Skip to content

NA GZ-140PRecalls & reported problems

Nihon Kohden Corporation · Vital Sign Telemeter · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04931921128244), received between Aug 14, 2019 and Aug 8, 2023 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • Data Problem1
  • Display or Visual Feedback Problem1
  • No Audible Alarm1
  • No Display/Image1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2023No Audible Alarm; Protective Measures Problem; No Audible Alarm; Protective Measures Problem

    THE CUSTOMER REPORTED THAT THIS GZ TRANSMITTER IS NOT OUTPUTTING ANY SOUND. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17484748 ↗
  • MalfunctionMay 11, 2023Data Problem; Data Problem

    THE CUSTOMER REPORTED THAT THE SPO2 WAS SET ON THE SIMULATOR TO 80%, BUT THE READINGS ON THE GZ TRANSMITTER WERE NEAR 85% WHICH IS OUT OF RANGE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16916735 ↗
  • MalfunctionMay 10, 2023Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE GZ TRANSMITTER WOULD NOT SHOW THE ECG WAVES OR NUMERICS ON THE WIRELESS SURGE SOLUTION (WSS) OR ON THE TRANSMITTER. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16907554 ↗
  • MalfunctionSep 6, 2020No Display/Image; Communication or Transmission Problem; Communication or Transmission Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE GZ TRANSMITTER WAS NOT DISPLAYING WAVEFORMS, BUT ALL OTHER PARAMETERS AVAILABLE. ISSUE WAS RESOLVED BY REBOOTING THE UNIFIED GATEWAY. NO ERROR MESSAGES DISPLAYED ON THE CENTRAL NURSE'S STATIONS (CNS) WHEN THIS WAS HAPPENING. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10498423 ↗
  • MalfunctionAug 4, 2020Display or Visual Feedback Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Display or Visual Feedback Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THE TELEMETRY TRANSMITTER WAS NOT DISPLAYING WAVEFORMS AND ONLY DISPLAYING NUMERICS. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10360506 ↗
  • MalfunctionAug 14, 2019Communication or Transmission Problem; Communication or Transmission Problem

    THE BIOMEDICAL ENGINEER REPORTED THAT THEY ARE HAVING COMMUNICATION ISSUES WITH THE TELEMETRY TRANSMITTERS, AND IT IS VERY INTERMITTENT ACROSS THE 5TH AND 6TH FLOOR. THE ISSUE HAS BEEN GOING ON SINCE THE INSTALL IN FEBRUARY. THIS WAS IN PATIENT USE. WE INQUIRED IF THERE WAS ANY PATIENT HARM REPORTED, AND A RESPONSE WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 8895588 ↗

FDA clearance and classification

The FDA regulates product code DRG (transmitters and receivers, physiological signal, radiofrequency) under 21 CFR 870.2910 as a Class II device. 21 CFR 870.2910 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA GZ-140P

Has the NA GZ-140P been recalled?

No FDA recall record in our data names this exact device identifier (DI 04931921128244), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA GZ-140P?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Aug 14, 2019 and Aug 8, 2023. In those reports, the most common specific device problems are “Communication or Transmission Problem” (2), “Incorrect, Inadequate or Imprecise Result or Readings” (2) and “Output Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA GZ-140P?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA GZ-140P MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the NA GZ-140P, and is it still sold?

The FDA's device database lists Nihon Kohden Corporation as the company responsible for it (brand NA, model GZ-140P). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices

All TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page