NA ZM-520PARecalls & reported problems
Nihon Kohden Corporation · Transmitter, ECG and Resp, 608 UHF MHz · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Nihon Kohden ZM-520PA transmitter sends ECG and breathing signals for cardiac monitoring. The 24 reports mostly describe the unit overheating or getting hot, with ECG dropouts, noisy waveforms, gaps in recorded data and missed alarms, and mention no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 04931921115077), received between Apr 4, 2017 and Sep 16, 2025 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device12
- Temperature Problem11
- Output Problem5
- Incorrect, Inadequate or Imprecise Result or Readings3
- Device Issue2
- Patient Data Problem2
- Smoking2
- Application Program Problem1
Patient problems
- No Known Impact Or Consequence To Patient14
- No Clinical Signs, Symptoms or Conditions12
- No Consequences Or Impact To Patient5
- No Information1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2025No Device Output; Output Problem; No Device Output; Output Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER WILL NO LONGER READ THE VA LEAD. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23069132 ↗MalfunctionFeb 2, 2024Display or Visual Feedback Problem; Image Display Error/Artifact; Display or Visual Feedback Problem; Image Display Error/Artifact
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE ZM TRANSMITTER WAS SHOWING ARTIFACT/NOISES WITHIN THE WAVEFROMS WHILE IN PATIENT USE. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN TECHNICAL SUPPORT HAD THE BME UN-MONITOR AND RE- MONITOR THE DEVICE AND THE BME REPORTED THE WAVEFORM BECOMES INTERMITTENT AFTERWARDS, THAT IT WOULD HAVE NO ARTIFACTS AT FIRST THEN WITHIN 1 MINUTE IT STARTS HAVING ARTIFACTS. NK TECHNICAL SUPPORT HAD THE BME TURN ON THE HUM FILTER WITHIN THE CNS BUT WHEN THE BME DID THE PROBLEM STILL PERSISTED. NK TECHNICAL SUPPORT HAD THE BME CHANGE THE ZM WITH NEW CABLES AND CLEAN THE CONNECTOR ON THE ZM ITSELF BUT THE ISSUE IS STILL PERSISTING. NK TECHNICAL SUPPORT HAD THE BME CHANGE THE ZM TRANSMITTER WITH A SPARE UNIT AND THE WAVEFORMS WERE NO LONGER HAVING ARTIFACTS. BME REQUESTED TO SEND THEIR ZM TRANSMITTER INTO NIHON KOHDEN FOR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18636926 ↗MalfunctionFeb 16, 2021Computer Software Problem; Data Problem; Patient Data Problem; Computer Software Problem; Data Problem; Patient Data Problem
THE BIOMED REPORTED THAT THE PREVIOUS DAY A PATIENT HAD CODED AND THEY NOTICED DATA WAS NOT RECORDED. THE DATA HAD ABOUT A 10-11MIN GAP WHERE IT WAS BLANK. WAVEFORMS WERE STILL DISPLAYING ON THE SCREEN BUT NOT RECORDED. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11335638 ↗MalfunctionFeb 16, 2021Patient Data Problem; Patient Data Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT A PATIENT CODED THE PREVIOUS DAY WHILE BEING MONITORED ON A TELEMETRY TRANSMITTER AND FOUND THAT THERE WAS A GAP OF ABOUT 10-11 MINUTES OF RECORDED DATA. WAVEFORMS WERE STILL DISPLAYING ON THE SCREEN BUT NOT RECORDING. THERE WAS NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11334566 ↗MalfunctionSep 16, 2020Overheating of Device; Unable to Obtain Readings; Output Problem; Temperature Problem; Overheating of Device; Unable to Obtain Readings; Output Problem; Temperature Problem
THE CUSTOMER REPORTED THAT THE TRANSMITTER WAS EXPERIENCING ECG DROPOUTS. THEY ALSO BELIEVE THE ECG BOARD MIGHT BE OVERHEATING. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10543551 ↗MalfunctionJul 7, 2020Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER STARTED TO HEAT UP AND GOT TOO HOT TO TOUCH OR HOLD. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10242149 ↗MalfunctionJun 18, 2020Device Alarm System; Protective Measures Problem; Device Alarm System; Protective Measures Problem
THE CUSTOMER REPORTED THAT THEIR TRANSMITTER MISSED AN ARRHYTHMIA ALARM. NO PATIENT HARM OR INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10171763 ↗MalfunctionMay 20, 2020Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE BIOMEDICAL ENGINEER REPORTED THAT THE BATTERIES WERE HOT IN THE TELEMETRY TRANSMITTER. THEY REPLACED THE BATTERIES, BUT THE ISSUE PERSISTED. THEY REQUESTED THAT THE UNIT BE REPAIRED. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10075969 ↗MalfunctionMar 19, 2020Overheating of Device; Battery Problem; Overheating of Device; Battery Problem
THE CUSTOMER REPORTED THAT THEIR TRANSMITTER BATTERY WAS SHORTING IF THEY PLACED THE BATTERY INSIDE OF THE TRANSMITTER WITH THE NEGATIVE END FIRST.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9857827 ↗MalfunctionDec 13, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER OVERHEATED AND THE TEMPERATURE FEELS HIGHER THAN USUAL, IN THE REAR OF THE TRANSMITTER, PARTICULARLY IN THE BATTERY AREA WHERE THE BATTERIES CONNECT. THE TRANSMITTER STILL WORKS AND TRANSMITS WAVEFORMS AND VITAL SIGNS AS NORMAL. THERE IS NOTHING ELSE WRONG WITH IT OTHER THAN IT BEING SUPER WARM IN THE REAR, BATTERY AREA. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9470244 ↗MalfunctionNov 25, 2019Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE CUSTOMER REPORTED THAT ONE OF THEIR TRANSMITTER WAS NOT DISPLAYING ANY HEART RATE READINGS. THE CUSTOMER ALSO REPORTED THAT THIS TRANSMITTER WAS NOT SENDING THE HEART RATE READING TO THE CENTRAL NURSE'S STATION (CNS).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9373390 ↗MalfunctionNov 11, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE CUSTOMER REPORTS THAT THE TRANSMITTER BATTERIES WILL GET VERY HOT WHEN PUT IN THE DEVICE AND IT WILL NOT TURN ON. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9307254 ↗MalfunctionAug 29, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE CUSTOMER REPORTED THAT THE TRANSMITTER OVERHEATED CAUSING THE BATTERY COMPARTMENT TO MELT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8947890 ↗MalfunctionAug 14, 2019Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TRANSMITTER SENT NO WAVE OR NUMERICS TO THE CENTRAL NURSE'S STATION (CNS) AND DID NOT DISPLAY THEM LOCALLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 8895549 ↗MalfunctionMar 1, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE CUSTOMER REPORTED THAT THE TRANSMITTER'S BATTERY OVERHEATED AND SPLIT APART.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8382895 ↗MalfunctionDec 26, 2018Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED SIGNAL LOSS ON MULTIPLE CHANNELS. NO CONSEQUENCE OR IMPACT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8198083 ↗MalfunctionAug 21, 2018Application Program Problem; Unintended Application Program Shut Down; Application Program Problem; Unintended Application Program Shut Down
THE CUSTOMER REPORTED THAT THE TRANSMITTER DROPPED TRANSMISSION OF THE PATIENT'S VITAL SIGNS WHILE ATTEMPTING A PATIENT TRANSFER BACK TO THE BEDSIDE. THE TRANSMITTER'S DISPLAY ALSO WENT BLANK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7803709 ↗MalfunctionAug 8, 2018Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem
THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER OVERHEATED AND MELTED THE BATTERY COMPARTMENT. THEY SENT THE TRANSMITTER IN FOR EVALUATION AND RECEIVED AN EXCHANGE. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME. THE BATTERIES REQUIRED FOR THIS INVESTIGATION WERE NOT RETURNED. NEW BATTERIES WERE INSERTED INTO THE UNIT AND HEATING COULD NOT BE DUPLICATED. INSPECTION OF THE NEGATIVE CONTACTS SHOWED THAT THERE WERE RESIN MELT MARKINGS AT THE SPRING, WHICH PER AN NKC INVESTIGATION ON A SIMILAR INCIDENT, IS INDICATIVE OF IMPROPER BATTERY INSERTION. ALSO NOTED WAS CORROSION AT THE POSITIVE CONTACT INDICATING FLUID INTRUSION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7764782 ↗MalfunctionJun 28, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Interpretation of Signal; Invalid Sensing; Device Sensing Problem; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Interpretation of Signal; Invalid Sensing; Device Sensing Problem; Output Problem
THE CUSTOMER REPORTED THAT THE TRANSMITTER IS NOT GETTING CORRECT READINGS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7648783 ↗MalfunctionApr 20, 2018Overheating of Device; Device Issue; Human Factors Issue; Misassembly by Users; Overheating of Device; Device Issue; Human Factors Issue; Misassembly by Users
THE CUSTOMER REPORTED THAT THE TRANSMITTER WAS OVERHEATING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7448219 ↗MalfunctionMar 13, 2018Inappropriate/Inadequate Shock/Stimulation; Device Issue; Inappropriate/Inadequate Shock/Stimulation; Device Issue
THE CUSTOMER REPORTED THAT A PATIENT COMPLAINED OF GETTING SHOCKED BY THE ECG ELECTRODES THAT WERE ATTACHED TO THE TRANSMITTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7335978 ↗MalfunctionFeb 28, 2018Thermal Decomposition of Device; Overheating of Device; Temperature Problem; Thermal Decomposition of Device; Overheating of Device; Temperature Problem
THE NURSE REPORTED THAT THE BATTERIES IN THE TRANSMITTER WERE OVERHEATING BUT THE BIOMEDICAL ENGINEER REPORTED THERE WAS NO OVERHEATING AND THAT HE KEPT GETTING SPORADIC READINGS, A "CHECK LEADS" ERROR MESSAGE, AND A FLAT LINE WHILE ON THE SIMULATOR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7304956 ↗MalfunctionSep 11, 2017Overheating of Device; Smoking; Temperature Problem; Overheating of Device; Smoking; Temperature Problem
THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER OVERHEATED AND WAS SMOKING. THE BATTERY CONTACTS WERE BURNED AND MELTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6855773 ↗MalfunctionApr 4, 2017Device Emits Odor; Smoking; Temperature Problem; Temperature Problem
THE CUSTOMER REPORTED THAT WHILE PREPARING THE TRANSMITTER FOR PATIENT USE, THE TRANSMITTER HAD A BURNING SMELL AND BEGAN TO SMOKE. THE TRANSMITTER HAD JUST BEEN USED ON ANOTHER PATIENT WITHOUT ISSUE.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 6459175 ↗
FDA clearance and classification
The FDA regulates product code DRT (monitor, cardiac (incl. cardiotachometer & rate alarm)) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA ZM-520PA
Has the NA ZM-520PA been recalled?
No FDA recall record in our data names this exact device identifier (DI 04931921115077), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA ZM-520PA?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between Apr 4, 2017 and Sep 16, 2025. In those reports, the most common specific device problems are “Overheating of Device” (12), “Temperature Problem” (11) and “Output Problem” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA ZM-520PA?
None of the reports are classified as deaths, 0 as injuries and 24 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the NA ZM-520PA, and is it still sold?
The FDA's device database lists Nihon Kohden Corporation as the company responsible for it (brand NA, model ZM-520PA). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) devices
- NA ZM-531PANihon Kohden Corporation · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)1 death167 reports
- NA B155 VSP3.0 Patient MonitorGE Medical Systems Information Technologies, Inc. · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)2 recalls0 reports
- NA ZM-530PANihon Kohden Corporation · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)2 deaths93 reports
- NA ZM-521PANihon Kohden Corporation · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)79 reports
- NA ZM-541PANihon Kohden Corporation · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)22 reports
- NA ZM-540PANihon Kohden Corporation · MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page