Skip to content

NA ZM-540PARecalls & reported problems

Nihon Kohden Corporation · Transmitter, Multi Parameter, 608 UHF MHz · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04931921106600), received between Feb 21, 2017 and Jun 30, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • Power Problem2
  • Temperature Problem2
  • Alarm Not Visible1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Alarm System1

Patient problems

  • No Known Impact Or Consequence To Patient5
  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE ZM TRANSMITTER WOULD NOT STAY POWERED ON. HE STATED THAT WITH NEW BATTERIES, IT WOULD SOMETIMES POWER OFF ON ITS OWN AS IF THERE WAS A BATTERY TERMINAL OR CONNECTION ISSUE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25668723 ↗
  • MalfunctionJun 15, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THIS TRANSMITTER FAILED THE AIR LEAK TEST AND SHUTS DOWN. DISCOVERED DURING TESTING AND NOT IN PATIENT USE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505346 ↗
  • MalfunctionSep 24, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE NURSE REPORTED THAT THEY COULD ONLY SEE ONE LEAD UNDER THE 12-LEAD DATA PAGE ON THE CENTRAL NURSE STATION (CNS) FOR THIS ZM TRANSMITTER. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23133264 ↗
  • MalfunctionSep 10, 2020Display or Visual Feedback Problem; Output Problem; Display or Visual Feedback Problem; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TRANSMITTER WAS NOT READING ECG LEADS AND THERE WAS NO ERROR MESSAGE. THE MONITORING TECH NOTICED IT WHEN THEY WENT INTO FULL DISCLOSURE AND ONLY SAW A QRS WAVE ON LEAD 2. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10519156 ↗
  • MalfunctionFeb 28, 2019Overheating of Device; Overheating of Device

    THE BIOMED REPORTED THAT THE MIDDLE BATTERY IS SHORTING OUT AND BELIEVES IT IS THAT TERMINAL THAT IS GETTING HOT AND CAUSING THE DEVICE TO SHUT DOWN. CUSTOMER WAS PROVIDED AN EXCHANGE TRANSMITTER. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8380439 ↗
  • MalfunctionAug 31, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED THAT THE TRANSMITTER HAS AN SPO2 FAILURE. NO ADDITIONAL INFORMATION REGARDING THE EVENT IS AVAILABLE. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7839601 ↗
  • MalfunctionAug 10, 2018Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER GAVE INCORRECT ECG READINGS. THE HEART RATE WAS RANDOMLY GOING TO HIGH NUMBERS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7772666 ↗
  • MalfunctionAug 17, 2017Electrostatic Discharge; Temperature Problem; Electrostatic Discharge; Temperature Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE NURSE WAS SHOCKED WHEN TRYING TO PLACE BATTERIES INTO THE TRANSMITTER. THE BIOMEDICAL ENGINEER THEN TESTED THE DEVICE AND THE BATTERIES OVERHEATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6802653 ↗
  • MalfunctionFeb 21, 2017Device Alarm System; No Audible Alarm; Alarm Not Visible; Device Alarm System; No Audible Alarm; Alarm Not Visible

    THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER DID NOT V-FIB ALARM WHILE THE PATIENT WAS BEING MONITORED ON THE CNS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6348143 ↗
  • MalfunctionFeb 21, 2017Temperature Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THE TRANSMITTER'S BATTERY OVERHEATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6348134 ↗

Other MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) devices

All MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page