ROTATABLE GRASPING FORCEPS FG-44NR-1Recalls & reported problems
Olympus Medical Systems Corp. · FG-44NR-1 ROTATABLE RAT TOOTH GRASP FORC · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170032196), received between Sep 9, 2020 and Apr 8, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Corroded1
- Device Reprocessing Problem1
- Failure to Clean Adequately1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Hemorrhage/Blood Loss/Bleeding1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2024Break; Break
THE CUSTOMER REPORTED THE DISTAL END OF THE ROTATABLE GRASPING FORCEPS WAS FAULTY. THE ISSUE WAS FOUND WHEN PREPARING FOR A THERAPEUTIC PROCEDURE TO REMOVE A FOREIGN BODY. THIS WAS THE SECOND TIME THE DEVICE WAS GOING TO BE USED. THE DISTAL END BROKE OUTSIDE OF THE PATIENT'S BODY AND A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PROCEDURE WAS NOT DELAYED OVER 15 MINUTES. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19060136 ↗MalfunctionMar 8, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS REPORTED, THE GRASPING FORCEPS HAD A BROWN CHARRED SUBSTANCE OBSERVED INSIDE THE STERILIZED PACK. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18866151 ↗MalfunctionMar 1, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED, THE ROTATABLE GRASPING FORCEPS HAD FOREIGN MATERIAL IN THE STERILE PACK. THIS ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18821965 ↗MalfunctionSep 14, 2022Corroded; Corroded
THE CUSTOMER REPORTED THAT, DURING REPROCESSING OF THE SUBJECT DEVICE, A BROWN RUST-LIKE SUBSTANCE WAS FOUND FLOATING IN THE CLEANING LIQUID AFTER THE DEVICE WAS IMMERSED. THE USER DETERMINED THE SUBSTANCE DERIVED FROM THE SUBJECT DEVICE. THE PERSON IN CHARGE OF THE FACILITY POINTED OUT THAT THE SUBJECT DEVICE RUSTED QUICKLY AND DETERIORATED FASTER THAN OTHER REUSABLE TREATMENT TOOLS. THE SUBJECT DEVICE WAS NOT USED AND THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15424221 ↗InjuryOct 4, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED "ENDOSCOPIC TREATMENT OF INTRADUCTAL PANCREATIC STENT FRAGMENTATION". THE LITERATURE REPORTED THE RESULT OF 33 PATIENTS WITH ENDOSCOPIC TREATMENT OF INTRADUCTAL PANCREATIC STENT FRAGMENTATION USING UNSPECIFIED ENDOSCOPE AND OLYMPUS ROTATABLE GRASPING FORCEPS (FG-44NR-1 OR FG-V422PR) OR ELECTROSURGICAL SNARE (SD-9U-1) FROM 2001 TO 2015. IN THE SUBJECT CASES, PERIAMPULLARY DUODENAL BLEEDING OCCURRED IN 2 PATIENTS. THEREFORE, OMSC WILL SUBMIT 2 MEDICAL DEVICE REPORTS (MDR) DEPENDING ON THE DEVICE AND EVENT TYPE. THIS REPORT IS 1 OF 2 REPORTS OF ROTATABLE GRASPING FORCEPS.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 10624704 ↗MalfunctionSep 9, 2020Break; Break
IT WAS REPORTED THAT FAULT WAS FOUND AT DISTAL END OF THE SUBJECT DEVICE AND THE PART BROKE OUTSIDE THE PATIENT DURING AN INSPECTION BEFORE USE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. ON (B)(6) 2020, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE ARM OF THE GRASPING JAWS WAS PARTIALLY MISSING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10512474 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.