun 26, 2024Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event; Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event
THE CALCAR PLANER INSTRUMENT (PART NUMBER 9110-3003 AND 9110-3002) USED DURING A HIP SURGERY ON (B)(6) 2024 AT THE SURGICAL INSTITUTE OF READING BY THE SURGEON BROKE WHILE GRINDING AND TORQUING THE BROACH TRIAL.METAL SHAVINGS WERE FOUND INSIDE THE INSTRUMENT, THERE WAS NO EVIDENCE OF THEM BEING LEFT IN THE PATIENT'S BODY.THE SURGEON COMPLETED THE CASE USING ANOTHER SIZE CALCAR PLANER, AND THE BROKEN INSTRUMENT WAS RETRIEVED AFTER THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19612011 ↗InjuryMay 17, 2024Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event; Material Disintegration; Structural Problem; Scratched Material; No Apparent Adverse Event
THE CALCAR PLANER INSTRUMENT (PART NUMBER 9110-3003 AND 9110-3002) USED DURING A HIP SURGERY ON (B)(6) 2024 AT THE SURGICAL INSTITUTE OF READING BY THE SURGEON BROKE WHILE GRINDING AND TORQUING THE BROACH TRIAL. METAL SHAVINGS WERE FOUND INSIDE THE INSTRUMENT, THERE WAS NO EVIDENCE OF THEM BEING LEFT IN THE PATIENT'S BODY. THE SURGEON COMPLETED THE CASE USING ANOTHER SIZE CALCAR PLANER, AND THE BROKEN INSTRUMENT WAS RETRIEVED AFTER THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19347048 ↗
Other Reamer devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page