U2 Total Knee System 9304-4103-RERecalls & reported problems

United Orthopedic Corp. · PS Housing impactor · Class 1

All models of U2 Total Knee System (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04719886890457), received between May 8, 2024 and Jun 26, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component4
  • Material Deformation4
  • Material Separation4
  • No Apparent Adverse Event4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2024Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event; Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event

    THE SURGEON HAD PREPARED THE FEMORAL POSTERIOR STABILIZED (PS) NOTCH, WHICH IS A SPECIFIC AREA ON THE FEMUR BONE WHERE THE PS COMPONENT OF A TOTAL KNEE REPLACEMENT IS POSITIONED. DURING THE PROCEDURE, THE SURGEON REMOVED THE PS IMPACTOR, AND IT WAS NOTED THAT THE INSTRUMENT ASSEMBLY HAD BECOME DAMAGED. THIS TYPE OF DAMAGE TO A SURGICAL INSTRUMENT IS NOT UNCOMMON AND CAN OCCUR DUE TO THE FORCES AND STRESSES INVOLVED IN THE BONE PREPARATION PROCESS. DESPITE THE DAMAGE TO THE INSTRUMENT, THE SURGEON WAS ABLE TO COMPLETE THE SURGERY WITHOUT ANY DELAYS OR COMPLICATIONS. THIS WAS POSSIBLE BECAUSE THE PS NOTCH HAD ALREADY BEEN SUCCESSFULLY PREPARED PRIOR TO THE INSTRUMENT DAMAGE. THE DAMAGED INSTRUMENT HAD ALREADY SERVED ITS PURPOSE, AND THE SURGEON WAS ABLE TO PROCEED WITH THE REMAINDER OF THE SURGICAL STEPS. THE SURGERY WAS COMPLETED WITHOUT ANY ISSUES OR DELAYS, AS THE INSTRUMENT HAD ALREADY PERFORMED ITS INTENDED FUNCTION OF PREPARING THE PS NOTCH PRIOR TO THE DAMAGE OCCURRING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19612013 ↗
  • MalfunctionJun 26, 2024Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event; Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event

    DURING THE PREPARATION FOR THE POSTERIOR STABILIZATION (PS) NOTCH PROCEDURE, THE SURGEON INADVERTENTLY MISALIGNED THE HOUSING IMPACTOR COMPONENT OF THE SURGICAL INSTRUMENT PRIOR TO STRIKING IT WITH A MALLET. THE GUIDE ARMS OF THE SURGICAL INSTRUMENT WERE DAMAGED DURING THE IMPACTION PROCESS, LIKELY DUE TO THE MISALIGNMENT. A REPLACEMENT GUIDE ARM COMPONENT WAS OBTAINED FROM ANOTHER SET OF THE SAME SURGICAL INSTRUMENT AND USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT WAS UNAFFECTED BY THE INSTRUMENT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19612012 ↗
  • MalfunctionMay 8, 2024Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event; Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event

    THE SURGEON HAD PREPARED THE FEMORAL POSTERIOR STABILIZED (PS) NOTCH, WHICH IS A SPECIFIC AREA ON THE FEMUR BONE WHERE THE PS COMPONENT OF A TOTAL KNEE REPLACEMENT IS POSITIONED. DURING THE PROCEDURE, THE SURGEON REMOVED THE PS IMPACTOR, AND IT WAS NOTED THAT THE INSTRUMENT ASSEMBLY HAD BECOME DAMAGED. THIS TYPE OF DAMAGE TO A SURGICAL INSTRUMENT IS NOT UNCOMMON AND CAN OCCUR DUE TO THE FORCES AND STRESSES INVOLVED IN THE BONE PREPARATION PROCESS. DESPITE THE DAMAGE TO THE INSTRUMENT, THE SURGEON WAS ABLE TO COMPLETE THE SURGERY WITHOUT ANY DELAYS OR COMPLICATIONS. THIS WAS POSSIBLE BECAUSE THE PS NOTCH HAD ALREADY BEEN SUCCESSFULLY PREPARED PRIOR TO THE INSTRUMENT DAMAGE. THE DAMAGED INSTRUMENT HAD ALREADY SERVED ITS PURPOSE, AND THE SURGEON WAS ABLE TO PROCEED WITH THE REMAINDER OF THE SURGICAL STEPS. THE SURGERY WAS COMPLETED WITHOUT ANY ISSUES OR DELAYS, AS THE INSTRUMENT HAD ALREADY PERFORMED ITS INTENDED FUNCTION OF PREPARING THE PS NOTCH PRIOR TO THE DAMAGE OCCURRING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19271561 ↗
  • MalfunctionMay 8, 2024Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event; Break; Material Separation; Detachment of Device or Device Component; Material Deformation; No Apparent Adverse Event

    DURING THE PREPARATION FOR THE POSTERIOR STABILIZATION (PS) NOTCH PROCEDURE, THE SURGEON INADVERTENTLY MISALIGNED THE HOUSING IMPACTOR COMPONENT OF THE SURGICAL INSTRUMENT PRIOR TO STRIKING IT WITH A MALLET. THE GUIDE ARMS OF THE SURGICAL INSTRUMENT WERE DAMAGED DURING THE IMPACTION PROCESS, LIKELY DUE TO THE MISALIGNMENT. A REPLACEMENT GUIDE ARM COMPONENT WAS OBTAINED FROM ANOTHER SET OF THE SAME SURGICAL INSTRUMENT AND USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT WAS UNAFFECTED BY THE INSTRUMENT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19271555 ↗