Nihon Kohden Corporation · Transmitter, ECG and Resp, 608 UHF MHz · Class 2

Quick summary

This Nihon Kohden ZM-520PA transmitter sends ECG and breathing signals for cardiac monitoring. The 24 reports mostly describe the unit overheating or getting hot, with ECG dropouts, noisy waveforms, gaps in recorded data and missed alarms, and mention no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 04931921115077), received between Apr 4, 2017 and Sep 16, 2025 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device12
  • Temperature Problem11
  • Output Problem5
  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Device Issue2
  • Patient Data Problem2
  • Smoking2
  • Application Program Problem1

Patient problems

  • No Known Impact Or Consequence To Patient14
  • No Clinical Signs, Symptoms or Conditions12
  • No Consequences Or Impact To Patient5
  • No Information1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025No Device Output; Output Problem; No Device Output; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER WILL NO LONGER READ THE VA LEAD. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23069132 ↗
  • MalfunctionFeb 2, 2024Display or Visual Feedback Problem; Image Display Error/Artifact; Display or Visual Feedback Problem; Image Display Error/Artifact

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE ZM TRANSMITTER WAS SHOWING ARTIFACT/NOISES WITHIN THE WAVEFROMS WHILE IN PATIENT USE. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN TECHNICAL SUPPORT HAD THE BME UN-MONITOR AND RE- MONITOR THE DEVICE AND THE BME REPORTED THE WAVEFORM BECOMES INTERMITTENT AFTERWARDS, THAT IT WOULD HAVE NO ARTIFACTS AT FIRST THEN WITHIN 1 MINUTE IT STARTS HAVING ARTIFACTS. NK TECHNICAL SUPPORT HAD THE BME TURN ON THE HUM FILTER WITHIN THE CNS BUT WHEN THE BME DID THE PROBLEM STILL PERSISTED. NK TECHNICAL SUPPORT HAD THE BME CHANGE THE ZM WITH NEW CABLES AND CLEAN THE CONNECTOR ON THE ZM ITSELF BUT THE ISSUE IS STILL PERSISTING. NK TECHNICAL SUPPORT HAD THE BME CHANGE THE ZM TRANSMITTER WITH A SPARE UNIT AND THE WAVEFORMS WERE NO LONGER HAVING ARTIFACTS. BME REQUESTED TO SEND THEIR ZM TRANSMITTER INTO NIHON KOHDEN FOR REPAIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18636926 ↗
  • MalfunctionFeb 16, 2021Computer Software Problem; Data Problem; Patient Data Problem; Computer Software Problem; Data Problem; Patient Data Problem

    THE BIOMED REPORTED THAT THE PREVIOUS DAY A PATIENT HAD CODED AND THEY NOTICED DATA WAS NOT RECORDED. THE DATA HAD ABOUT A 10-11MIN GAP WHERE IT WAS BLANK. WAVEFORMS WERE STILL DISPLAYING ON THE SCREEN BUT NOT RECORDED. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11335638 ↗
  • MalfunctionFeb 16, 2021Patient Data Problem; Patient Data Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT A PATIENT CODED THE PREVIOUS DAY WHILE BEING MONITORED ON A TELEMETRY TRANSMITTER AND FOUND THAT THERE WAS A GAP OF ABOUT 10-11 MINUTES OF RECORDED DATA. WAVEFORMS WERE STILL DISPLAYING ON THE SCREEN BUT NOT RECORDING. THERE WAS NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11334566 ↗
  • MalfunctionSep 16, 2020Overheating of Device; Unable to Obtain Readings; Output Problem; Temperature Problem; Overheating of Device; Unable to Obtain Readings; Output Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THE TRANSMITTER WAS EXPERIENCING ECG DROPOUTS. THEY ALSO BELIEVE THE ECG BOARD MIGHT BE OVERHEATING. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10543551 ↗
  • MalfunctionJul 7, 2020Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER STARTED TO HEAT UP AND GOT TOO HOT TO TOUCH OR HOLD. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10242149 ↗
  • MalfunctionJun 18, 2020Device Alarm System; Protective Measures Problem; Device Alarm System; Protective Measures Problem

    THE CUSTOMER REPORTED THAT THEIR TRANSMITTER MISSED AN ARRHYTHMIA ALARM. NO PATIENT HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10171763 ↗
  • MalfunctionMay 20, 2020Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE BATTERIES WERE HOT IN THE TELEMETRY TRANSMITTER. THEY REPLACED THE BATTERIES, BUT THE ISSUE PERSISTED. THEY REQUESTED THAT THE UNIT BE REPAIRED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10075969 ↗
  • MalfunctionMar 19, 2020Overheating of Device; Battery Problem; Overheating of Device; Battery Problem

    THE CUSTOMER REPORTED THAT THEIR TRANSMITTER BATTERY WAS SHORTING IF THEY PLACED THE BATTERY INSIDE OF THE TRANSMITTER WITH THE NEGATIVE END FIRST.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 9857827 ↗
  • MalfunctionDec 13, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TELEMETRY TRANSMITTER OVERHEATED AND THE TEMPERATURE FEELS HIGHER THAN USUAL, IN THE REAR OF THE TRANSMITTER, PARTICULARLY IN THE BATTERY AREA WHERE THE BATTERIES CONNECT. THE TRANSMITTER STILL WORKS AND TRANSMITS WAVEFORMS AND VITAL SIGNS AS NORMAL. THERE IS NOTHING ELSE WRONG WITH IT OTHER THAN IT BEING SUPER WARM IN THE REAR, BATTERY AREA. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9470244 ↗
  • MalfunctionNov 25, 2019Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT ONE OF THEIR TRANSMITTER WAS NOT DISPLAYING ANY HEART RATE READINGS. THE CUSTOMER ALSO REPORTED THAT THIS TRANSMITTER WAS NOT SENDING THE HEART RATE READING TO THE CENTRAL NURSE'S STATION (CNS).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9373390 ↗
  • MalfunctionNov 11, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE CUSTOMER REPORTS THAT THE TRANSMITTER BATTERIES WILL GET VERY HOT WHEN PUT IN THE DEVICE AND IT WILL NOT TURN ON. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 9307254 ↗
  • MalfunctionAug 29, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE CUSTOMER REPORTED THAT THE TRANSMITTER OVERHEATED CAUSING THE BATTERY COMPARTMENT TO MELT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8947890 ↗
  • MalfunctionAug 14, 2019Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE TRANSMITTER SENT NO WAVE OR NUMERICS TO THE CENTRAL NURSE'S STATION (CNS) AND DID NOT DISPLAY THEM LOCALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 8895549 ↗
  • MalfunctionMar 1, 2019Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE CUSTOMER REPORTED THAT THE TRANSMITTER'S BATTERY OVERHEATED AND SPLIT APART.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8382895 ↗
  • MalfunctionDec 26, 2018Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED SIGNAL LOSS ON MULTIPLE CHANNELS. NO CONSEQUENCE OR IMPACT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8198083 ↗
  • MalfunctionAug 21, 2018Application Program Problem; Unintended Application Program Shut Down; Application Program Problem; Unintended Application Program Shut Down

    THE CUSTOMER REPORTED THAT THE TRANSMITTER DROPPED TRANSMISSION OF THE PATIENT'S VITAL SIGNS WHILE ATTEMPTING A PATIENT TRANSFER BACK TO THE BEDSIDE. THE TRANSMITTER'S DISPLAY ALSO WENT BLANK.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7803709 ↗
  • MalfunctionAug 8, 2018Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER OVERHEATED AND MELTED THE BATTERY COMPARTMENT. THEY SENT THE TRANSMITTER IN FOR EVALUATION AND RECEIVED AN EXCHANGE. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME. THE BATTERIES REQUIRED FOR THIS INVESTIGATION WERE NOT RETURNED. NEW BATTERIES WERE INSERTED INTO THE UNIT AND HEATING COULD NOT BE DUPLICATED. INSPECTION OF THE NEGATIVE CONTACTS SHOWED THAT THERE WERE RESIN MELT MARKINGS AT THE SPRING, WHICH PER AN NKC INVESTIGATION ON A SIMILAR INCIDENT, IS INDICATIVE OF IMPROPER BATTERY INSERTION. ALSO NOTED WAS CORROSION AT THE POSITIVE CONTACT INDICATING FLUID INTRUSION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7764782 ↗
  • MalfunctionJun 28, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Interpretation of Signal; Invalid Sensing; Device Sensing Problem; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Interpretation of Signal; Invalid Sensing; Device Sensing Problem; Output Problem

    THE CUSTOMER REPORTED THAT THE TRANSMITTER IS NOT GETTING CORRECT READINGS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7648783 ↗
  • MalfunctionApr 20, 2018Overheating of Device; Device Issue; Human Factors Issue; Misassembly by Users; Overheating of Device; Device Issue; Human Factors Issue; Misassembly by Users

    THE CUSTOMER REPORTED THAT THE TRANSMITTER WAS OVERHEATING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7448219 ↗
  • MalfunctionMar 13, 2018Inappropriate/Inadequate Shock/Stimulation; Device Issue; Inappropriate/Inadequate Shock/Stimulation; Device Issue

    THE CUSTOMER REPORTED THAT A PATIENT COMPLAINED OF GETTING SHOCKED BY THE ECG ELECTRODES THAT WERE ATTACHED TO THE TRANSMITTER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7335978 ↗
  • MalfunctionFeb 28, 2018Thermal Decomposition of Device; Overheating of Device; Temperature Problem; Thermal Decomposition of Device; Overheating of Device; Temperature Problem

    THE NURSE REPORTED THAT THE BATTERIES IN THE TRANSMITTER WERE OVERHEATING BUT THE BIOMEDICAL ENGINEER REPORTED THERE WAS NO OVERHEATING AND THAT HE KEPT GETTING SPORADIC READINGS, A "CHECK LEADS" ERROR MESSAGE, AND A FLAT LINE WHILE ON THE SIMULATOR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7304956 ↗
  • MalfunctionSep 11, 2017Overheating of Device; Smoking; Temperature Problem; Overheating of Device; Smoking; Temperature Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE TRANSMITTER OVERHEATED AND WAS SMOKING. THE BATTERY CONTACTS WERE BURNED AND MELTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6855773 ↗
  • MalfunctionApr 4, 2017Device Emits Odor; Smoking; Temperature Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT WHILE PREPARING THE TRANSMITTER FOR PATIENT USE, THE TRANSMITTER HAD A BURNING SMELL AND BEGAN TO SMOKE. THE TRANSMITTER HAD JUST BEEN USED ON ANOTHER PATIENT WITHOUT ISSUE.

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 6459175 ↗

Other MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) devices

All MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) devices →