BIOPSY FORCEPS FB-19K-1Recalls & reported problems
Olympus Medical Systems Corp. · FB-19K-1 STAND FENEST BIOPSY FORCEPS · Class 2
All models of BIOPSY FORCEPS (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04953170029608), received between Apr 13, 2022 and Jul 24, 2024 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately6
- Contamination3
- Contamination /Decontamination Problem1
- Device Reprocessing Problem1
- Difficult to Open or Close1
- Microbial Contamination of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BIOPSY FORCEPS EXHIBITED PROBLEMS CAUSED BY IMPROPER REPROCESSING (FOREIGN MATTER). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19816380 ↗MalfunctionDec 4, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THE BIOPSY FORCEPS WERE NOT OPENING OR CLOSING. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE GRASPING CUPS PREVENTING IT FROM OPENING AND CLOSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18257761 ↗MalfunctionJul 27, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THE BIOPSY FORCEPS HANDLE DID NOT MOVE AND THE CUP DID NOT OPEN WHEN TRYING TO SLIDE THE HANDLE. THE REPORTED ISSUE OCCURRED DURING PREPARATION OF USE FOR AN UNKNOWN PROCEDURE. THE PROCEDURE WAS SUBSEQUENTLY COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT. UPON DEVICE RETURN AND EVALUATION, IT WAS OBSERVED THAT THERE WAS FOREIGN MATERIAL THAT ADHERED TO THE LINKAGE MECHANISM AND TO THE INSIDE OF THE CUP WHICH IS IDENTIFIED AS A REPORTABLE EVENT PER THE RISK SUMMARY APPLICATION (RSA). THE NEW AWARE DATE FOR THIS REPORTABLE MALFUNCTION IS (B)(6) 2023. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17406948 ↗MalfunctionJun 29, 2023Contamination /Decontamination Problem; Failure to Clean Adequately; Contamination /Decontamination Problem; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT THE BIOPSY FORCEPS EXHIBITED SLOW CUP MOVEMENT. UPON INSPECTION AND TESTING OF THE CUSTOMER RETURNED DEVICE, FOREIGN MATERIAL WAS FOUND ADHERED TO THE LINKAGE MECHANISM DUE TO INSUFFICIENT CLEANING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION RECEIVED THAT AFTER CLEANING/STERILIZING, THE ACTUAL PRODUCT WAS STORED FOR USE IN THE NEXT CASE AND BECAUSE THE PREVIOUS CASE DID NOT WORK WELL, THE CUSTOMER CHECKED THE OPERATION OF THE CUP, AND DISCOVERED THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17226005 ↗MalfunctionOct 1, 2022Contamination; Failure to Clean Adequately; Contamination; Failure to Clean Adequately
CUSTOMER RETURNED THE DEVICE FOR EVALUATION AND REPAIR FOR THE CUP AT THE TIP BEING BROKEN DURING REPROCESSING AFTER FIRST TIME USE OF THE DEVICE. THERE IS NO PATIENT INVOLVEMENT. THE PREVIOUS BIOPSY PROCEDURE WAS SUCCESSFULLY COMPLETED. THE REPROCESSING HAD BEEN PERFORMED BY IMMERSION IN DISINFECTANT, IMMERSION TIME AND DISINFECTANT TYPE IS UNKNOWN. UPON EVALUATION OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE ARM OF THE CUP (A PART OF THE LINKAGE MECHANISM AT THE DISTAL END) WERE BROKEN. THE BROKEN ARMS HAD BROWN FOREIGN MATERIALS AROUND THE BROKEN PLACE. THE BROKEN ARMS WERE NOT MISSING. THIS MEDWATCH IS BEING SUBMITTED FOR THE FOREIGN MATERIAL FOUND ON THE DEVICE DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15522894 ↗MalfunctionSep 13, 2022Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS THAT WHILE USING THE BIOPSY FORCEPS, THE CUPS COULD NO LONGER BE OPENED AND CLOSED, AFTER BEING USED SEVERAL TIMES SINCE THEY WERE BRAND NEW. THE ISSUE WAS FOUND AT PREPARATION FOR USE, THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE AND THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED AND EVALUATED, AND FOREIGN MATERIAL WAS FOUND ADHERING TO THE FORCEPS. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15414913 ↗MalfunctionJun 14, 2022Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED THAT, DURING PREPARATION FOR USE, THERE WAS A CATCH WHEN THE SLIDER WAS MOVED BACK AND FORTH DURING THE PRE-CASE INSPECTION. THE CUP OPENED AND CLOSED, BUT THERE WAS A CATCH. THE PROCEDURE WAS COMPLETED WITH ANOTHER SIMILAR DEVICE. THE SUBJECT DEVICE WAS SENT TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. DURING INSPECTION AND TESTING, FOREIGN MATTER WAS FOUND ON THE FORCEPS. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (FOREIGN MATERIAL). THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14690003 ↗MalfunctionJun 6, 2022Contamination; Difficult to Open or Close; Contamination; Difficult to Open or Close
THE CUSTOMER REPORTED THE BIOPSY FORCEPS CUP FAILED TO OPEN. THE EVENT OCCURRED DURING PREPARATION FOR USE AND A SIMILAR DEVICE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14606658 ↗MalfunctionMay 18, 2022Contamination; Contamination
A USER FACILITY REPORTED THAT THE BIOPSY FORCEPS CUP FAILED TO OPEN. THE EVENT OCCURRED DURING PREPARATION FOR USE AND A SIMILAR DEVICE WAS USED TO COMPLETE THE OPERATION. DURING A STANDARD SERVICE INSPECTION OF THE CUSTOMER RETURNED DEVICE, FOREIGN MATTER WAS ADHERED TO THE FORCEPS. THIS REPORT IS TO CAPTURE THE REPORTABLE MALFUNCTION FOUND AT INSPECTION. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14427387 ↗MalfunctionApr 13, 2022Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER ORIGINALLY RETURNED HIS OLYMPUS BIOPSY FORCEPS AS DURING AN UPPER ENDOSCOPY CASE, THE CUPS WOULD NOT CLOSE. ACCORDING TO THE INITIAL REPORTER, THE FORCEPS AND SCOPE WERE BOTH REMOVED AND THE PROCEDURE WAS COMPLETED USING ANOTHER SET OF FORCEPS WITH NO PATIENT HARM REPORTED. DURING THE DEVICE EVALUATION, IT WAS DISCOVERED THAT THE BIOPSY FORCEPS HAD EXPERIENCED A REPROCESSING ISSUE CAUSING THEM TO RUST. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE REPROCESSING ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14104958 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page