Nihon Kohden SVM-7200 Series Vital Signs Monitor SVM-7260Recalls & reported problems

Nihon Kohden Corporation · Vital Signs Monitor with Nihon Kohden SpO2 technology of SVM-7200 Series Vital Signs Monitor · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04931921005903), received between Oct 4, 2024 and Sep 30, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings7
  • Output Problem6
  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948180 ↗
  • MalfunctionAug 7, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50176254 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE NIBP READINGS WERE INCONSISTENTLY NOT IN SPECIFICATIONS. THE NURSE JUST COULD NOT GET ACCURATE READINGS AND AT TIME NO READINGS. SOMETIMES MONITOR WOULD JUST KEEP PUMPING, SO THEY WOULD SHUT IT OFF AND USE ANOTHER MONITOR. THE BIOMEDICAL ENGINEER (BME) BROUGHT IT TO THE SHOP AND CONNECTED TO ONE OF THEIR PRONK SIMULATOR CUBES. THIS TEST DEVICE WILL RUN BLOOD PRESSURE, ECG, AND SPO2 SIMULATED FUNCTIONS. WHEN RUNNING BP SIMULATIONS A NORMAL READING OF 120/80, THE SPECIFICATIONS'S ARE + OR - 3. THIS MONITOR WOULD ONLY GET ONE OR TWO READINGS IN SPECIFICATION. THEN THE NEXT FEW WOULD-BE WAY OUT OF SPECIFICATION RANGE. THEY COULD NOT GET ENOUGH OF THEM WITHIN SPECIFICATION READINGS CONSISTENTLY. ALSO, THE PUMP MAKES ERRATIC, STRANGE OR UNUSUAL SOUNDS. NOT IN PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22722730 ↗
  • MalfunctionJul 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE VITAL SIGN MONITORS WERE NOT PROPERLY TAKING NIBP READINGS, AND THE PUMP SOUNDED LIKE IT WAS GIVING OUT. THEY MENTIONED THAT THE READINGS WERE IN THE 200/90+ RANGE. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22414184 ↗
  • MalfunctionJul 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE VITAL SIGN MONITORS WERE NOT PROPERLY TAKING NIBP READINGS, AND THE PUMP SOUNDED LIKE IT WAS GIVING OUT. THEY MENTIONED THAT THE READINGS WERE IN THE 200/90+ RANGE. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22413747 ↗
  • MalfunctionJul 2, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE VITAL SIGN MONITORS WERE NOT PROPERLY TAKING NIBP READINGS, AND THE PUMP SOUNDED LIKE IT WAS GIVING OUT. THEY MENTIONED THAT THE READINGS WERE IN THE 200/90+ RANGE. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22387152 ↗
  • MalfunctionOct 4, 2024Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THE VITAL SIGNS MONITOR (SVM) HAD INCORRECT NON-INVASIVE BLOOD PRESSURE (NIBP) READINGS WHILE IN PATIENT USE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20367150 ↗